Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation

BACKGROUND: Cardiac index (CI; cardiac output indexed to body surface area) is routinely measured during kidney transplant surgery. Bioimpedance cardiometry is a transthoracic impedance as the non-invasive alternative for hemodynamic monitoring, using semi-invasive uncalibrated pulse wave or contour...

Descripción completa

Detalles Bibliográficos
Autores principales: Aditianingsih, Dita, Hidayat, Jefferson, Ginting, Vivi Medina
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Kowsar 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8782196/
https://www.ncbi.nlm.nih.gov/pubmed/35075410
http://dx.doi.org/10.5812/aapm.117918
_version_ 1784638261200683008
author Aditianingsih, Dita
Hidayat, Jefferson
Ginting, Vivi Medina
author_facet Aditianingsih, Dita
Hidayat, Jefferson
Ginting, Vivi Medina
author_sort Aditianingsih, Dita
collection PubMed
description BACKGROUND: Cardiac index (CI; cardiac output indexed to body surface area) is routinely measured during kidney transplant surgery. Bioimpedance cardiometry is a transthoracic impedance as the non-invasive alternative for hemodynamic monitoring, using semi-invasive uncalibrated pulse wave or contour (UPC) analysis. OBJECTIVES: We performed a cross-sectional observational study on 50 kidney transplant patients to compare the CI measurement agreement, concordance rate, and trending ability between bioimpedance and UPC analysis. METHODS: For each patient, CI was measured by bioimpedance analysis (ICON(TM)) and UPC analysis (EV1000(TM)) devices at three time points: after induction, during incision, and at reperfusion. The device measurement accuracy was assessed by the bias value, limit of agreement (LoA), and percentage error (PE) using Bland-Altman analyses. Trending ability was assessed by angular bias and polar concordance through four-quadrant and polar plot analyses. RESULTS: From each time point and pooled measurement, the correlation coefficients were 0.267, 0.327, 0.321, and 0.348. Bland-Altman analyses showed mean bias values of 1.18, 1.06, 1.48, and 1.30, LoA of -1.35 to 3.72, -1.39 to 3.51, -1.07 to 4.04, and -1.17 to 3.78, and PE of 82.21, 78.50, 68.74, and 74.58%, respectively. Polar plot analyses revealed angular bias values of -10.37º, -15.01º, -18.68º, and -12.62º, with radial LoA of 89.79º, 85.86º, 83.38º, and 87.82º, respectively. The four-quadrant plot concordance rates were 70.77, 67.35, 65.90, and 69.79%. These analyses showed poor agreement, weak concordance, and low trending ability of bioimpedance cardiometry to UPC analysis. CONCLUSIONS: Bioimpedance and UPC analysis for CI measurements were not interchangeable in patients undergoing kidney transplant surgery. Cardiac index monitoring using bioimpedance cardiometry during kidney transplantation should be interpreted cautiously because it showed poor reliability due to low accuracy, precision, and trending ability for CI measurement.
format Online
Article
Text
id pubmed-8782196
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Kowsar
record_format MEDLINE/PubMed
spelling pubmed-87821962022-01-23 Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation Aditianingsih, Dita Hidayat, Jefferson Ginting, Vivi Medina Anesth Pain Med Research Article BACKGROUND: Cardiac index (CI; cardiac output indexed to body surface area) is routinely measured during kidney transplant surgery. Bioimpedance cardiometry is a transthoracic impedance as the non-invasive alternative for hemodynamic monitoring, using semi-invasive uncalibrated pulse wave or contour (UPC) analysis. OBJECTIVES: We performed a cross-sectional observational study on 50 kidney transplant patients to compare the CI measurement agreement, concordance rate, and trending ability between bioimpedance and UPC analysis. METHODS: For each patient, CI was measured by bioimpedance analysis (ICON(TM)) and UPC analysis (EV1000(TM)) devices at three time points: after induction, during incision, and at reperfusion. The device measurement accuracy was assessed by the bias value, limit of agreement (LoA), and percentage error (PE) using Bland-Altman analyses. Trending ability was assessed by angular bias and polar concordance through four-quadrant and polar plot analyses. RESULTS: From each time point and pooled measurement, the correlation coefficients were 0.267, 0.327, 0.321, and 0.348. Bland-Altman analyses showed mean bias values of 1.18, 1.06, 1.48, and 1.30, LoA of -1.35 to 3.72, -1.39 to 3.51, -1.07 to 4.04, and -1.17 to 3.78, and PE of 82.21, 78.50, 68.74, and 74.58%, respectively. Polar plot analyses revealed angular bias values of -10.37º, -15.01º, -18.68º, and -12.62º, with radial LoA of 89.79º, 85.86º, 83.38º, and 87.82º, respectively. The four-quadrant plot concordance rates were 70.77, 67.35, 65.90, and 69.79%. These analyses showed poor agreement, weak concordance, and low trending ability of bioimpedance cardiometry to UPC analysis. CONCLUSIONS: Bioimpedance and UPC analysis for CI measurements were not interchangeable in patients undergoing kidney transplant surgery. Cardiac index monitoring using bioimpedance cardiometry during kidney transplantation should be interpreted cautiously because it showed poor reliability due to low accuracy, precision, and trending ability for CI measurement. Kowsar 2021-10-31 /pmc/articles/PMC8782196/ /pubmed/35075410 http://dx.doi.org/10.5812/aapm.117918 Text en Copyright © 2021, Author(s) https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits copy and redistribute the material just in noncommercial usages, provided the original work is properly cited.
spellingShingle Research Article
Aditianingsih, Dita
Hidayat, Jefferson
Ginting, Vivi Medina
Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation
title Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation
title_full Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation
title_fullStr Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation
title_full_unstemmed Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation
title_short Comparison of Bioimpedance Versus Pulse Contour Analysis for Intraoperative Cardiac Index Monitoring in Patients Undergoing Kidney Transplantation
title_sort comparison of bioimpedance versus pulse contour analysis for intraoperative cardiac index monitoring in patients undergoing kidney transplantation
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8782196/
https://www.ncbi.nlm.nih.gov/pubmed/35075410
http://dx.doi.org/10.5812/aapm.117918
work_keys_str_mv AT aditianingsihdita comparisonofbioimpedanceversuspulsecontouranalysisforintraoperativecardiacindexmonitoringinpatientsundergoingkidneytransplantation
AT hidayatjefferson comparisonofbioimpedanceversuspulsecontouranalysisforintraoperativecardiacindexmonitoringinpatientsundergoingkidneytransplantation
AT gintingvivimedina comparisonofbioimpedanceversuspulsecontouranalysisforintraoperativecardiacindexmonitoringinpatientsundergoingkidneytransplantation