Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy

BACKGROUND: Balloon guide catheters (BGCs) achieve proximal flow control during thrombectomy but antegrade intracranial flow often persists via the Circle of Willis. Closely sizing an aspiration catheter to the target vessel might achieve greater flow control and improve technical performance. Our o...

Descripción completa

Detalles Bibliográficos
Autores principales: Nogueira, Raul G, Ryan, David, Mullins, Liam, Thornton, John, Fitzgerald, Seán
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8785049/
https://www.ncbi.nlm.nih.gov/pubmed/33722968
http://dx.doi.org/10.1136/neurintsurg-2021-017316
_version_ 1784638878499471360
author Nogueira, Raul G
Ryan, David
Mullins, Liam
Thornton, John
Fitzgerald, Seán
author_facet Nogueira, Raul G
Ryan, David
Mullins, Liam
Thornton, John
Fitzgerald, Seán
author_sort Nogueira, Raul G
collection PubMed
description BACKGROUND: Balloon guide catheters (BGCs) achieve proximal flow control during thrombectomy but antegrade intracranial flow often persists via the Circle of Willis. Closely sizing an aspiration catheter to the target vessel might achieve greater flow control and improve technical performance. Our objective was to measure the impact of aspiration catheter size on distal flow control and flow reversal with and without the use of BGCs. Clot retrieval testing was performed to establish the impact of these parameters on revascularization. METHODS: An in vitro thrombectomy model replicated in vivo conditions. Flow was measured continuously using ultrasonic flow sensors placed 20 cm distal to the catheter tip in the middlel cerebral artery (MCA). Four aspiration catheters of increasing size were evaluated: ACE 60 and 64 (Penumbra), SOFIA Plus (MicroVention), and Millipede 088 (Perfuze). Two clot analog types (red blood cell-rich and fibrin/platelet-rich) were used for clot retrieval testing. RESULTS: The larger area of the ‘superbore’ Millipede 088 catheter resulted in a larger reduction in antegrade flow than standard aspiration catheters, even when the latter were combined with a BGC. During aspiration, 6Fr catheters were unable to cause flow reversal in the distal MCA while the Millipede 088 achieved significant distal flow reversal (−146 mL/min) (P<0.0001*) (*denotes significance). The solo use of Millipede 088 resulted in better recanalization outcomes and significantly reduced distal emboli for internal carotid artery (P=0.015*) and MCA (P=0.014*) occlusions compared with all other devices and combinations. CONCLUSIONS: Maximizing the catheter-to-vessel size facilitates near flow-arrest on catheter insertion, potentially negating the need for a BGC. A 0.088 inch aspiration catheter enables significant flow reversal in the distal MCA during aspiration.
format Online
Article
Text
id pubmed-8785049
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-87850492022-02-04 Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy Nogueira, Raul G Ryan, David Mullins, Liam Thornton, John Fitzgerald, Seán J Neurointerv Surg Basic Science BACKGROUND: Balloon guide catheters (BGCs) achieve proximal flow control during thrombectomy but antegrade intracranial flow often persists via the Circle of Willis. Closely sizing an aspiration catheter to the target vessel might achieve greater flow control and improve technical performance. Our objective was to measure the impact of aspiration catheter size on distal flow control and flow reversal with and without the use of BGCs. Clot retrieval testing was performed to establish the impact of these parameters on revascularization. METHODS: An in vitro thrombectomy model replicated in vivo conditions. Flow was measured continuously using ultrasonic flow sensors placed 20 cm distal to the catheter tip in the middlel cerebral artery (MCA). Four aspiration catheters of increasing size were evaluated: ACE 60 and 64 (Penumbra), SOFIA Plus (MicroVention), and Millipede 088 (Perfuze). Two clot analog types (red blood cell-rich and fibrin/platelet-rich) were used for clot retrieval testing. RESULTS: The larger area of the ‘superbore’ Millipede 088 catheter resulted in a larger reduction in antegrade flow than standard aspiration catheters, even when the latter were combined with a BGC. During aspiration, 6Fr catheters were unable to cause flow reversal in the distal MCA while the Millipede 088 achieved significant distal flow reversal (−146 mL/min) (P<0.0001*) (*denotes significance). The solo use of Millipede 088 resulted in better recanalization outcomes and significantly reduced distal emboli for internal carotid artery (P=0.015*) and MCA (P=0.014*) occlusions compared with all other devices and combinations. CONCLUSIONS: Maximizing the catheter-to-vessel size facilitates near flow-arrest on catheter insertion, potentially negating the need for a BGC. A 0.088 inch aspiration catheter enables significant flow reversal in the distal MCA during aspiration. BMJ Publishing Group 2022-02 2021-03-15 /pmc/articles/PMC8785049/ /pubmed/33722968 http://dx.doi.org/10.1136/neurintsurg-2021-017316 Text en © Author(s) (or their employer(s)) 2022. Re-use permitted under CC BY. Published by BMJ. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed in accordance with the Creative Commons Attribution 4.0 Unported (CC BY 4.0) license, which permits others to copy, redistribute, remix, transform and build upon this work for any purpose, provided the original work is properly cited, a link to the licence is given, and indication of whether changes were made. See: https://creativecommons.org/licenses/by/4.0/.
spellingShingle Basic Science
Nogueira, Raul G
Ryan, David
Mullins, Liam
Thornton, John
Fitzgerald, Seán
Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy
title Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy
title_full Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy
title_fullStr Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy
title_full_unstemmed Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy
title_short Maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy
title_sort maximizing the catheter-to-vessel size optimizes distal flow control resulting in improved revascularization in vitro for aspiration thrombectomy
topic Basic Science
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8785049/
https://www.ncbi.nlm.nih.gov/pubmed/33722968
http://dx.doi.org/10.1136/neurintsurg-2021-017316
work_keys_str_mv AT nogueiraraulg maximizingthecathetertovesselsizeoptimizesdistalflowcontrolresultinginimprovedrevascularizationinvitroforaspirationthrombectomy
AT ryandavid maximizingthecathetertovesselsizeoptimizesdistalflowcontrolresultinginimprovedrevascularizationinvitroforaspirationthrombectomy
AT mullinsliam maximizingthecathetertovesselsizeoptimizesdistalflowcontrolresultinginimprovedrevascularizationinvitroforaspirationthrombectomy
AT thorntonjohn maximizingthecathetertovesselsizeoptimizesdistalflowcontrolresultinginimprovedrevascularizationinvitroforaspirationthrombectomy
AT fitzgeraldsean maximizingthecathetertovesselsizeoptimizesdistalflowcontrolresultinginimprovedrevascularizationinvitroforaspirationthrombectomy