Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy

PURPOSE: A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was es...

Descripción completa

Detalles Bibliográficos
Autores principales: Karius, Andre, Karolczak, Marek, Strnad, Vratislav, Bert, Christoph
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833290/
https://www.ncbi.nlm.nih.gov/pubmed/34905285
http://dx.doi.org/10.1002/acm2.13501
_version_ 1784648899552608256
author Karius, Andre
Karolczak, Marek
Strnad, Vratislav
Bert, Christoph
author_facet Karius, Andre
Karolczak, Marek
Strnad, Vratislav
Bert, Christoph
author_sort Karius, Andre
collection PubMed
description PURPOSE: A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was essential to optimize both the imaging workflow and image quality for achieving the best possible clinical outcomes. We present the results of our investigations. METHODS: The novel CBCT‐system features a ring gantry with 121 cm clearance as well as a 43.2 × 43.2 cm(2) flat‐panel detector, and is controlled via a tablet‐personal computer (PC). For evaluating its imaging performance, the geometric reproducibility as well as imaging fidelity, computed tomography (CT)‐number accuracy, uniformity, contrast‐noise‐ratio (CNR), noise characteristics, and spatial resolution as fundamental image quality parameters were assessed. As dose metric the weighted cone‐beam dose index (CBDI(w)) was measured. Image quality was evaluated using standard quality assurance (QA) as well as anthropomorphic upper torso and breast phantoms. Both in‐house and manufacturer protocols for abdomen, pelvis, and breast imaging were examined. RESULTS: Using the in‐house protocols, the QA phantom scans showed altogether a high image quality, with high CT‐number accuracy (R (2 )> 0.97) and uniformity (<12 Hounsfield Unit (HU) cupping), reasonable noise and imaging fidelity, and good CNR at bone–tissue transitions of up to 28:1. Spatial resolution was strongly limited by geometric instabilities of the device. The breast phantom scans fulfilled clinical requirements, whereas the abdomen and pelvis scans showed severe artifacts, particularly at air/bone–tissue transitions. CONCLUSION: With the novel CBCT‐system, achieving a high image quality appears possible in principle. However, adaptations of the standard protocols, performance enhancements in image reconstruction referring to artifact reductions, as well as the extinction of geometric instabilities are imperative.
format Online
Article
Text
id pubmed-8833290
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-88332902022-02-14 Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy Karius, Andre Karolczak, Marek Strnad, Vratislav Bert, Christoph J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was essential to optimize both the imaging workflow and image quality for achieving the best possible clinical outcomes. We present the results of our investigations. METHODS: The novel CBCT‐system features a ring gantry with 121 cm clearance as well as a 43.2 × 43.2 cm(2) flat‐panel detector, and is controlled via a tablet‐personal computer (PC). For evaluating its imaging performance, the geometric reproducibility as well as imaging fidelity, computed tomography (CT)‐number accuracy, uniformity, contrast‐noise‐ratio (CNR), noise characteristics, and spatial resolution as fundamental image quality parameters were assessed. As dose metric the weighted cone‐beam dose index (CBDI(w)) was measured. Image quality was evaluated using standard quality assurance (QA) as well as anthropomorphic upper torso and breast phantoms. Both in‐house and manufacturer protocols for abdomen, pelvis, and breast imaging were examined. RESULTS: Using the in‐house protocols, the QA phantom scans showed altogether a high image quality, with high CT‐number accuracy (R (2 )> 0.97) and uniformity (<12 Hounsfield Unit (HU) cupping), reasonable noise and imaging fidelity, and good CNR at bone–tissue transitions of up to 28:1. Spatial resolution was strongly limited by geometric instabilities of the device. The breast phantom scans fulfilled clinical requirements, whereas the abdomen and pelvis scans showed severe artifacts, particularly at air/bone–tissue transitions. CONCLUSION: With the novel CBCT‐system, achieving a high image quality appears possible in principle. However, adaptations of the standard protocols, performance enhancements in image reconstruction referring to artifact reductions, as well as the extinction of geometric instabilities are imperative. John Wiley and Sons Inc. 2021-12-14 /pmc/articles/PMC8833290/ /pubmed/34905285 http://dx.doi.org/10.1002/acm2.13501 Text en © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Karius, Andre
Karolczak, Marek
Strnad, Vratislav
Bert, Christoph
Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
title Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
title_full Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
title_fullStr Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
title_full_unstemmed Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
title_short Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
title_sort technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based x‐ray system for brachytherapy
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8833290/
https://www.ncbi.nlm.nih.gov/pubmed/34905285
http://dx.doi.org/10.1002/acm2.13501
work_keys_str_mv AT kariusandre technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy
AT karolczakmarek technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy
AT strnadvratislav technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy
AT bertchristoph technicalevaluationoftheconebeamcomputedtomographyimagingperformanceofanovelmobilegantrybasedxraysystemforbrachytherapy