Accuracy and Practical Considerations for Doubly Labeled Water Analysis in Nutrition Studies Using a Laser-Based Isotope Instrument (Off-Axis Integrated Cavity Output Spectroscopy)

BACKGROUND: Given the utility of the doubly labeled water (DLW) method for determination of energy expenditure, additional techniques for isotope analysis of the samples are welcome. Laser-based instruments are one such new analytical tool, but their accuracy and feasibility for DLW studies are gros...

Descripción completa

Detalles Bibliográficos
Autores principales: Reynard, Linda M, Wong, William W, Tuross, Noreen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8754563/
https://www.ncbi.nlm.nih.gov/pubmed/34718673
http://dx.doi.org/10.1093/jn/nxab324
Descripción
Sumario:BACKGROUND: Given the utility of the doubly labeled water (DLW) method for determination of energy expenditure, additional techniques for isotope analysis of the samples are welcome. Laser-based instruments are one such new analytical tool, but their accuracy and feasibility for DLW studies are grossly understudied. OBJECTIVES: We assessed the accuracy of laser-based isotope ratio measurements as part of the DLW method for estimation of carbon dioxide production rate (rCO(2)) and total energy expenditure (TEE), in between-group comparison study designs. METHODS: Urine samples from a previous study were analyzed with a laser-based instrument [off-axis integrated cavity output spectroscopy (OA-ICOS)]. In that study, participants consumed a high-, moderate-, or low-carbohydrate diet for 20 wk; urine samples were obtained in weeks 18–20 before and after a (2)H- and (18)O-enriched water dose. Isotope ratios (δ(2)H and δ(18)O), rCO(2), and TEE calculated by standard methods were compared to results previously obtained with the standard technique of isotope ratio mass spectrometry (IRMS). Bias, SD, and bias ± 1.96SD bands between IRMS and OA-ICOS were computed. RESULTS: The between OA-ICOS and IRMS rCO(2) and TEE trends were equivalent (within 1.2% and 4.1%, respectively), in spite of the differences in measured δ(18)O values at high enrichment levels. The OA-ICOS δ(18)O values displayed an increasing offset from the IRMS results as the (18)O enrichment increased (mean ± SD 4.6–5.7‰ ± 2‰ offset at the time point with highest (18)O enrichment, ∼135‰), whereas the hydrogen isotope ratio (δ(2)H) differed only slightly between the methods (mean offset −4.9‰ for all time points). The between-diet differences in TEE from the previous study were recapitulated with a smaller subset of participants and time points. CONCLUSIONS: OA-ICOS analysis is an accurate and feasible technique for the DLW method. Given the δ(18)O offset observed at high enrichment, validation of each OA-ICOS instrumental setup against established methods (e.g., IRMS) is recommended.