Cargando…

Extended Use of The Spanner® Temporary Prostatic Stent in Catheter-Dependent Men with Comorbidities

PURPOSE: This US FDA investigational device exemption (IDE) study evaluated the extended use of The Spanner® Temporary Prostatic Stent in catheter-dependent men with urinary retention who were not deemed candidates for corrective surgery but demonstrated bladder contractility. MATERIALS AND METHODS:...

Descripción completa

Detalles Bibliográficos
Autores principales: Cambio, Angelo J., Roach, Richard M., Arnold, Paul, Cambio, Joseph, Gluck, Clifford D., Heron, Sean P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8831058/
https://www.ncbi.nlm.nih.gov/pubmed/35154313
http://dx.doi.org/10.1155/2022/7367851
Descripción
Sumario:PURPOSE: This US FDA investigational device exemption (IDE) study evaluated the extended use of The Spanner® Temporary Prostatic Stent in catheter-dependent men with urinary retention who were not deemed candidates for corrective surgery but demonstrated bladder contractility. MATERIALS AND METHODS: The Spanner was placed for 3 cycles of 30 days in catheter-dependent men with comorbid conditions, confirmed detrusor contractility, and catheter-associated discomfort. At each visit, postvoid residual, maximum flow rate, international prostate symptom score, quality of life, and adverse events were assessed. Voiding success was defined as PVR ≤ 150 ml at all visits. RESULTS: One hundred seven men were enrolled at 8 US sites; 82/107 (76.6%) completed the trial, and 79/107 (73.8%) successfully maintained PVR ≤ 150 ml for the trial duration. Patients were 77.1 ± 10.6 years old; 63/107 (58.9%) were dependent on Foley and 40/107 (37.4%) on intermittent catheterization for 36.0 ± 39.3 days and 30.2 ± 45.8 days, respectively. 25/107 (23.4%) discontinuations were primarily due to voluntary patient withdrawal 9/107 (8.4%), investigator-initiated withdrawal 8/107 (7.5%), or lack of effectiveness 4/107 (3.7%). During Spanner use, the mean Q(max) was 11.2 ± 6.6, mean IPSS was 7.5 ± 6.4, and mean QOL was 2.0 ± 1.6. The most prevalent device-related adverse events were asymptomatic bacteriuria 25/107 (23.4%), discomfort 10/107 (9.4%), and urinary urgency 8/107 (7.5%). No device-related serious AEs were reported. CONCLUSIONS: This study demonstrates that catheter-dependent men with sufficient bladder contractility can achieve volitional voiding and successful bladder drainage using The Spanner Temporary Prostatic Stent for extended periods of time.