Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial

BACKGROUND AND OBJECTIVES: EUS is widely used in the clinical practice. This study aimed to evaluate the efficacy of a novel echoendoscope regarding image quality, maneuverability, stability of the entire machine system, and safety. SETTING AND DESIGN: We conducted this open-label, multicenter, rand...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Sheng, Guo, Jintao, Liu, Xiang, Ge, Nan, Wang, Guoxin, Hu, Jinlong, Zhang, Kai, Sun, Siyu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8785674/
https://www.ncbi.nlm.nih.gov/pubmed/34975041
http://dx.doi.org/10.4103/EUS-D-21-00100
_version_ 1784639013066375168
author Wang, Sheng
Guo, Jintao
Liu, Xiang
Ge, Nan
Wang, Guoxin
Hu, Jinlong
Zhang, Kai
Sun, Siyu
author_facet Wang, Sheng
Guo, Jintao
Liu, Xiang
Ge, Nan
Wang, Guoxin
Hu, Jinlong
Zhang, Kai
Sun, Siyu
author_sort Wang, Sheng
collection PubMed
description BACKGROUND AND OBJECTIVES: EUS is widely used in the clinical practice. This study aimed to evaluate the efficacy of a novel echoendoscope regarding image quality, maneuverability, stability of the entire machine system, and safety. SETTING AND DESIGN: We conducted this open-label, multicenter, randomized, parallel-group, noninferiority clinical trial in three tertiary hospitals between November 2018 and April 2019. SUBJECTS AND METHODS: One hundred and thirty patients were included. The stratified segment randomization method was employed. Sixty-five patients in the test group received an EUS examination using the new material radial echoendoscope, and 65 patients in the control group received an EUS examination using the existing endoscope. We recorded the image quality, maneuverability, stability of the entire machine system, and safety. For the main outcome, comprehensive image quality, the groups were compared with the noninferiority test, using the confidence interval method and 10% as the noninferiority threshold. The Pearson Chi-square test was used to compare the incidence of adverse events between the groups. RESULTS: Sixty-five patients in the test group and 63 patients in the control group were analyzed. Two patients in control group did not complete an EUS examination due to machine problems. There were no significant differences in image quality, maneuverability of the echoendoscope, stability of the entire machine system, and safety between the groups. CONCLUSIONS: This new material radial echoendoscope showed good capabilities for image quality, maneuverability, stability of the entire machine system, and safety. Our data suggest that the new echoendoscope may provide an additional option for endoscopic physicians.
format Online
Article
Text
id pubmed-8785674
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-87856742022-02-03 Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial Wang, Sheng Guo, Jintao Liu, Xiang Ge, Nan Wang, Guoxin Hu, Jinlong Zhang, Kai Sun, Siyu Endosc Ultrasound Original Article BACKGROUND AND OBJECTIVES: EUS is widely used in the clinical practice. This study aimed to evaluate the efficacy of a novel echoendoscope regarding image quality, maneuverability, stability of the entire machine system, and safety. SETTING AND DESIGN: We conducted this open-label, multicenter, randomized, parallel-group, noninferiority clinical trial in three tertiary hospitals between November 2018 and April 2019. SUBJECTS AND METHODS: One hundred and thirty patients were included. The stratified segment randomization method was employed. Sixty-five patients in the test group received an EUS examination using the new material radial echoendoscope, and 65 patients in the control group received an EUS examination using the existing endoscope. We recorded the image quality, maneuverability, stability of the entire machine system, and safety. For the main outcome, comprehensive image quality, the groups were compared with the noninferiority test, using the confidence interval method and 10% as the noninferiority threshold. The Pearson Chi-square test was used to compare the incidence of adverse events between the groups. RESULTS: Sixty-five patients in the test group and 63 patients in the control group were analyzed. Two patients in control group did not complete an EUS examination due to machine problems. There were no significant differences in image quality, maneuverability of the echoendoscope, stability of the entire machine system, and safety between the groups. CONCLUSIONS: This new material radial echoendoscope showed good capabilities for image quality, maneuverability, stability of the entire machine system, and safety. Our data suggest that the new echoendoscope may provide an additional option for endoscopic physicians. Wolters Kluwer - Medknow 2021-12-27 /pmc/articles/PMC8785674/ /pubmed/34975041 http://dx.doi.org/10.4103/EUS-D-21-00100 Text en Copyright: © 2021 SPRING MEDIA PUBLISHING CO. LTD https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Wang, Sheng
Guo, Jintao
Liu, Xiang
Ge, Nan
Wang, Guoxin
Hu, Jinlong
Zhang, Kai
Sun, Siyu
Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial
title Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial
title_full Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial
title_fullStr Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial
title_full_unstemmed Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial
title_short Evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: An open-label, multicenter, randomized, parallel-group, noninferiority clinical trial
title_sort evaluation of a novel radial echoendosonoscope with a piezoelectric-composite transducer: an open-label, multicenter, randomized, parallel-group, noninferiority clinical trial
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8785674/
https://www.ncbi.nlm.nih.gov/pubmed/34975041
http://dx.doi.org/10.4103/EUS-D-21-00100
work_keys_str_mv AT wangsheng evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial
AT guojintao evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial
AT liuxiang evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial
AT genan evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial
AT wangguoxin evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial
AT hujinlong evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial
AT zhangkai evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial
AT sunsiyu evaluationofanovelradialechoendosonoscopewithapiezoelectriccompositetransduceranopenlabelmulticenterrandomizedparallelgroupnoninferiorityclinicaltrial