(19)F MRI Imaging Strategies to Reduce Isoflurane Artifacts in In Vivo Images
PURPOSE: Isoflurane (ISO) is the most commonly used preclinical inhalation anesthetic. This is a problem in (19)F MRI of fluorine contrast agents, as ISO signals cause artifacts that interfere with unambiguous image interpretation and quantification; the two most attractive properties of heteronucle...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8760223/ https://www.ncbi.nlm.nih.gov/pubmed/34671928 http://dx.doi.org/10.1007/s11307-021-01653-6 |
Sumario: | PURPOSE: Isoflurane (ISO) is the most commonly used preclinical inhalation anesthetic. This is a problem in (19)F MRI of fluorine contrast agents, as ISO signals cause artifacts that interfere with unambiguous image interpretation and quantification; the two most attractive properties of heteronuclear MRI. We aimed to avoid these artifacts using MRI strategies that can be applied by any pre-clinical researcher. PROCEDURES: Three strategies to avoid ISO chemical shift displacement artifacts (CSDA) in (19)F MRI are described and demonstrated with measurements of (19)F-containing agents in phantoms and in vivo (n = 3 for all strategies). The success of these strategies is compared to a standard Rapid Acquisition with Relaxation Enhancement (RARE) sequence, with phantom and in vivo validation. ISO artifacts can successfully be avoided by (1) shifting them outside the region of interest using a narrow signal acquisition bandwidth, (2) suppression of ISO by planning a frequency-selective suppression pulse before signal acquisition or by (3) preventing ISO excitation with a 3D sequence with a narrow excitation bandwidth. RESULTS: All three strategies result in complete ISO signal avoidance (p < 0.0001 for all methods). Using a narrow acquisition bandwidth can result in loss of signal to noise ratio and distortion of the image, and a frequency-selective suppression pulse can be incomplete when B(1)-inhomogeneities are present. Preventing ISO excitation with a narrow excitation pulse in a 3D sequence yields the most robust results (relative SNR 151 ± 28% compared to 2D multislice methods, p = 0.006). CONCLUSION: We optimized three easily implementable methods to avoid ISO signal artifacts and validated their performance in phantoms and in vivo. We make recommendation on the parameters that pre-clinical studies should report in their method section to make the used approach insightful. |
---|