Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients

BACKGROUND: Assessment of platelet function is important in the management of patients who are subject to operation as well as at potential risk of hemorrhagic complications. OBJECTIVE: This study aimed to evaluate a new platelet assays (Anysis-Epinephrine, Anysis-ADP) and to compare them with PFA-2...

Descripción completa

Detalles Bibliográficos
Autores principales: Piao, Jinxiang, Yoo, Chaeyoung, Kim, SunYoung, Whang, Youn-Wha, Choi, Cheol Ung, Shin, Sehyun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: IOS Press 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8764596/
https://www.ncbi.nlm.nih.gov/pubmed/34657880
http://dx.doi.org/10.3233/CH-211185
_version_ 1784634199750213632
author Piao, Jinxiang
Yoo, Chaeyoung
Kim, SunYoung
Whang, Youn-Wha
Choi, Cheol Ung
Shin, Sehyun
author_facet Piao, Jinxiang
Yoo, Chaeyoung
Kim, SunYoung
Whang, Youn-Wha
Choi, Cheol Ung
Shin, Sehyun
author_sort Piao, Jinxiang
collection PubMed
description BACKGROUND: Assessment of platelet function is important in the management of patients who are subject to operation as well as at potential risk of hemorrhagic complications. OBJECTIVE: This study aimed to evaluate a new platelet assays (Anysis-Epinephrine, Anysis-ADP) and to compare them with PFA-200 in cardiology visiting patients and inpatients. METHODS: Citrated blood samples were collected from 184 patients for ADP test and 163 patients for EPI test, who visited Korea University Guro Hospital with written consent. The PFA-200 assay gives a test result the closure time (CT) until the blood flow rate decreases to 10% of the initial value, whereas Anysis-200 assay does a blood flow migration distance (MD) until blood flow completely stops. According to the results of PFA closure time (CT), the tested samples were classified as either negative control or positive group. The measurements were simultaneously conducted with two devices and compared. RESULTS: The sensitivity and specificity of Anysis-200 C/EPI kit in comparison to PFA-200 C/EPI kit was 87.5% and 85.7%, respectively. Regarding C/ADP kit, the sensitivity and specificity of Anysis-200 was 96.9% and 87.5%, respectively. In addition, the sums of sensitivity and specificity are greater than 150% for both of EPI and ADP. Also, it was found that likelihood ratio and odd ratio for each assay provide useful additional information. Since the Cohen’s kappa coefficients value between the two devices was relatively high, the equivalence between the two devices was confirmed. CONCLUSIONS: Anysis-200, a novel platelet function analyzer has showed excellent agreements with PFA-200 with high agreement rates and precision. Anysis-200 assay would be useful in assessing bleeding risk management as well as abnormal platelet reactivity at point of care.
format Online
Article
Text
id pubmed-8764596
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher IOS Press
record_format MEDLINE/PubMed
spelling pubmed-87645962022-02-04 Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients Piao, Jinxiang Yoo, Chaeyoung Kim, SunYoung Whang, Youn-Wha Choi, Cheol Ung Shin, Sehyun Clin Hemorheol Microcirc Research Article BACKGROUND: Assessment of platelet function is important in the management of patients who are subject to operation as well as at potential risk of hemorrhagic complications. OBJECTIVE: This study aimed to evaluate a new platelet assays (Anysis-Epinephrine, Anysis-ADP) and to compare them with PFA-200 in cardiology visiting patients and inpatients. METHODS: Citrated blood samples were collected from 184 patients for ADP test and 163 patients for EPI test, who visited Korea University Guro Hospital with written consent. The PFA-200 assay gives a test result the closure time (CT) until the blood flow rate decreases to 10% of the initial value, whereas Anysis-200 assay does a blood flow migration distance (MD) until blood flow completely stops. According to the results of PFA closure time (CT), the tested samples were classified as either negative control or positive group. The measurements were simultaneously conducted with two devices and compared. RESULTS: The sensitivity and specificity of Anysis-200 C/EPI kit in comparison to PFA-200 C/EPI kit was 87.5% and 85.7%, respectively. Regarding C/ADP kit, the sensitivity and specificity of Anysis-200 was 96.9% and 87.5%, respectively. In addition, the sums of sensitivity and specificity are greater than 150% for both of EPI and ADP. Also, it was found that likelihood ratio and odd ratio for each assay provide useful additional information. Since the Cohen’s kappa coefficients value between the two devices was relatively high, the equivalence between the two devices was confirmed. CONCLUSIONS: Anysis-200, a novel platelet function analyzer has showed excellent agreements with PFA-200 with high agreement rates and precision. Anysis-200 assay would be useful in assessing bleeding risk management as well as abnormal platelet reactivity at point of care. IOS Press 2021-12-17 /pmc/articles/PMC8764596/ /pubmed/34657880 http://dx.doi.org/10.3233/CH-211185 Text en © 2021 – The authors. Published by IOS Press https://creativecommons.org/licenses/by-nc/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution Non-Commercial (CC BY-NC 4.0) License (https://creativecommons.org/licenses/by-nc/4.0/) , which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Piao, Jinxiang
Yoo, Chaeyoung
Kim, SunYoung
Whang, Youn-Wha
Choi, Cheol Ung
Shin, Sehyun
Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients
title Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients
title_full Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients
title_fullStr Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients
title_full_unstemmed Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients
title_short Performance comparison of the PFA-200 and Anysis-200: Assessment of bleeding risk screening in cardiology patients
title_sort performance comparison of the pfa-200 and anysis-200: assessment of bleeding risk screening in cardiology patients
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8764596/
https://www.ncbi.nlm.nih.gov/pubmed/34657880
http://dx.doi.org/10.3233/CH-211185
work_keys_str_mv AT piaojinxiang performancecomparisonofthepfa200andanysis200assessmentofbleedingriskscreeningincardiologypatients
AT yoochaeyoung performancecomparisonofthepfa200andanysis200assessmentofbleedingriskscreeningincardiologypatients
AT kimsunyoung performancecomparisonofthepfa200andanysis200assessmentofbleedingriskscreeningincardiologypatients
AT whangyounwha performancecomparisonofthepfa200andanysis200assessmentofbleedingriskscreeningincardiologypatients
AT choicheolung performancecomparisonofthepfa200andanysis200assessmentofbleedingriskscreeningincardiologypatients
AT shinsehyun performancecomparisonofthepfa200andanysis200assessmentofbleedingriskscreeningincardiologypatients