Cargando…

Pulmonary Artery Perforation by Atrial Appendage Closure Device and the Contribution of Auricular Contraction to Its Pathogenesis

We report the case of an 83-year-old man with paroxysmal atrial fibrillation who underwent successful percutaneous left atrial appendage closure with the LAmbre device, being in sinus rhythm at implantation. Suddenly, the patient experienced cardiac tamponade and died within a few minutes. Autopsy r...

Descripción completa

Detalles Bibliográficos
Autores principales: Bretones-Pino, Teresa, Rivera-López, Ricardo, Carrero-Castaño, Alejandro, Molina-Navarro, Eduardo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8767028/
https://www.ncbi.nlm.nih.gov/pubmed/35072032
http://dx.doi.org/10.1016/j.cjco.2021.07.021
Descripción
Sumario:We report the case of an 83-year-old man with paroxysmal atrial fibrillation who underwent successful percutaneous left atrial appendage closure with the LAmbre device, being in sinus rhythm at implantation. Suddenly, the patient experienced cardiac tamponade and died within a few minutes. Autopsy revealed a slight protrusion of the LAmbre device into the atrial appendage wall, and pulmonary artery laceration. This is the first published report of pulmonary artery perforation by the LAmbre device. This case highlights the need for a detailed imaging study before this procedure is performed, to assess left atrial appendage movement/contraction in patients in sinus rhythm.