Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy

BACKGROUND: To assess the position accuracy of the six-degree-of-freedom (6-DoF) couch based on cone beam computed tomography (CBCT) and exploit the correlation of the six degrees errors. METHODS: CT scans of an anthropomorphic phantom and patients were obtained at 3 mm slice thicknesses using a Phi...

Descripción completa

Detalles Bibliográficos
Autores principales: Fu, Caofei, Ma, Changsheng, Shang, Dongping, Qiu, Qingtao, Meng, Huipeng, Duan, Jinghao, Yin, Yong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: AME Publishing Company 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8798941/
https://www.ncbi.nlm.nih.gov/pubmed/35117212
http://dx.doi.org/10.21037/tcr-20-1528
_version_ 1784641944566104064
author Fu, Caofei
Ma, Changsheng
Shang, Dongping
Qiu, Qingtao
Meng, Huipeng
Duan, Jinghao
Yin, Yong
author_facet Fu, Caofei
Ma, Changsheng
Shang, Dongping
Qiu, Qingtao
Meng, Huipeng
Duan, Jinghao
Yin, Yong
author_sort Fu, Caofei
collection PubMed
description BACKGROUND: To assess the position accuracy of the six-degree-of-freedom (6-DoF) couch based on cone beam computed tomography (CBCT) and exploit the correlation of the six degrees errors. METHODS: CT scans of an anthropomorphic phantom and patients were obtained at 3 mm slice thicknesses using a Philips scanner at the head, first supine and prostrate positions. An Eclipse Treatment Planning System was used to create a treatment plan. Different levels of known position errors were introduced to simulate patient position status for the anthropomorphic phantom. CBCT datasets for each position were acquired and registered to original CT datasets to evaluate the accuracy of the 6-DoF couch and determine the setup errors of patients. The setup errors of 200 CBCT datasets from 14 patients with pelvic tumors were analyzed. The correlations between six degrees position errors were finally extracted. RESULTS: For the phantom study, the difference between known introduced errors and the setup errors were almost negligible. The deviation (mean ± one standard deviation) in registration methods were (0.01±0.02) cm, (0.04±0.075) cm, (0.02±0.004) cm, (0.01±0.04)°, (0.1±0.08)°, (0.03±0.05)° and (0.01±0.01) cm, (0.03±0.007) cm, (0.01±0.01) cm, (0.05±0.06)°, (0.08±0.08)°, (0.04±0.05)° for supine and prone position, respectively. The deviation in positions were (0.07±0.10) cm, (0.16±0.02) cm, (0.08±0.06) cm, (0.54±0.46)°, (0.24±0.16)°, (0.09±0.09)° and (0.06±0.09) cm, (0.19±0.09) cm, (0.09±0.07) cm, (0.49±0.49)°, (0.16±0.08)°, (0.1±0.13)° for bone and soft tissue registration methods, respectively. For patient data, the setup errors were (−0.07±0.22) cm, (0.14±0.35) cm, (−0.12±0.4) cm, (0.79±1.6)°, (0.41±0.71)°, (−0.03±0.8)° for supine position and (0.16±0.27) cm, (0.19±0.48) cm, (−0.05±0.34) cm, (1.1±1.49)°, (0.65±1.00)°, (−0.23±0.75)° for prone position, respectively. There is a significant moderate correlation between the longitudinal and pitch directions and between the vertical and pitch directions when the patient is in the supine position. CONCLUSIONS: The six-dimensional couch positioning verification system based on CBCT has high accuracy and can meet the requirements of precise radiotherapy for pelvic tumors. There is a certain correlation between translation direction and rotation direction.
format Online
Article
Text
id pubmed-8798941
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher AME Publishing Company
record_format MEDLINE/PubMed
spelling pubmed-87989412022-02-02 Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy Fu, Caofei Ma, Changsheng Shang, Dongping Qiu, Qingtao Meng, Huipeng Duan, Jinghao Yin, Yong Transl Cancer Res Original Article BACKGROUND: To assess the position accuracy of the six-degree-of-freedom (6-DoF) couch based on cone beam computed tomography (CBCT) and exploit the correlation of the six degrees errors. METHODS: CT scans of an anthropomorphic phantom and patients were obtained at 3 mm slice thicknesses using a Philips scanner at the head, first supine and prostrate positions. An Eclipse Treatment Planning System was used to create a treatment plan. Different levels of known position errors were introduced to simulate patient position status for the anthropomorphic phantom. CBCT datasets for each position were acquired and registered to original CT datasets to evaluate the accuracy of the 6-DoF couch and determine the setup errors of patients. The setup errors of 200 CBCT datasets from 14 patients with pelvic tumors were analyzed. The correlations between six degrees position errors were finally extracted. RESULTS: For the phantom study, the difference between known introduced errors and the setup errors were almost negligible. The deviation (mean ± one standard deviation) in registration methods were (0.01±0.02) cm, (0.04±0.075) cm, (0.02±0.004) cm, (0.01±0.04)°, (0.1±0.08)°, (0.03±0.05)° and (0.01±0.01) cm, (0.03±0.007) cm, (0.01±0.01) cm, (0.05±0.06)°, (0.08±0.08)°, (0.04±0.05)° for supine and prone position, respectively. The deviation in positions were (0.07±0.10) cm, (0.16±0.02) cm, (0.08±0.06) cm, (0.54±0.46)°, (0.24±0.16)°, (0.09±0.09)° and (0.06±0.09) cm, (0.19±0.09) cm, (0.09±0.07) cm, (0.49±0.49)°, (0.16±0.08)°, (0.1±0.13)° for bone and soft tissue registration methods, respectively. For patient data, the setup errors were (−0.07±0.22) cm, (0.14±0.35) cm, (−0.12±0.4) cm, (0.79±1.6)°, (0.41±0.71)°, (−0.03±0.8)° for supine position and (0.16±0.27) cm, (0.19±0.48) cm, (−0.05±0.34) cm, (1.1±1.49)°, (0.65±1.00)°, (−0.23±0.75)° for prone position, respectively. There is a significant moderate correlation between the longitudinal and pitch directions and between the vertical and pitch directions when the patient is in the supine position. CONCLUSIONS: The six-dimensional couch positioning verification system based on CBCT has high accuracy and can meet the requirements of precise radiotherapy for pelvic tumors. There is a certain correlation between translation direction and rotation direction. AME Publishing Company 2020-10 /pmc/articles/PMC8798941/ /pubmed/35117212 http://dx.doi.org/10.21037/tcr-20-1528 Text en 2020 Translational Cancer Research. All rights reserved. https://creativecommons.org/licenses/by-nc-nd/4.0/Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.
spellingShingle Original Article
Fu, Caofei
Ma, Changsheng
Shang, Dongping
Qiu, Qingtao
Meng, Huipeng
Duan, Jinghao
Yin, Yong
Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy
title Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy
title_full Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy
title_fullStr Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy
title_full_unstemmed Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy
title_short Geometric accuracy evaluation of a six-degree-of-freedom (6-DoF) couch with cone beam computed tomography (CBCT) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy
title_sort geometric accuracy evaluation of a six-degree-of-freedom (6-dof) couch with cone beam computed tomography (cbct) using a phantom and correlation study of the position errors in pelvic tumor radiotherapy
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8798941/
https://www.ncbi.nlm.nih.gov/pubmed/35117212
http://dx.doi.org/10.21037/tcr-20-1528
work_keys_str_mv AT fucaofei geometricaccuracyevaluationofasixdegreeoffreedom6dofcouchwithconebeamcomputedtomographycbctusingaphantomandcorrelationstudyofthepositionerrorsinpelvictumorradiotherapy
AT machangsheng geometricaccuracyevaluationofasixdegreeoffreedom6dofcouchwithconebeamcomputedtomographycbctusingaphantomandcorrelationstudyofthepositionerrorsinpelvictumorradiotherapy
AT shangdongping geometricaccuracyevaluationofasixdegreeoffreedom6dofcouchwithconebeamcomputedtomographycbctusingaphantomandcorrelationstudyofthepositionerrorsinpelvictumorradiotherapy
AT qiuqingtao geometricaccuracyevaluationofasixdegreeoffreedom6dofcouchwithconebeamcomputedtomographycbctusingaphantomandcorrelationstudyofthepositionerrorsinpelvictumorradiotherapy
AT menghuipeng geometricaccuracyevaluationofasixdegreeoffreedom6dofcouchwithconebeamcomputedtomographycbctusingaphantomandcorrelationstudyofthepositionerrorsinpelvictumorradiotherapy
AT duanjinghao geometricaccuracyevaluationofasixdegreeoffreedom6dofcouchwithconebeamcomputedtomographycbctusingaphantomandcorrelationstudyofthepositionerrorsinpelvictumorradiotherapy
AT yinyong geometricaccuracyevaluationofasixdegreeoffreedom6dofcouchwithconebeamcomputedtomographycbctusingaphantomandcorrelationstudyofthepositionerrorsinpelvictumorradiotherapy