Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate

INTRODUCTION: Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precisio...

Descripción completa

Detalles Bibliográficos
Autores principales: Čičak, Helena, Šonjić, Pavica, Šimundić, Ana-Maria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833250/
https://www.ncbi.nlm.nih.gov/pubmed/35210928
http://dx.doi.org/10.11613/BM.2022.010708
_version_ 1784648889620496384
author Čičak, Helena
Šonjić, Pavica
Šimundić, Ana-Maria
author_facet Čičak, Helena
Šonjić, Pavica
Šimundić, Ana-Maria
author_sort Čičak, Helena
collection PubMed
description INTRODUCTION: Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precision, method comparison and accuracy of automated ESR analysers: Roller 20PN (Alifax S.p.A., Polverara, Italy) and iSED (Alcor Scientific, Smithfield, USA). MATERIALS AND METHODS: Blood samples (N = 752 for Roller 20PN and N = 213 for iSED) were sampled into K(2)EDTA (Kima, Italy) tubes for automated and 3.8% Na-citrate tubes (Kima, Italy) for Westergren method. The data was divided into three groups according to the ESR values obtained with the Westergren method: Group Low (L) (ESR ≤ 20 mm), Group Medium (M) (ESR 21-60 mm), and Group High (H) (ESR ≥ 61 mm). Method agreement was assessed by Bland-Altman analysis and Passing-Bablok regression. RESULTS: Analyser iSED has shown better comparability with Westergren method (bias 0.0 (95%Cl -1.4 to 1.5) range than Roller 20 PN (bias = - 6.4 (95%Cl - 7.1 to -5.7) in the whole measuring. For Roller 20 PN, Passing-Bablok regression has shown constant and proportional difference for Groups L and M, and for iSED only for Group H. Roller 20 PN had lower sensitivity (0.51 (95%Cl: 0.45-0.57) than iSED (0.72 (95%Cl: 0.59-0.80) while they had comparable specificity (> 0.90) and accuracy (≥ 0.80) in comparison with the Westergren method. CONCLUSION: Both analysers are not comparable with the Westergren method and should not be used interchangeably.
format Online
Article
Text
id pubmed-8833250
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-88332502022-02-23 Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate Čičak, Helena Šonjić, Pavica Šimundić, Ana-Maria Biochem Med (Zagreb) Original Articles INTRODUCTION: Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precision, method comparison and accuracy of automated ESR analysers: Roller 20PN (Alifax S.p.A., Polverara, Italy) and iSED (Alcor Scientific, Smithfield, USA). MATERIALS AND METHODS: Blood samples (N = 752 for Roller 20PN and N = 213 for iSED) were sampled into K(2)EDTA (Kima, Italy) tubes for automated and 3.8% Na-citrate tubes (Kima, Italy) for Westergren method. The data was divided into three groups according to the ESR values obtained with the Westergren method: Group Low (L) (ESR ≤ 20 mm), Group Medium (M) (ESR 21-60 mm), and Group High (H) (ESR ≥ 61 mm). Method agreement was assessed by Bland-Altman analysis and Passing-Bablok regression. RESULTS: Analyser iSED has shown better comparability with Westergren method (bias 0.0 (95%Cl -1.4 to 1.5) range than Roller 20 PN (bias = - 6.4 (95%Cl - 7.1 to -5.7) in the whole measuring. For Roller 20 PN, Passing-Bablok regression has shown constant and proportional difference for Groups L and M, and for iSED only for Group H. Roller 20 PN had lower sensitivity (0.51 (95%Cl: 0.45-0.57) than iSED (0.72 (95%Cl: 0.59-0.80) while they had comparable specificity (> 0.90) and accuracy (≥ 0.80) in comparison with the Westergren method. CONCLUSION: Both analysers are not comparable with the Westergren method and should not be used interchangeably. Croatian Society of Medical Biochemistry and Laboratory Medicine 2022-02-15 2022-02-15 /pmc/articles/PMC8833250/ /pubmed/35210928 http://dx.doi.org/10.11613/BM.2022.010708 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Čičak, Helena
Šonjić, Pavica
Šimundić, Ana-Maria
Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate
title Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate
title_full Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate
title_fullStr Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate
title_full_unstemmed Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate
title_short Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate
title_sort verification of automatic analysers roller 20pn and ised for measuring erythrocyte sedimentation rate
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833250/
https://www.ncbi.nlm.nih.gov/pubmed/35210928
http://dx.doi.org/10.11613/BM.2022.010708
work_keys_str_mv AT cicakhelena verificationofautomaticanalysersroller20pnandisedformeasuringerythrocytesedimentationrate
AT sonjicpavica verificationofautomaticanalysersroller20pnandisedformeasuringerythrocytesedimentationrate
AT simundicanamaria verificationofautomaticanalysersroller20pnandisedformeasuringerythrocytesedimentationrate