Cargando…

Constant Potential Coulometric Measurements with Ca(2+)-Selective Electrode: Analysis Using Calibration Plot vs. Analysis Using the Charge Curve Fitting

The possibility of analysis using charge curve fitting in constant potential coulometric mode instead of using a calibration plot is explored, for the first time. The results are compared with the analysis based on the use of a calibration plot. A Ca(2+) ion-selective electrode, with and without an...

Descripción completa

Detalles Bibliográficos
Autores principales: Bondar, Anna, Mikhelson, Konstantin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8838552/
https://www.ncbi.nlm.nih.gov/pubmed/35161889
http://dx.doi.org/10.3390/s22031145

Ejemplares similares