Bearing failure in a mobile bearing unicompartmental knee arthroplasty: an uncommon presentation of an implant-specific complication
BACKGROUND: We present two cases of unicompartmental knee arthroplasty (UKA) bearing failure in this report—one case of bearing dislocation and one case of bearing fracture. The causes of failure in both cases are evaluated in depth and recommendations are provided regarding intraoperative technique...