Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
PURPOSE: A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was es...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833290/ https://www.ncbi.nlm.nih.gov/pubmed/34905285 http://dx.doi.org/10.1002/acm2.13501 |
_version_ | 1784648899552608256 |
---|---|
author | Karius, Andre Karolczak, Marek Strnad, Vratislav Bert, Christoph |
author_facet | Karius, Andre Karolczak, Marek Strnad, Vratislav Bert, Christoph |
author_sort | Karius, Andre |
collection | PubMed |
description | PURPOSE: A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was essential to optimize both the imaging workflow and image quality for achieving the best possible clinical outcomes. We present the results of our investigations. METHODS: The novel CBCT‐system features a ring gantry with 121 cm clearance as well as a 43.2 × 43.2 cm(2) flat‐panel detector, and is controlled via a tablet‐personal computer (PC). For evaluating its imaging performance, the geometric reproducibility as well as imaging fidelity, computed tomography (CT)‐number accuracy, uniformity, contrast‐noise‐ratio (CNR), noise characteristics, and spatial resolution as fundamental image quality parameters were assessed. As dose metric the weighted cone‐beam dose index (CBDI(w)) was measured. Image quality was evaluated using standard quality assurance (QA) as well as anthropomorphic upper torso and breast phantoms. Both in‐house and manufacturer protocols for abdomen, pelvis, and breast imaging were examined. RESULTS: Using the in‐house protocols, the QA phantom scans showed altogether a high image quality, with high CT‐number accuracy (R (2 )> 0.97) and uniformity (<12 Hounsfield Unit (HU) cupping), reasonable noise and imaging fidelity, and good CNR at bone–tissue transitions of up to 28:1. Spatial resolution was strongly limited by geometric instabilities of the device. The breast phantom scans fulfilled clinical requirements, whereas the abdomen and pelvis scans showed severe artifacts, particularly at air/bone–tissue transitions. CONCLUSION: With the novel CBCT‐system, achieving a high image quality appears possible in principle. However, adaptations of the standard protocols, performance enhancements in image reconstruction referring to artifact reductions, as well as the extinction of geometric instabilities are imperative. |
format | Online Article Text |
id | pubmed-8833290 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-88332902022-02-14 Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy Karius, Andre Karolczak, Marek Strnad, Vratislav Bert, Christoph J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was essential to optimize both the imaging workflow and image quality for achieving the best possible clinical outcomes. We present the results of our investigations. METHODS: The novel CBCT‐system features a ring gantry with 121 cm clearance as well as a 43.2 × 43.2 cm(2) flat‐panel detector, and is controlled via a tablet‐personal computer (PC). For evaluating its imaging performance, the geometric reproducibility as well as imaging fidelity, computed tomography (CT)‐number accuracy, uniformity, contrast‐noise‐ratio (CNR), noise characteristics, and spatial resolution as fundamental image quality parameters were assessed. As dose metric the weighted cone‐beam dose index (CBDI(w)) was measured. Image quality was evaluated using standard quality assurance (QA) as well as anthropomorphic upper torso and breast phantoms. Both in‐house and manufacturer protocols for abdomen, pelvis, and breast imaging were examined. RESULTS: Using the in‐house protocols, the QA phantom scans showed altogether a high image quality, with high CT‐number accuracy (R (2 )> 0.97) and uniformity (<12 Hounsfield Unit (HU) cupping), reasonable noise and imaging fidelity, and good CNR at bone–tissue transitions of up to 28:1. Spatial resolution was strongly limited by geometric instabilities of the device. The breast phantom scans fulfilled clinical requirements, whereas the abdomen and pelvis scans showed severe artifacts, particularly at air/bone–tissue transitions. CONCLUSION: With the novel CBCT‐system, achieving a high image quality appears possible in principle. However, adaptations of the standard protocols, performance enhancements in image reconstruction referring to artifact reductions, as well as the extinction of geometric instabilities are imperative. John Wiley and Sons Inc. 2021-12-14 /pmc/articles/PMC8833290/ /pubmed/34905285 http://dx.doi.org/10.1002/acm2.13501 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Karius, Andre Karolczak, Marek Strnad, Vratislav Bert, Christoph Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy |
title | Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy |
title_full | Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy |
title_fullStr | Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy |
title_full_unstemmed | Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy |
title_short | Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy |
title_sort | technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based x‐ray system for brachytherapy |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833290/ https://www.ncbi.nlm.nih.gov/pubmed/34905285 http://dx.doi.org/10.1002/acm2.13501 |
work_keys_str_mv | AT kariusandre technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy AT karolczakmarek technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy AT strnadvratislav technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy AT bertchristoph technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy |