Cargando…
Prediction Error Stabilization and Long-Term Standard Results with a Monofocal Intraocular Lens
The aim of this study was to assess the stability and differences between objective (O-Rx) and subjective (S-Rx) refraction for the assessment of the prediction error (PE). A secondary aim was to report the results of a monofocal intraocular lens (IOL). 100 subjects were included for whom S-Rx and O...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8788560/ https://www.ncbi.nlm.nih.gov/pubmed/35076638 http://dx.doi.org/10.3390/vision6010005 |
Sumario: | The aim of this study was to assess the stability and differences between objective (O-Rx) and subjective (S-Rx) refraction for the assessment of the prediction error (PE). A secondary aim was to report the results of a monofocal intraocular lens (IOL). 100 subjects were included for whom S-Rx and O-Rx were obtained for all visits, and for visual performance, posterior capsular opacification incidence and Nd:YAG rates at 12 months. Either S-Rx and O-Rx showed a hyperopic shift from 1 to 6 months (p < 0.05) and stabilization after 6 months. S-Rx was related with the axial length (rho = −0.29, p = 0.007), obtaining a major tendency towards hyperopia in short eyes implanted with high-power IOLs. O-Rx showed a myopic shift in comparison to S-Rx (p < 0.05). This resulted in a decrease of the number of eyes in ±0.50 D and ±1.00 D from 79 to 67% and from 94 to 90%, respectively. The median (interquartile range) uncorrected and corrected visual acuities were 0.1 (0.29) and 0 (0.12) logMAR, respectively, and seven eyes required Nd:YAG capsulotomy at 12 months. Some caution should be taken in PE studies in which O-Rx is used or S-Rx is measured in a 1-month follow-up. Constant optimization should be conducted for this IOL after S-Rx stabilization. |
---|