Multi-locus transcranial magnetic stimulation system for electronically targeted brain stimulation

BACKGROUND: Transcranial magnetic stimulation (TMS) allows non-invasive stimulation of the cortex. In multi-locus TMS (mTMS), the stimulating electric field (E-field) is controlled electronically without coil movement by adjusting currents in the coils of a transducer. OBJECTIVE: To develop an mTMS...

Descripción completa

Detalles Bibliográficos
Autores principales: Nieminen, Jaakko O., Sinisalo, Heikki, Souza, Victor H., Malmi, Mikko, Yuryev, Mikhail, Tervo, Aino E., Stenroos, Matti, Milardovich, Diego, Korhonen, Juuso T., Koponen, Lari M., Ilmoniemi, Risto J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8807400/
https://www.ncbi.nlm.nih.gov/pubmed/34818580
http://dx.doi.org/10.1016/j.brs.2021.11.014
Descripción
Sumario:BACKGROUND: Transcranial magnetic stimulation (TMS) allows non-invasive stimulation of the cortex. In multi-locus TMS (mTMS), the stimulating electric field (E-field) is controlled electronically without coil movement by adjusting currents in the coils of a transducer. OBJECTIVE: To develop an mTMS system that allows adjusting the location and orientation of the E-field maximum within a cortical region. METHODS: We designed and manufactured a planar 5-coil mTMS transducer to allow controlling the maximum of the induced E-field within a cortical region approximately 30 mm in diameter. We developed electronics with a design consisting of independently controlled H-bridge circuits to drive up to six TMS coils. To control the hardware, we programmed software that runs on a field-programmable gate array and a computer. To induce the desired E-field in the cortex, we developed an optimization method to calculate the currents needed in the coils. We characterized the mTMS system and conducted a proof-of-concept motor-mapping experiment on a healthy volunteer. In the motor mapping, we kept the transducer placement fixed while electronically shifting the E-field maximum on the precentral gyrus and measuring electromyography from the contralateral hand. RESULTS: The transducer consists of an oval coil, two figure-of-eight coils, and two four-leaf-clover coils stacked on top of each other. The technical characterization indicated that the mTMS system performs as designed. The measured motor evoked potential amplitudes varied consistently as a function of the location of the E-field maximum. CONCLUSION: The developed mTMS system enables electronically targeted brain stimulation within a cortical region.