Cargando…
Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay
OBJECTIVE: We assessed the clinical performance of novel Roche Elecsys SARS-CoV-2 Antigen fully automated electrochemiluminescence immunoassay (ECLIA). DESIGN AND METHODS: We tested 160 subjects, 110 (68.8%), with positive molecular test for SARS-CoV-2 infection in nasopharyngeal samples, with Alton...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8760093/ https://www.ncbi.nlm.nih.gov/pubmed/35071720 http://dx.doi.org/10.1016/j.plabm.2022.e00265 |
_version_ | 1784633245709631488 |
---|---|
author | Salvagno, Gian Luca Pighi, Laura De Nitto, Simone Lippi, Giuseppe |
author_facet | Salvagno, Gian Luca Pighi, Laura De Nitto, Simone Lippi, Giuseppe |
author_sort | Salvagno, Gian Luca |
collection | PubMed |
description | OBJECTIVE: We assessed the clinical performance of novel Roche Elecsys SARS-CoV-2 Antigen fully automated electrochemiluminescence immunoassay (ECLIA). DESIGN AND METHODS: We tested 160 subjects, 110 (68.8%), with positive molecular test for SARS-CoV-2 infection in nasopharyngeal samples, with Altona Diagnostics RealStar SARS-CoV-2 RT-PCR Kit and Roche Elecsys SARS-CoV-2 Antigen. RESULTS: Highly significant correlation was found between Elecsys SARS-CoV-2 Antigen ECLIA and cycle threshold (Ct) values of SARS-CoV-2 S and E genes (both r = −0.91; p < 0.001). The area under the curve (AUC), sensitivity and specificity of Elecsys SARS-CoV-2 Antigen ECLIA were 0.83, 0.43 and 1.00 in all samples, 0.99, 0.87 and 0.99 in those with both Ct values < 30, as well as 1.00, 1.00 and 0.89 in samples with both Ct values < 25. CONCLUSION: Roche Elecsys SARS-CoV-2 Antigen ECLIA may be a surrogate of molecular testing for identification of super-spreaders. |
format | Online Article Text |
id | pubmed-8760093 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-87600932022-01-18 Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay Salvagno, Gian Luca Pighi, Laura De Nitto, Simone Lippi, Giuseppe Pract Lab Med Article OBJECTIVE: We assessed the clinical performance of novel Roche Elecsys SARS-CoV-2 Antigen fully automated electrochemiluminescence immunoassay (ECLIA). DESIGN AND METHODS: We tested 160 subjects, 110 (68.8%), with positive molecular test for SARS-CoV-2 infection in nasopharyngeal samples, with Altona Diagnostics RealStar SARS-CoV-2 RT-PCR Kit and Roche Elecsys SARS-CoV-2 Antigen. RESULTS: Highly significant correlation was found between Elecsys SARS-CoV-2 Antigen ECLIA and cycle threshold (Ct) values of SARS-CoV-2 S and E genes (both r = −0.91; p < 0.001). The area under the curve (AUC), sensitivity and specificity of Elecsys SARS-CoV-2 Antigen ECLIA were 0.83, 0.43 and 1.00 in all samples, 0.99, 0.87 and 0.99 in those with both Ct values < 30, as well as 1.00, 1.00 and 0.89 in samples with both Ct values < 25. CONCLUSION: Roche Elecsys SARS-CoV-2 Antigen ECLIA may be a surrogate of molecular testing for identification of super-spreaders. Elsevier 2022-01-15 /pmc/articles/PMC8760093/ /pubmed/35071720 http://dx.doi.org/10.1016/j.plabm.2022.e00265 Text en © 2022 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Article Salvagno, Gian Luca Pighi, Laura De Nitto, Simone Lippi, Giuseppe Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay |
title | Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay |
title_full | Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay |
title_fullStr | Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay |
title_full_unstemmed | Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay |
title_short | Clinical performance of the Roche Elecsys SARS-CoV-2 antigen fully automated electrochemiluminescence immunoassay |
title_sort | clinical performance of the roche elecsys sars-cov-2 antigen fully automated electrochemiluminescence immunoassay |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8760093/ https://www.ncbi.nlm.nih.gov/pubmed/35071720 http://dx.doi.org/10.1016/j.plabm.2022.e00265 |
work_keys_str_mv | AT salvagnogianluca clinicalperformanceoftherocheelecsyssarscov2antigenfullyautomatedelectrochemiluminescenceimmunoassay AT pighilaura clinicalperformanceoftherocheelecsyssarscov2antigenfullyautomatedelectrochemiluminescenceimmunoassay AT denittosimone clinicalperformanceoftherocheelecsyssarscov2antigenfullyautomatedelectrochemiluminescenceimmunoassay AT lippigiuseppe clinicalperformanceoftherocheelecsyssarscov2antigenfullyautomatedelectrochemiluminescenceimmunoassay |