Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion
A venipuncture is the most common non-invasive medical procedure, and is frequently used with patients; however, a high probability of post-injection complications accompanies intravenous injection. The most common complication is a hematoma, which is associated with puncture of the uppermost and lo...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8779690/ https://www.ncbi.nlm.nih.gov/pubmed/35062626 http://dx.doi.org/10.3390/s22020665 |
_version_ | 1784637639119339520 |
---|---|
author | Kudashov, Ivan Shchukin, Sergey Al-harosh, Mugeb Shcherbachev, Andrew |
author_facet | Kudashov, Ivan Shchukin, Sergey Al-harosh, Mugeb Shcherbachev, Andrew |
author_sort | Kudashov, Ivan |
collection | PubMed |
description | A venipuncture is the most common non-invasive medical procedure, and is frequently used with patients; however, a high probability of post-injection complications accompanies intravenous injection. The most common complication is a hematoma, which is associated with puncture of the uppermost and lowermost walls. To simplify and reduce complications of the venipuncture procedure, and as well as automation of this process, a device that can provide information of the needle tip position into patient’s tissues needs to be developed. This paper presents a peripheral vascular puncture control system based on electrical impedance measurements. A special electrode system was designed to achieve the maximum sensitivity for puncture identification using a traditional needle, which is usually used in clinical practice. An experimental study on subjects showed that the electrical impedance signal changed significantly once the standard needle entered the blood vessel. On basis of theoretical and experimental studies, a decision rule of puncture identification based on the analysis of amplitude-time parameters of experimental signals was proposed. The proposed method was tested on 15 test and 9 control samples, with the results showing that 97% accuracy was obtained. |
format | Online Article Text |
id | pubmed-8779690 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-87796902022-01-22 Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion Kudashov, Ivan Shchukin, Sergey Al-harosh, Mugeb Shcherbachev, Andrew Sensors (Basel) Article A venipuncture is the most common non-invasive medical procedure, and is frequently used with patients; however, a high probability of post-injection complications accompanies intravenous injection. The most common complication is a hematoma, which is associated with puncture of the uppermost and lowermost walls. To simplify and reduce complications of the venipuncture procedure, and as well as automation of this process, a device that can provide information of the needle tip position into patient’s tissues needs to be developed. This paper presents a peripheral vascular puncture control system based on electrical impedance measurements. A special electrode system was designed to achieve the maximum sensitivity for puncture identification using a traditional needle, which is usually used in clinical practice. An experimental study on subjects showed that the electrical impedance signal changed significantly once the standard needle entered the blood vessel. On basis of theoretical and experimental studies, a decision rule of puncture identification based on the analysis of amplitude-time parameters of experimental signals was proposed. The proposed method was tested on 15 test and 9 control samples, with the results showing that 97% accuracy was obtained. MDPI 2022-01-15 /pmc/articles/PMC8779690/ /pubmed/35062626 http://dx.doi.org/10.3390/s22020665 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Kudashov, Ivan Shchukin, Sergey Al-harosh, Mugeb Shcherbachev, Andrew Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion |
title | Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion |
title_full | Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion |
title_fullStr | Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion |
title_full_unstemmed | Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion |
title_short | Smart Bio-Impedance-Based Sensor for Guiding Standard Needle Insertion |
title_sort | smart bio-impedance-based sensor for guiding standard needle insertion |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8779690/ https://www.ncbi.nlm.nih.gov/pubmed/35062626 http://dx.doi.org/10.3390/s22020665 |
work_keys_str_mv | AT kudashovivan smartbioimpedancebasedsensorforguidingstandardneedleinsertion AT shchukinsergey smartbioimpedancebasedsensorforguidingstandardneedleinsertion AT alharoshmugeb smartbioimpedancebasedsensorforguidingstandardneedleinsertion AT shcherbachevandrew smartbioimpedancebasedsensorforguidingstandardneedleinsertion |