Constant Potential Coulometric Measurements with Ca(2+)-Selective Electrode: Analysis Using Calibration Plot vs. Analysis Using the Charge Curve Fitting

The possibility of analysis using charge curve fitting in constant potential coulometric mode instead of using a calibration plot is explored, for the first time. The results are compared with the analysis based on the use of a calibration plot. A Ca(2+) ion-selective electrode, with and without an...