Cargando…

Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system

In the electron beam radiation therapy, customized blocks are mostly used to shape treatment fields to generate conformal doses. The goal of this study is to investigate quantitatively dosimetric uncertainties associated with heterogeneities, detectors used in the measurement of the beam data commis...

Descripción completa

Detalles Bibliográficos
Autores principales: Ali, Imad, Alsbou, Nesreen, Ahmad, Salahuddin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8803289/
https://www.ncbi.nlm.nih.gov/pubmed/34822731
http://dx.doi.org/10.1002/acm2.13478
_version_ 1784642837283864576
author Ali, Imad
Alsbou, Nesreen
Ahmad, Salahuddin
author_facet Ali, Imad
Alsbou, Nesreen
Ahmad, Salahuddin
author_sort Ali, Imad
collection PubMed
description In the electron beam radiation therapy, customized blocks are mostly used to shape treatment fields to generate conformal doses. The goal of this study is to investigate quantitatively dosimetric uncertainties associated with heterogeneities, detectors used in the measurement of the beam data commissioning, and modeling of the interactions of high energy electrons with tissue. These uncertainties were investigated both by measurements with different detectors and calculations using electron Monte Carlo algorithm (eMC) in the Eclipse treatment planning system. Dose distributions for different field sizes were calculated using eMC and measured with a multiple‐diode‐array detector (MapCheck2) for cone sizes ranging from 6 to 25 cm. The dose distributions were calculated using the CT images of the MapCheck2 and water‐equivalent phantoms. In the umbra region (<20% isodose line), the eMC underestimated dose by a factor of 3 for high energy electron beams due to lack of consideration of bremsstrahlung emitted laterally that was not accounted by eMC in the low dose region outside the field. In the penumbra (20%–80% isodose line), the eMC overestimated dose (40%) for high energy 20 MeV electrons compared to the measured dose with small diodes in the high gradient dose region. This was mainly due to lack of consideration of volume averaging of the ion chamber used in beam data commissioning which was input to the eMC dose calculation algorithm. Large uncertainties in the CT numbers (25%) resulted from the image artifacts in the CT images of the MapCheck2 phantom due to metal artifacts. The eMC algorithm used the electron and material densities extracted from the CT numbers which resulted large dosimetric uncertainties (10%) in the material densities and corresponding stopping power ratios. The dose calculations with eMC are associated with large uncertainties particularly in penumbra and umbra regions and around heterogeneities which affect the low dose level that cover nearby normal tissue or critical structures.
format Online
Article
Text
id pubmed-8803289
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-88032892022-02-04 Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system Ali, Imad Alsbou, Nesreen Ahmad, Salahuddin J Appl Clin Med Phys Radiation Oncology Physics In the electron beam radiation therapy, customized blocks are mostly used to shape treatment fields to generate conformal doses. The goal of this study is to investigate quantitatively dosimetric uncertainties associated with heterogeneities, detectors used in the measurement of the beam data commissioning, and modeling of the interactions of high energy electrons with tissue. These uncertainties were investigated both by measurements with different detectors and calculations using electron Monte Carlo algorithm (eMC) in the Eclipse treatment planning system. Dose distributions for different field sizes were calculated using eMC and measured with a multiple‐diode‐array detector (MapCheck2) for cone sizes ranging from 6 to 25 cm. The dose distributions were calculated using the CT images of the MapCheck2 and water‐equivalent phantoms. In the umbra region (<20% isodose line), the eMC underestimated dose by a factor of 3 for high energy electron beams due to lack of consideration of bremsstrahlung emitted laterally that was not accounted by eMC in the low dose region outside the field. In the penumbra (20%–80% isodose line), the eMC overestimated dose (40%) for high energy 20 MeV electrons compared to the measured dose with small diodes in the high gradient dose region. This was mainly due to lack of consideration of volume averaging of the ion chamber used in beam data commissioning which was input to the eMC dose calculation algorithm. Large uncertainties in the CT numbers (25%) resulted from the image artifacts in the CT images of the MapCheck2 phantom due to metal artifacts. The eMC algorithm used the electron and material densities extracted from the CT numbers which resulted large dosimetric uncertainties (10%) in the material densities and corresponding stopping power ratios. The dose calculations with eMC are associated with large uncertainties particularly in penumbra and umbra regions and around heterogeneities which affect the low dose level that cover nearby normal tissue or critical structures. John Wiley and Sons Inc. 2021-11-25 /pmc/articles/PMC8803289/ /pubmed/34822731 http://dx.doi.org/10.1002/acm2.13478 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Ali, Imad
Alsbou, Nesreen
Ahmad, Salahuddin
Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system
title Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system
title_full Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system
title_fullStr Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system
title_full_unstemmed Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system
title_short Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system
title_sort quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron monte carlo dose algorithm in the eclipse treatment planning system
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8803289/
https://www.ncbi.nlm.nih.gov/pubmed/34822731
http://dx.doi.org/10.1002/acm2.13478
work_keys_str_mv AT aliimad quantitativeevaluationofdosimetricuncertaintiesinelectrontherapybymeasurementandcalculationusingtheelectronmontecarlodosealgorithmintheeclipsetreatmentplanningsystem
AT alsbounesreen quantitativeevaluationofdosimetricuncertaintiesinelectrontherapybymeasurementandcalculationusingtheelectronmontecarlodosealgorithmintheeclipsetreatmentplanningsystem
AT ahmadsalahuddin quantitativeevaluationofdosimetricuncertaintiesinelectrontherapybymeasurementandcalculationusingtheelectronmontecarlodosealgorithmintheeclipsetreatmentplanningsystem