Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate
INTRODUCTION: Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precisio...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833250/ https://www.ncbi.nlm.nih.gov/pubmed/35210928 http://dx.doi.org/10.11613/BM.2022.010708 |
_version_ | 1784648889620496384 |
---|---|
author | Čičak, Helena Šonjić, Pavica Šimundić, Ana-Maria |
author_facet | Čičak, Helena Šonjić, Pavica Šimundić, Ana-Maria |
author_sort | Čičak, Helena |
collection | PubMed |
description | INTRODUCTION: Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precision, method comparison and accuracy of automated ESR analysers: Roller 20PN (Alifax S.p.A., Polverara, Italy) and iSED (Alcor Scientific, Smithfield, USA). MATERIALS AND METHODS: Blood samples (N = 752 for Roller 20PN and N = 213 for iSED) were sampled into K(2)EDTA (Kima, Italy) tubes for automated and 3.8% Na-citrate tubes (Kima, Italy) for Westergren method. The data was divided into three groups according to the ESR values obtained with the Westergren method: Group Low (L) (ESR ≤ 20 mm), Group Medium (M) (ESR 21-60 mm), and Group High (H) (ESR ≥ 61 mm). Method agreement was assessed by Bland-Altman analysis and Passing-Bablok regression. RESULTS: Analyser iSED has shown better comparability with Westergren method (bias 0.0 (95%Cl -1.4 to 1.5) range than Roller 20 PN (bias = - 6.4 (95%Cl - 7.1 to -5.7) in the whole measuring. For Roller 20 PN, Passing-Bablok regression has shown constant and proportional difference for Groups L and M, and for iSED only for Group H. Roller 20 PN had lower sensitivity (0.51 (95%Cl: 0.45-0.57) than iSED (0.72 (95%Cl: 0.59-0.80) while they had comparable specificity (> 0.90) and accuracy (≥ 0.80) in comparison with the Westergren method. CONCLUSION: Both analysers are not comparable with the Westergren method and should not be used interchangeably. |
format | Online Article Text |
id | pubmed-8833250 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-88332502022-02-23 Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate Čičak, Helena Šonjić, Pavica Šimundić, Ana-Maria Biochem Med (Zagreb) Original Articles INTRODUCTION: Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precision, method comparison and accuracy of automated ESR analysers: Roller 20PN (Alifax S.p.A., Polverara, Italy) and iSED (Alcor Scientific, Smithfield, USA). MATERIALS AND METHODS: Blood samples (N = 752 for Roller 20PN and N = 213 for iSED) were sampled into K(2)EDTA (Kima, Italy) tubes for automated and 3.8% Na-citrate tubes (Kima, Italy) for Westergren method. The data was divided into three groups according to the ESR values obtained with the Westergren method: Group Low (L) (ESR ≤ 20 mm), Group Medium (M) (ESR 21-60 mm), and Group High (H) (ESR ≥ 61 mm). Method agreement was assessed by Bland-Altman analysis and Passing-Bablok regression. RESULTS: Analyser iSED has shown better comparability with Westergren method (bias 0.0 (95%Cl -1.4 to 1.5) range than Roller 20 PN (bias = - 6.4 (95%Cl - 7.1 to -5.7) in the whole measuring. For Roller 20 PN, Passing-Bablok regression has shown constant and proportional difference for Groups L and M, and for iSED only for Group H. Roller 20 PN had lower sensitivity (0.51 (95%Cl: 0.45-0.57) than iSED (0.72 (95%Cl: 0.59-0.80) while they had comparable specificity (> 0.90) and accuracy (≥ 0.80) in comparison with the Westergren method. CONCLUSION: Both analysers are not comparable with the Westergren method and should not be used interchangeably. Croatian Society of Medical Biochemistry and Laboratory Medicine 2022-02-15 2022-02-15 /pmc/articles/PMC8833250/ /pubmed/35210928 http://dx.doi.org/10.11613/BM.2022.010708 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Articles Čičak, Helena Šonjić, Pavica Šimundić, Ana-Maria Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate |
title | Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate |
title_full | Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate |
title_fullStr | Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate |
title_full_unstemmed | Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate |
title_short | Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate |
title_sort | verification of automatic analysers roller 20pn and ised for measuring erythrocyte sedimentation rate |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833250/ https://www.ncbi.nlm.nih.gov/pubmed/35210928 http://dx.doi.org/10.11613/BM.2022.010708 |
work_keys_str_mv | AT cicakhelena verificationofautomaticanalysersroller20pnandisedformeasuringerythrocytesedimentationrate AT sonjicpavica verificationofautomaticanalysersroller20pnandisedformeasuringerythrocytesedimentationrate AT simundicanamaria verificationofautomaticanalysersroller20pnandisedformeasuringerythrocytesedimentationrate |