Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients
BACKGROUND: To evaluate the dissociation effect of optical fluorescence platelet counting of BC-6800 hematology analyzer on ethylene diamine tetraacetic acid-dependent pseudo thrombocytopenia (EDTA-PTCP) samples. METHODS: Twenty-three finally identified EDTA-PTCP samples were recruited in this study...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
AME Publishing Company
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8798320/ https://www.ncbi.nlm.nih.gov/pubmed/35117170 http://dx.doi.org/10.21037/tcr.2019.12.58 |
_version_ | 1784641775351103488 |
---|---|
author | Bao, Yejiang Wang, Jia Wang, Anjun Bian, Jia Jin, Yuan |
author_facet | Bao, Yejiang Wang, Jia Wang, Anjun Bian, Jia Jin, Yuan |
author_sort | Bao, Yejiang |
collection | PubMed |
description | BACKGROUND: To evaluate the dissociation effect of optical fluorescence platelet counting of BC-6800 hematology analyzer on ethylene diamine tetraacetic acid-dependent pseudo thrombocytopenia (EDTA-PTCP) samples. METHODS: Twenty-three finally identified EDTA-PTCP samples were recruited in this study using criteria as follow: (I) impedance platelet counts lower than 100×10(9)/L with instrument “platelet aggregation” flag; (II) existence of platelet clumps in the blood smear; (III) obviously higher platelet counts without “platelet aggregation” flag and no platelet clumps in blood smear after repeating phlebotomy using citrate anticoagulated tubes. The BC-6800 hematology analyzer and the XE-2100 hematology analyzer were used to test 23 EDTA-PTCP samples and 30 controls on both the impedance channel and the reticulocyte channel. The dissociation rate was defined as optical fluorescence platelet counts in the EDTA tubes/impedance platelet counts in citrate tubes ×100%. RESULTS: BC-6800 analyzer’s optical fluorescence platelet counts of EDTA-PTCP samples were significantly higher than impedance platelet counts (t=4.33, P=0.00) and comparable with the platelet counts of re-collected samples in tubes containing citrate anticoagulant. On BC-6800 hematology analyzer, 22 of 23 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 93%. On the XE-2100 hematology analyzer, 1 of the 17 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 56%. CONCLUSIONS: Optical fluorescence platelet counting of BC-6800 Hematology Analyzer is effective for the correction of spurious low platelet counts in EDTA-PTCP patients, and its dissociation effect on EDTA-PTCP samples was independent of fluorescent dye staining. |
format | Online Article Text |
id | pubmed-8798320 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | AME Publishing Company |
record_format | MEDLINE/PubMed |
spelling | pubmed-87983202022-02-02 Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients Bao, Yejiang Wang, Jia Wang, Anjun Bian, Jia Jin, Yuan Transl Cancer Res Original Article BACKGROUND: To evaluate the dissociation effect of optical fluorescence platelet counting of BC-6800 hematology analyzer on ethylene diamine tetraacetic acid-dependent pseudo thrombocytopenia (EDTA-PTCP) samples. METHODS: Twenty-three finally identified EDTA-PTCP samples were recruited in this study using criteria as follow: (I) impedance platelet counts lower than 100×10(9)/L with instrument “platelet aggregation” flag; (II) existence of platelet clumps in the blood smear; (III) obviously higher platelet counts without “platelet aggregation” flag and no platelet clumps in blood smear after repeating phlebotomy using citrate anticoagulated tubes. The BC-6800 hematology analyzer and the XE-2100 hematology analyzer were used to test 23 EDTA-PTCP samples and 30 controls on both the impedance channel and the reticulocyte channel. The dissociation rate was defined as optical fluorescence platelet counts in the EDTA tubes/impedance platelet counts in citrate tubes ×100%. RESULTS: BC-6800 analyzer’s optical fluorescence platelet counts of EDTA-PTCP samples were significantly higher than impedance platelet counts (t=4.33, P=0.00) and comparable with the platelet counts of re-collected samples in tubes containing citrate anticoagulant. On BC-6800 hematology analyzer, 22 of 23 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 93%. On the XE-2100 hematology analyzer, 1 of the 17 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 56%. CONCLUSIONS: Optical fluorescence platelet counting of BC-6800 Hematology Analyzer is effective for the correction of spurious low platelet counts in EDTA-PTCP patients, and its dissociation effect on EDTA-PTCP samples was independent of fluorescent dye staining. AME Publishing Company 2020-01 /pmc/articles/PMC8798320/ /pubmed/35117170 http://dx.doi.org/10.21037/tcr.2019.12.58 Text en 2020 Translational Cancer Research. All rights reserved. https://creativecommons.org/licenses/by-nc-nd/4.0/Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/. |
spellingShingle | Original Article Bao, Yejiang Wang, Jia Wang, Anjun Bian, Jia Jin, Yuan Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients |
title | Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients |
title_full | Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients |
title_fullStr | Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients |
title_full_unstemmed | Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients |
title_short | Correction of spurious low platelet counts by optical fluorescence platelet counting of BC-6800 hematology analyzer in EDTA-dependent pseudo thrombocytopenia patients |
title_sort | correction of spurious low platelet counts by optical fluorescence platelet counting of bc-6800 hematology analyzer in edta-dependent pseudo thrombocytopenia patients |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8798320/ https://www.ncbi.nlm.nih.gov/pubmed/35117170 http://dx.doi.org/10.21037/tcr.2019.12.58 |
work_keys_str_mv | AT baoyejiang correctionofspuriouslowplateletcountsbyopticalfluorescenceplateletcountingofbc6800hematologyanalyzerinedtadependentpseudothrombocytopeniapatients AT wangjia correctionofspuriouslowplateletcountsbyopticalfluorescenceplateletcountingofbc6800hematologyanalyzerinedtadependentpseudothrombocytopeniapatients AT wanganjun correctionofspuriouslowplateletcountsbyopticalfluorescenceplateletcountingofbc6800hematologyanalyzerinedtadependentpseudothrombocytopeniapatients AT bianjia correctionofspuriouslowplateletcountsbyopticalfluorescenceplateletcountingofbc6800hematologyanalyzerinedtadependentpseudothrombocytopeniapatients AT jinyuan correctionofspuriouslowplateletcountsbyopticalfluorescenceplateletcountingofbc6800hematologyanalyzerinedtadependentpseudothrombocytopeniapatients |