Rapid Quantitative Analysis of IR Absorption Spectra for Trace Gas Detection by Artificial Neural Networks Trained with Synthetic Data

Infrared absorption spectroscopy is a widely used tool to quantify and monitor compositions of gases. The concentration information is often retrieved by fitting absorption profiles to the acquired spectra, utilizing spectroscopic databases. In complex gas matrices an expanded parameter space leads...

Descripción completa

Detalles Bibliográficos
Autores principales: Goldschmidt, Jens, Nitzsche, Leonard, Wolf, Sebastian, Lambrecht, Armin, Wöllenstein, Jürgen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8839408/
https://www.ncbi.nlm.nih.gov/pubmed/35161602
http://dx.doi.org/10.3390/s22030857
Descripción
Sumario:Infrared absorption spectroscopy is a widely used tool to quantify and monitor compositions of gases. The concentration information is often retrieved by fitting absorption profiles to the acquired spectra, utilizing spectroscopic databases. In complex gas matrices an expanded parameter space leads to long computation times of the fitting routines due to the increased number of spectral features that need to be computed for each iteration during the fit. This hinders the capability of real-time analysis of the gas matrix. Here, an artificial neural network (ANN) is employed for rapid prediction of gas concentrations in complex infrared absorption spectra composed of mixtures of CO and N(2)O. Experimental data is acquired with a mid-infrared dual frequency comb spectrometer. To circumvent the experimental collection of huge amounts of training data, the network is trained on synthetically generated spectra. The spectra are based on simulated absorption profiles making use of the HITRAN database. In addition, the spectrometer’s influence on the measured spectra is characterized and included in the synthetic training data generation. The ANN was tested on measured spectra and compared to a non-linear least squares fitting algorithm. An average evaluation time of 303 µs for a single measured spectrum was achieved. Coefficients of determination were 0.99997 for the predictions of N(2)O concentrations and 0.99987 for the predictions of CO concentrations, with uncertainties on the predicted concentrations between 0.04 and 0.18 ppm for 0 to 100 ppm N(2)O and between 0.05 and 0.18 ppm for 0 to 60 ppm CO.