Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal

We examined the performance of human Schlemm’s canal (SC) imaging using different OCT devices: CIRRUS 5000 (840 nm, spectral-domain (SD)-OCT), PLEX Elite 9000 (1060 nm, swept-source (SS)-OCT) and CASIA SS-1000 (1310 nm, SS-OCT), and analyzed potential impact factors on visualization and the quantita...

Descripción completa

Detalles Bibliográficos
Autores principales: Wu, Xuan, Tan, Bingyao, Gan, Jinyuan, Lam, Adeline R., Chen, Yibing, Liu, Xinyu, Chua, Jacqueline, Wong, Damon W. K., Ang, Marcus, Schmetterer, Leopold, Yao, Xinwen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8792050/
https://www.ncbi.nlm.nih.gov/pubmed/35082345
http://dx.doi.org/10.1038/s41598-022-05410-9
_version_ 1784640323559882752
author Wu, Xuan
Tan, Bingyao
Gan, Jinyuan
Lam, Adeline R.
Chen, Yibing
Liu, Xinyu
Chua, Jacqueline
Wong, Damon W. K.
Ang, Marcus
Schmetterer, Leopold
Yao, Xinwen
author_facet Wu, Xuan
Tan, Bingyao
Gan, Jinyuan
Lam, Adeline R.
Chen, Yibing
Liu, Xinyu
Chua, Jacqueline
Wong, Damon W. K.
Ang, Marcus
Schmetterer, Leopold
Yao, Xinwen
author_sort Wu, Xuan
collection PubMed
description We examined the performance of human Schlemm’s canal (SC) imaging using different OCT devices: CIRRUS 5000 (840 nm, spectral-domain (SD)-OCT), PLEX Elite 9000 (1060 nm, swept-source (SS)-OCT) and CASIA SS-1000 (1310 nm, SS-OCT), and analyzed potential impact factors on visualization and the quantitative assessment of SC morphology in a pilot study. Ten healthy subjects were imaged using three OCT devices by a single experienced operator on the same day. Each eye underwent two cubic scans by each device, one on nasal and the other on temporal quadrant. The B-scan showing the largest SC was manually selected for processing. Four quantitative metrics, including one morphological metric as cross-sectional area (CSA), and three performance metrics as contrast, continuity, and coverage, were derived from the datasets. Repeated-measures ANOVA was used to investigate the difference between these parameters from the three devices (P < 0.05). We found the CSA measured from CIRRUS was significantly larger than PLEX, followed by CASIA. The contrast was highest in CIRRUS, followed by PLEX and CASIA. The coverage was also higher in CIRRUS as compared to PLEX and CASIA. No significant difference was seen in the continuity from the three devices. In summary, we showed the measurements from the three devices were not interchangeable.
format Online
Article
Text
id pubmed-8792050
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-87920502022-01-28 Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal Wu, Xuan Tan, Bingyao Gan, Jinyuan Lam, Adeline R. Chen, Yibing Liu, Xinyu Chua, Jacqueline Wong, Damon W. K. Ang, Marcus Schmetterer, Leopold Yao, Xinwen Sci Rep Article We examined the performance of human Schlemm’s canal (SC) imaging using different OCT devices: CIRRUS 5000 (840 nm, spectral-domain (SD)-OCT), PLEX Elite 9000 (1060 nm, swept-source (SS)-OCT) and CASIA SS-1000 (1310 nm, SS-OCT), and analyzed potential impact factors on visualization and the quantitative assessment of SC morphology in a pilot study. Ten healthy subjects were imaged using three OCT devices by a single experienced operator on the same day. Each eye underwent two cubic scans by each device, one on nasal and the other on temporal quadrant. The B-scan showing the largest SC was manually selected for processing. Four quantitative metrics, including one morphological metric as cross-sectional area (CSA), and three performance metrics as contrast, continuity, and coverage, were derived from the datasets. Repeated-measures ANOVA was used to investigate the difference between these parameters from the three devices (P < 0.05). We found the CSA measured from CIRRUS was significantly larger than PLEX, followed by CASIA. The contrast was highest in CIRRUS, followed by PLEX and CASIA. The coverage was also higher in CIRRUS as compared to PLEX and CASIA. No significant difference was seen in the continuity from the three devices. In summary, we showed the measurements from the three devices were not interchangeable. Nature Publishing Group UK 2022-01-26 /pmc/articles/PMC8792050/ /pubmed/35082345 http://dx.doi.org/10.1038/s41598-022-05410-9 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Article
Wu, Xuan
Tan, Bingyao
Gan, Jinyuan
Lam, Adeline R.
Chen, Yibing
Liu, Xinyu
Chua, Jacqueline
Wong, Damon W. K.
Ang, Marcus
Schmetterer, Leopold
Yao, Xinwen
Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal
title Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal
title_full Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal
title_fullStr Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal
title_full_unstemmed Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal
title_short Evaluation of different OCT systems in quantitative imaging of human Schlemm's canal
title_sort evaluation of different oct systems in quantitative imaging of human schlemm's canal
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8792050/
https://www.ncbi.nlm.nih.gov/pubmed/35082345
http://dx.doi.org/10.1038/s41598-022-05410-9
work_keys_str_mv AT wuxuan evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT tanbingyao evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT ganjinyuan evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT lamadeliner evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT chenyibing evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT liuxinyu evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT chuajacqueline evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT wongdamonwk evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT angmarcus evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT schmettererleopold evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal
AT yaoxinwen evaluationofdifferentoctsystemsinquantitativeimagingofhumanschlemmscanal