Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating
Objective The aim of this study was to evaluate the use of low peak kilovoltage (kVp) low-volume iodinated contrast protocol for performing coronary computed tomography (CT) angiography (CCTA) in patients using retrospective electrocardiogram (ECG) gating. Materials and Methods Hundred prospective...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Thieme Medical and Scientific Publishers Pvt. Ltd.
2022
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8817802/ https://www.ncbi.nlm.nih.gov/pubmed/35136504 http://dx.doi.org/10.1055/s-0041-1741102 |
_version_ | 1784645716017152000 |
---|---|
author | Kapoor, Atul Mahajan, Goldaa Kapoor, Aprajita |
author_facet | Kapoor, Atul Mahajan, Goldaa Kapoor, Aprajita |
author_sort | Kapoor, Atul |
collection | PubMed |
description | Objective The aim of this study was to evaluate the use of low peak kilovoltage (kVp) low-volume iodinated contrast protocol for performing coronary computed tomography (CT) angiography (CCTA) in patients using retrospective electrocardiogram (ECG) gating. Materials and Methods Hundred prospective patients undergoing CCTA were studied in two groups, A and B, using 70 kilovoltage (kV) and 120 kV protocols with half and standard intravenous volumes of injected iodinated contrast, respectively. All patients had heart rates less than 100 beats/min and body mass index (BMI) less than 31 kg/m (2) . Both the groups were evaluated for signal-to-noise (S/N) and contrast-to-noise (C/N) ratios along with radiation dose delivered in millisievert (mSv), and for image quality (IQ), on per patient and per segment basis. Results Patients with group A showed statistically reduced radiation dose of 1.86 mSv compared with 6.86 mSv in group B patients. Marked reduction in image noise with statistically improved S/N and C/N ratios in all coronary vessels was seen in group A. S/N ratios in group A were 20.25, 18.68, 19.04, 17.41, and 18.69 for aorta, left main, left anterior descending, right coronary, and left circumflex arteries while they were 13.34, 11.12, 10.96, 9.74, and 8.67 in group B patients. C/N ratios were also higher in all vessels in group A patients, that is, 19.48, 19.48, 19.04, 19.48, and 17.68, compared with group B patients, who had 12.43, 10.03, 9.23, 9.57, and 8.23 ratios ( p < 0.0001). No significant difference in IQ per patient and per vessel was seen between both the groups. Discussion Retrospective ECG-gated low-kVp low-volume iodinated contrast protocol provides good diagnostic quality angiograms in patients with BMI up to 31 kg/m (2) and with heart rates of less than 100 beats/min with three times reduced radiation dose. The reduced volume of contrast reduces the cost as well as the chance of contrast-induced nephropathy. |
format | Online Article Text |
id | pubmed-8817802 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Thieme Medical and Scientific Publishers Pvt. Ltd. |
record_format | MEDLINE/PubMed |
spelling | pubmed-88178022022-02-07 Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating Kapoor, Atul Mahajan, Goldaa Kapoor, Aprajita Indian J Radiol Imaging Objective The aim of this study was to evaluate the use of low peak kilovoltage (kVp) low-volume iodinated contrast protocol for performing coronary computed tomography (CT) angiography (CCTA) in patients using retrospective electrocardiogram (ECG) gating. Materials and Methods Hundred prospective patients undergoing CCTA were studied in two groups, A and B, using 70 kilovoltage (kV) and 120 kV protocols with half and standard intravenous volumes of injected iodinated contrast, respectively. All patients had heart rates less than 100 beats/min and body mass index (BMI) less than 31 kg/m (2) . Both the groups were evaluated for signal-to-noise (S/N) and contrast-to-noise (C/N) ratios along with radiation dose delivered in millisievert (mSv), and for image quality (IQ), on per patient and per segment basis. Results Patients with group A showed statistically reduced radiation dose of 1.86 mSv compared with 6.86 mSv in group B patients. Marked reduction in image noise with statistically improved S/N and C/N ratios in all coronary vessels was seen in group A. S/N ratios in group A were 20.25, 18.68, 19.04, 17.41, and 18.69 for aorta, left main, left anterior descending, right coronary, and left circumflex arteries while they were 13.34, 11.12, 10.96, 9.74, and 8.67 in group B patients. C/N ratios were also higher in all vessels in group A patients, that is, 19.48, 19.48, 19.04, 19.48, and 17.68, compared with group B patients, who had 12.43, 10.03, 9.23, 9.57, and 8.23 ratios ( p < 0.0001). No significant difference in IQ per patient and per vessel was seen between both the groups. Discussion Retrospective ECG-gated low-kVp low-volume iodinated contrast protocol provides good diagnostic quality angiograms in patients with BMI up to 31 kg/m (2) and with heart rates of less than 100 beats/min with three times reduced radiation dose. The reduced volume of contrast reduces the cost as well as the chance of contrast-induced nephropathy. Thieme Medical and Scientific Publishers Pvt. Ltd. 2022-01-10 /pmc/articles/PMC8817802/ /pubmed/35136504 http://dx.doi.org/10.1055/s-0041-1741102 Text en Indian Radiological Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. ( https://creativecommons.org/licenses/by-nc-nd/4.0/ ) https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives License, which permits unrestricted reproduction and distribution, for non-commercial purposes only; and use and reproduction, but not distribution, of adapted material for non-commercial purposes only, provided the original work is properly cited. |
spellingShingle | Kapoor, Atul Mahajan, Goldaa Kapoor, Aprajita Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating |
title | Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating |
title_full | Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating |
title_fullStr | Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating |
title_full_unstemmed | Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating |
title_short | Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating |
title_sort | evaluation of low-kvp low-volume iodinated contrast protocol for coronary ct angiography using retrospective ecg gating |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8817802/ https://www.ncbi.nlm.nih.gov/pubmed/35136504 http://dx.doi.org/10.1055/s-0041-1741102 |
work_keys_str_mv | AT kapooratul evaluationoflowkvplowvolumeiodinatedcontrastprotocolforcoronaryctangiographyusingretrospectiveecggating AT mahajangoldaa evaluationoflowkvplowvolumeiodinatedcontrastprotocolforcoronaryctangiographyusingretrospectiveecggating AT kapooraprajita evaluationoflowkvplowvolumeiodinatedcontrastprotocolforcoronaryctangiographyusingretrospectiveecggating |