A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion
A central venous catheter (CVC) should be inserted at the optimum position to infuse medicines, blood products, nutrients, or fluids. Positioning of the catheter tip is commonly performed under landmark or fluoroscopic guidance. However, Japanese regulations do not allow the performing of fluoroscop...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Fukushima Society of Medical Science
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8784190/ https://www.ncbi.nlm.nih.gov/pubmed/34937810 http://dx.doi.org/10.5387/fms.2021-27 |
_version_ | 1784638679776493568 |
---|---|
author | Tashiro, Masami Kubo, Hitoshi Kanezawa, Chie Ito, Hiroshi |
author_facet | Tashiro, Masami Kubo, Hitoshi Kanezawa, Chie Ito, Hiroshi |
author_sort | Tashiro, Masami |
collection | PubMed |
description | A central venous catheter (CVC) should be inserted at the optimum position to infuse medicines, blood products, nutrients, or fluids. Positioning of the catheter tip is commonly performed under landmark or fluoroscopic guidance. However, Japanese regulations do not allow the performing of fluoroscopy-guided procedures outside of the fluoroscopy room. We hypothesized that a new image-guided CVC placement technique by combining a wireless flat-panel detector (FPD) and a mobile X-ray system could be applied at the bedside to support CVC insertion. A CVC attached to a chest phantom in conjunction with the polymethyl methacrylate (PMMA) phantom was imaged, contrast-to-noise ratio (CNR) was measured with images, and radiologists and emergency physicians rated the catheter images using a Likert scale for visual evaluation. The minimum dose of the FPD and mobile X-ray system was reduced by at least 98% compared with that of the X-ray fluoroscopy system. The CNR decreased with the increasing PMMA phantom thickness. However, results of the visual evaluation were maintained at the clinically usable score with low-dose imaging up to a 6-cm thickness of the PMMA phantom. In conclusion, the combination of FPD and mobile X-ray systems is particularly effective in the emergency room setting where such procedures are required to be performed with urgency. |
format | Online Article Text |
id | pubmed-8784190 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | The Fukushima Society of Medical Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-87841902022-02-02 A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion Tashiro, Masami Kubo, Hitoshi Kanezawa, Chie Ito, Hiroshi Fukushima J Med Sci Original Article A central venous catheter (CVC) should be inserted at the optimum position to infuse medicines, blood products, nutrients, or fluids. Positioning of the catheter tip is commonly performed under landmark or fluoroscopic guidance. However, Japanese regulations do not allow the performing of fluoroscopy-guided procedures outside of the fluoroscopy room. We hypothesized that a new image-guided CVC placement technique by combining a wireless flat-panel detector (FPD) and a mobile X-ray system could be applied at the bedside to support CVC insertion. A CVC attached to a chest phantom in conjunction with the polymethyl methacrylate (PMMA) phantom was imaged, contrast-to-noise ratio (CNR) was measured with images, and radiologists and emergency physicians rated the catheter images using a Likert scale for visual evaluation. The minimum dose of the FPD and mobile X-ray system was reduced by at least 98% compared with that of the X-ray fluoroscopy system. The CNR decreased with the increasing PMMA phantom thickness. However, results of the visual evaluation were maintained at the clinically usable score with low-dose imaging up to a 6-cm thickness of the PMMA phantom. In conclusion, the combination of FPD and mobile X-ray systems is particularly effective in the emergency room setting where such procedures are required to be performed with urgency. The Fukushima Society of Medical Science 2021-12-21 2021 /pmc/articles/PMC8784190/ /pubmed/34937810 http://dx.doi.org/10.5387/fms.2021-27 Text en © 2021 The Fukushima Society of Medical Science https://creativecommons.org/licenses/by-nc-sa/4.0/This article is licensed under a Creative Commons [Attribution-NonCommercial-ShareAlike 4.0 International] license. https://creativecommons.org/licenses/by-nc-sa/4.0/ |
spellingShingle | Original Article Tashiro, Masami Kubo, Hitoshi Kanezawa, Chie Ito, Hiroshi A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion |
title | A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion |
title_full | A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion |
title_fullStr | A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion |
title_full_unstemmed | A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion |
title_short | A proposed combination of flat-panel detector and mobile X-ray systems for low-dose image-guided central venous catheter insertion |
title_sort | proposed combination of flat-panel detector and mobile x-ray systems for low-dose image-guided central venous catheter insertion |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8784190/ https://www.ncbi.nlm.nih.gov/pubmed/34937810 http://dx.doi.org/10.5387/fms.2021-27 |
work_keys_str_mv | AT tashiromasami aproposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion AT kubohitoshi aproposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion AT kanezawachie aproposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion AT itohiroshi aproposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion AT tashiromasami proposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion AT kubohitoshi proposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion AT kanezawachie proposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion AT itohiroshi proposedcombinationofflatpaneldetectorandmobilexraysystemsforlowdoseimageguidedcentralvenouscatheterinsertion |