Cargando…
Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study
BACKGROUND: Combined medial and lateral plate fixation is recommended for complex tibial plateau fractures with medial fragments or no cortical bone contact. Although such fixation is adequate to resist forces during range of motion, it may be insufficient to support immediate postoperative weightbe...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8790864/ https://www.ncbi.nlm.nih.gov/pubmed/35078451 http://dx.doi.org/10.1186/s12891-022-05024-2 |
_version_ | 1784640109224656896 |
---|---|
author | Thamyongkit, Sorawut Abbasi, Pooyan Parks, Brent G. Shafiq, Babar Hasenboehler, Erik A. |
author_facet | Thamyongkit, Sorawut Abbasi, Pooyan Parks, Brent G. Shafiq, Babar Hasenboehler, Erik A. |
author_sort | Thamyongkit, Sorawut |
collection | PubMed |
description | BACKGROUND: Combined medial and lateral plate fixation is recommended for complex tibial plateau fractures with medial fragments or no cortical bone contact. Although such fixation is adequate to resist forces during range of motion, it may be insufficient to support immediate postoperative weightbearing. Here, we analyzed displacement, stiffness, and fixation failure during simulated full weightbearing of bicondylar tibial plateau fractures treated with combined medial and lateral locking plate fixation. METHODS: We used 10 fresh-frozen adult human cadaveric tibias and mated femurs. Osteotomies were performed with an oscillating saw and cutting template to simulate an AO Foundation and Orthopaedic Trauma Association (AO/OTA) 41-C2 fracture (simple articular, multifragmentary metaphyseal fracture). Specimens were anatomically reduced and stabilized with combined medial and lateral locking plates (AxSOS, Stryker, Mahwah, NJ). Specimens were loaded axially to simulate 4 weeks of walking in a person weighing 70 kg. The specimens were cyclically loaded from 200 N to a maximum of 2800 N. Then, if no failure, loading continued for 200,000 cycles. We measured displacement of each bone fragment and defined fixation failure as ≥5 mm of displacement. Construct stiffness and load at failure were calculated. Categorical and continuous data were analyzed using Chi-squared and unpaired t-tests, respectively. RESULTS: Mean total displacement values after 10,000 loading cycles were as follows: lateral, 0.4 ± 0.8 mm; proximal medial, 0.3 ± 0.7 mm; distal medial, 0.3 ± 0.6 mm; and central 0.4 ± 0.5 mm. Mean stiffness of the construct was 562 ± 164 N/mm. Fixation failure occurred in 6 of 10 specimens that reached 5 mm of plastic deformation before test completion. In the failure group, the mean load at failure was 2467 ± 532 N, and the mean number of cycles before failure was 53,155. After test completion, the greatest displacement was found at the distal medial fracture site (2.3 ± 1.4 mm) and lateral fracture site (2.2 ± 1.7 mm). CONCLUSIONS: Although combined medial and lateral plate fixation of complex tibial plateau fractures provides adequate stability to allow early range of motion, immediate full weightbearing is not recommended. |
format | Online Article Text |
id | pubmed-8790864 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-87908642022-01-26 Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study Thamyongkit, Sorawut Abbasi, Pooyan Parks, Brent G. Shafiq, Babar Hasenboehler, Erik A. BMC Musculoskelet Disord Research BACKGROUND: Combined medial and lateral plate fixation is recommended for complex tibial plateau fractures with medial fragments or no cortical bone contact. Although such fixation is adequate to resist forces during range of motion, it may be insufficient to support immediate postoperative weightbearing. Here, we analyzed displacement, stiffness, and fixation failure during simulated full weightbearing of bicondylar tibial plateau fractures treated with combined medial and lateral locking plate fixation. METHODS: We used 10 fresh-frozen adult human cadaveric tibias and mated femurs. Osteotomies were performed with an oscillating saw and cutting template to simulate an AO Foundation and Orthopaedic Trauma Association (AO/OTA) 41-C2 fracture (simple articular, multifragmentary metaphyseal fracture). Specimens were anatomically reduced and stabilized with combined medial and lateral locking plates (AxSOS, Stryker, Mahwah, NJ). Specimens were loaded axially to simulate 4 weeks of walking in a person weighing 70 kg. The specimens were cyclically loaded from 200 N to a maximum of 2800 N. Then, if no failure, loading continued for 200,000 cycles. We measured displacement of each bone fragment and defined fixation failure as ≥5 mm of displacement. Construct stiffness and load at failure were calculated. Categorical and continuous data were analyzed using Chi-squared and unpaired t-tests, respectively. RESULTS: Mean total displacement values after 10,000 loading cycles were as follows: lateral, 0.4 ± 0.8 mm; proximal medial, 0.3 ± 0.7 mm; distal medial, 0.3 ± 0.6 mm; and central 0.4 ± 0.5 mm. Mean stiffness of the construct was 562 ± 164 N/mm. Fixation failure occurred in 6 of 10 specimens that reached 5 mm of plastic deformation before test completion. In the failure group, the mean load at failure was 2467 ± 532 N, and the mean number of cycles before failure was 53,155. After test completion, the greatest displacement was found at the distal medial fracture site (2.3 ± 1.4 mm) and lateral fracture site (2.2 ± 1.7 mm). CONCLUSIONS: Although combined medial and lateral plate fixation of complex tibial plateau fractures provides adequate stability to allow early range of motion, immediate full weightbearing is not recommended. BioMed Central 2022-01-25 /pmc/articles/PMC8790864/ /pubmed/35078451 http://dx.doi.org/10.1186/s12891-022-05024-2 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Thamyongkit, Sorawut Abbasi, Pooyan Parks, Brent G. Shafiq, Babar Hasenboehler, Erik A. Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study |
title | Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study |
title_full | Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study |
title_fullStr | Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study |
title_full_unstemmed | Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study |
title_short | Weightbearing after combined medial and lateral plate fixation of AO/OTA 41-C2 bicondylar tibial plateau fractures: a biomechanical study |
title_sort | weightbearing after combined medial and lateral plate fixation of ao/ota 41-c2 bicondylar tibial plateau fractures: a biomechanical study |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8790864/ https://www.ncbi.nlm.nih.gov/pubmed/35078451 http://dx.doi.org/10.1186/s12891-022-05024-2 |
work_keys_str_mv | AT thamyongkitsorawut weightbearingaftercombinedmedialandlateralplatefixationofaoota41c2bicondylartibialplateaufracturesabiomechanicalstudy AT abbasipooyan weightbearingaftercombinedmedialandlateralplatefixationofaoota41c2bicondylartibialplateaufracturesabiomechanicalstudy AT parksbrentg weightbearingaftercombinedmedialandlateralplatefixationofaoota41c2bicondylartibialplateaufracturesabiomechanicalstudy AT shafiqbabar weightbearingaftercombinedmedialandlateralplatefixationofaoota41c2bicondylartibialplateaufracturesabiomechanicalstudy AT hasenboehlererika weightbearingaftercombinedmedialandlateralplatefixationofaoota41c2bicondylartibialplateaufracturesabiomechanicalstudy |