Cargando…
Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing
BACKGROUND: The safety and efficacy analysis of flow redirection lumen device (FRED) demonstrated the excellent safety profile of FREDs for aneurysm treatment. We describe the first case in which FRED deployment for a paraclinoid aneurysm resulted in in-stent stenosis, necessitating balloon angiopla...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Scientific Scholar
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8813601/ https://www.ncbi.nlm.nih.gov/pubmed/35127222 http://dx.doi.org/10.25259/SNI_1127_2021 |
_version_ | 1784644894193614848 |
---|---|
author | Beppu, Mikiya Kuramoto, Yoji Abe, Soichiro Namitome, Satoshi Yoshimura, Shinichi |
author_facet | Beppu, Mikiya Kuramoto, Yoji Abe, Soichiro Namitome, Satoshi Yoshimura, Shinichi |
author_sort | Beppu, Mikiya |
collection | PubMed |
description | BACKGROUND: The safety and efficacy analysis of flow redirection lumen device (FRED) demonstrated the excellent safety profile of FREDs for aneurysm treatment. We describe the first case in which FRED deployment for a paraclinoid aneurysm resulted in in-stent stenosis, necessitating balloon angioplasty, and an additional stent. CASE DESCRIPTION: A 50-year-old woman had a left paraclinoid aneurysm with a maximum diameter of 6.1 mm. We planned FRED deployment. We experienced in-stent stenosis just after the deployment of a FRED. Devices such as guidewires and catheters could not cross the lesion through the FRED because of an obstruction in the FRED. Balloon angioplasty and subsequent stenting resolved thrombosis and kinking. FRED has potential for kinking locally. CONCLUSION: Surgeons should consider this possibility when treating cerebral aneurysm using FRED. Cone-beam computed tomography after deployment of FRED may be useful for evaluating the stent shape. |
format | Online Article Text |
id | pubmed-8813601 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Scientific Scholar |
record_format | MEDLINE/PubMed |
spelling | pubmed-88136012022-02-04 Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing Beppu, Mikiya Kuramoto, Yoji Abe, Soichiro Namitome, Satoshi Yoshimura, Shinichi Surg Neurol Int Case Report BACKGROUND: The safety and efficacy analysis of flow redirection lumen device (FRED) demonstrated the excellent safety profile of FREDs for aneurysm treatment. We describe the first case in which FRED deployment for a paraclinoid aneurysm resulted in in-stent stenosis, necessitating balloon angioplasty, and an additional stent. CASE DESCRIPTION: A 50-year-old woman had a left paraclinoid aneurysm with a maximum diameter of 6.1 mm. We planned FRED deployment. We experienced in-stent stenosis just after the deployment of a FRED. Devices such as guidewires and catheters could not cross the lesion through the FRED because of an obstruction in the FRED. Balloon angioplasty and subsequent stenting resolved thrombosis and kinking. FRED has potential for kinking locally. CONCLUSION: Surgeons should consider this possibility when treating cerebral aneurysm using FRED. Cone-beam computed tomography after deployment of FRED may be useful for evaluating the stent shape. Scientific Scholar 2022-01-20 /pmc/articles/PMC8813601/ /pubmed/35127222 http://dx.doi.org/10.25259/SNI_1127_2021 Text en Copyright: © 2022 Surgical Neurology International https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-Share Alike 4.0 License, which allows others to remix, transform, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Case Report Beppu, Mikiya Kuramoto, Yoji Abe, Soichiro Namitome, Satoshi Yoshimura, Shinichi Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing |
title | Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing |
title_full | Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing |
title_fullStr | Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing |
title_full_unstemmed | Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing |
title_short | Localized kinking during deployment of a flow redirection lumen device (FRED) could be due to excessive pushing |
title_sort | localized kinking during deployment of a flow redirection lumen device (fred) could be due to excessive pushing |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8813601/ https://www.ncbi.nlm.nih.gov/pubmed/35127222 http://dx.doi.org/10.25259/SNI_1127_2021 |
work_keys_str_mv | AT beppumikiya localizedkinkingduringdeploymentofaflowredirectionlumendevicefredcouldbeduetoexcessivepushing AT kuramotoyoji localizedkinkingduringdeploymentofaflowredirectionlumendevicefredcouldbeduetoexcessivepushing AT abesoichiro localizedkinkingduringdeploymentofaflowredirectionlumendevicefredcouldbeduetoexcessivepushing AT namitomesatoshi localizedkinkingduringdeploymentofaflowredirectionlumendevicefredcouldbeduetoexcessivepushing AT yoshimurashinichi localizedkinkingduringdeploymentofaflowredirectionlumendevicefredcouldbeduetoexcessivepushing |