Lateral Resolution of a Commercial Optical Coherence Tomography Instrument

PURPOSE: The lateral resolution of an optical coherence tomography (OCT) instrument was considered to be equal to the illumination spot size on the retina. To evaluate the potential lateral resolution of the Spectralis OCT, an instrument calculated to have a 14 µm resolution. METHODS: The lateral po...

Descripción completa

Detalles Bibliográficos
Autores principales: Spaide, Richard F., Otto, Tilman, Caujolle, Sophie, Kübler, Johannes, Aumann, Silke, Fischer, Joerg, Reisman, Charles, Spahr, Hendrik, Lessmann, Annette
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8787587/
https://www.ncbi.nlm.nih.gov/pubmed/35044444
http://dx.doi.org/10.1167/tvst.11.1.28
_version_ 1784639390406934528
author Spaide, Richard F.
Otto, Tilman
Caujolle, Sophie
Kübler, Johannes
Aumann, Silke
Fischer, Joerg
Reisman, Charles
Spahr, Hendrik
Lessmann, Annette
author_facet Spaide, Richard F.
Otto, Tilman
Caujolle, Sophie
Kübler, Johannes
Aumann, Silke
Fischer, Joerg
Reisman, Charles
Spahr, Hendrik
Lessmann, Annette
author_sort Spaide, Richard F.
collection PubMed
description PURPOSE: The lateral resolution of an optical coherence tomography (OCT) instrument was considered to be equal to the illumination spot size on the retina. To evaluate the potential lateral resolution of the Spectralis OCT, an instrument calculated to have a 14 µm resolution. METHODS: The lateral point spread function (PSF) was evaluated using diamond abrasive powder 0 to 1 µm in diameter in silicone elastomer and a validated target with 800 nm FeO particles in urethane. The amplitude transfer function was calculated from human OCT images. Finally, resolution was measured using the 1951 USAF target. RESULTS: Measurement of the lateral PSF from 1215 diamond particle images yielded a full-width half maximum (FWHM) to be 5.11 µm and for 732 FeO particles, 4.9 µm. From the amplitude transfer function, the FWHM of the diffraction limited PSF was calculated to be 5.0 µm. The USAF target imaging showed a lateral resolution of 4.6 µm. CONCLUSIONS: Although a calculation of the spot size of the illumination beam was reported in the past as the lateral resolution of the OCT instrument, the actual lateral resolution is better by a factor of at least 2.5 times. The clinically used A-scan spacing was derived from the calculated, and not the true resolution, and results in under sampling. This set of findings likely apply to all commercial clinical instruments. TRANSLATIONAL RELEVANCE: The scan density parameters of past and present commercial OCT instruments were based on earlier translational concepts, which now appear to have been incorrect.
format Online
Article
Text
id pubmed-8787587
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-87875872022-01-26 Lateral Resolution of a Commercial Optical Coherence Tomography Instrument Spaide, Richard F. Otto, Tilman Caujolle, Sophie Kübler, Johannes Aumann, Silke Fischer, Joerg Reisman, Charles Spahr, Hendrik Lessmann, Annette Transl Vis Sci Technol Article PURPOSE: The lateral resolution of an optical coherence tomography (OCT) instrument was considered to be equal to the illumination spot size on the retina. To evaluate the potential lateral resolution of the Spectralis OCT, an instrument calculated to have a 14 µm resolution. METHODS: The lateral point spread function (PSF) was evaluated using diamond abrasive powder 0 to 1 µm in diameter in silicone elastomer and a validated target with 800 nm FeO particles in urethane. The amplitude transfer function was calculated from human OCT images. Finally, resolution was measured using the 1951 USAF target. RESULTS: Measurement of the lateral PSF from 1215 diamond particle images yielded a full-width half maximum (FWHM) to be 5.11 µm and for 732 FeO particles, 4.9 µm. From the amplitude transfer function, the FWHM of the diffraction limited PSF was calculated to be 5.0 µm. The USAF target imaging showed a lateral resolution of 4.6 µm. CONCLUSIONS: Although a calculation of the spot size of the illumination beam was reported in the past as the lateral resolution of the OCT instrument, the actual lateral resolution is better by a factor of at least 2.5 times. The clinically used A-scan spacing was derived from the calculated, and not the true resolution, and results in under sampling. This set of findings likely apply to all commercial clinical instruments. TRANSLATIONAL RELEVANCE: The scan density parameters of past and present commercial OCT instruments were based on earlier translational concepts, which now appear to have been incorrect. The Association for Research in Vision and Ophthalmology 2022-01-19 /pmc/articles/PMC8787587/ /pubmed/35044444 http://dx.doi.org/10.1167/tvst.11.1.28 Text en Copyright 2022 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
spellingShingle Article
Spaide, Richard F.
Otto, Tilman
Caujolle, Sophie
Kübler, Johannes
Aumann, Silke
Fischer, Joerg
Reisman, Charles
Spahr, Hendrik
Lessmann, Annette
Lateral Resolution of a Commercial Optical Coherence Tomography Instrument
title Lateral Resolution of a Commercial Optical Coherence Tomography Instrument
title_full Lateral Resolution of a Commercial Optical Coherence Tomography Instrument
title_fullStr Lateral Resolution of a Commercial Optical Coherence Tomography Instrument
title_full_unstemmed Lateral Resolution of a Commercial Optical Coherence Tomography Instrument
title_short Lateral Resolution of a Commercial Optical Coherence Tomography Instrument
title_sort lateral resolution of a commercial optical coherence tomography instrument
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8787587/
https://www.ncbi.nlm.nih.gov/pubmed/35044444
http://dx.doi.org/10.1167/tvst.11.1.28
work_keys_str_mv AT spaiderichardf lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT ottotilman lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT caujollesophie lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT kublerjohannes lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT aumannsilke lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT fischerjoerg lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT reismancharles lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT spahrhendrik lateralresolutionofacommercialopticalcoherencetomographyinstrument
AT lessmannannette lateralresolutionofacommercialopticalcoherencetomographyinstrument