BACKGROUND: Iliofemoral venous obstruction is recognized with increasing frequency as the underlying cause of lower extremity symptoms including edema, pain, skin changes, and, in advanced cases, ulceration. This study sought to evaluate the safety and effectiveness of the Abre venous self-expanding...

Descripción completa

Detalles Bibliográficos
Autores principales: Murphy, Erin, Gibson, Kathleen, Sapoval, Marc, Dexter, David J., Kolluri, Raghu, Razavi, Mahmood, Black, Stephen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8843393/
https://www.ncbi.nlm.nih.gov/pubmed/35105153
http://dx.doi.org/10.1161/CIRCINTERVENTIONS.121.010960
_version_ 1784651246579220480
author Murphy, Erin
Gibson, Kathleen
Sapoval, Marc
Dexter, David J.
Kolluri, Raghu
Razavi, Mahmood
Black, Stephen
author_facet Murphy, Erin
Gibson, Kathleen
Sapoval, Marc
Dexter, David J.
Kolluri, Raghu
Razavi, Mahmood
Black, Stephen
author_sort Murphy, Erin
collection PubMed
description BACKGROUND: Iliofemoral venous obstruction is recognized with increasing frequency as the underlying cause of lower extremity symptoms including edema, pain, skin changes, and, in advanced cases, ulceration. This study sought to evaluate the safety and effectiveness of the Abre venous self-expanding stent system for the treatment of symptomatic iliofemoral venous outflow obstruction. METHODS: The ABRE Study (A Multi-Center, Non-Randomized Study to Evaluate the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction) is a single-arm, multicenter, prospective study that included 200 subjects from 24 global sites. The primary end points were 12-month primary patency and major adverse events within 30 days. Secondary end points included lesion and procedure success, primary-assisted and secondary patency, major adverse events, stent migration, stent fracture, and quality of life changes. End point-related adverse events and imaging studies were adjudicated by independent clinical events committee and core laboratories, respectively. RESULTS: Venous obstruction cause was classified as acute deep vein thrombosis (16.5%, 33/200), post-thrombotic syndrome (47.5%, 95/200), or nonthrombotic iliac vein lesion (36.0%, 72/200). The common iliac and external iliac veins were stented in 96.0% (192/200), 80.5% (161/200) of subjects, respectively. Stent implant into the common femoral vein was required in 44.0% (88/200). Primary patency at 12 months was 88.0% (162/184). Four (2.0%) major adverse events occurred within 30 days. Twelve-month primary-assisted and secondary patency were 91.8% (169/184) and 92.9% (171/184), respectively. No stent fractures or migrations were reported. Mean target limb Villalta score decreased from 11.2±5.6 at baseline to 4.1±4.8 at 12 months, and the mean target limb revised Venous Clinical Severity Score decreased from 8.8±4.7 at baseline to 4.3±3.6 at 12 months. Clinically meaningful improvements in quality of life and venous functional assessment scores from baseline were demonstrated through 12 months in all measures. CONCLUSIONS: Symptomatic iliofemoral venous obstruction can be successfully treated with an Abre venous stent. Study outcomes demonstrated a high patency rate with a good safety profile. Patients demonstrated a significant reduction in clinical symptoms and improvement in quality of life that was maintained through 12-month follow-up. REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT03038438.
format Online
Article
Text
id pubmed-8843393
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-88433932022-02-17 Pivotal Study Evaluating the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction Murphy, Erin Gibson, Kathleen Sapoval, Marc Dexter, David J. Kolluri, Raghu Razavi, Mahmood Black, Stephen Circ Cardiovasc Interv Original Articles BACKGROUND: Iliofemoral venous obstruction is recognized with increasing frequency as the underlying cause of lower extremity symptoms including edema, pain, skin changes, and, in advanced cases, ulceration. This study sought to evaluate the safety and effectiveness of the Abre venous self-expanding stent system for the treatment of symptomatic iliofemoral venous outflow obstruction. METHODS: The ABRE Study (A Multi-Center, Non-Randomized Study to Evaluate the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction) is a single-arm, multicenter, prospective study that included 200 subjects from 24 global sites. The primary end points were 12-month primary patency and major adverse events within 30 days. Secondary end points included lesion and procedure success, primary-assisted and secondary patency, major adverse events, stent migration, stent fracture, and quality of life changes. End point-related adverse events and imaging studies were adjudicated by independent clinical events committee and core laboratories, respectively. RESULTS: Venous obstruction cause was classified as acute deep vein thrombosis (16.5%, 33/200), post-thrombotic syndrome (47.5%, 95/200), or nonthrombotic iliac vein lesion (36.0%, 72/200). The common iliac and external iliac veins were stented in 96.0% (192/200), 80.5% (161/200) of subjects, respectively. Stent implant into the common femoral vein was required in 44.0% (88/200). Primary patency at 12 months was 88.0% (162/184). Four (2.0%) major adverse events occurred within 30 days. Twelve-month primary-assisted and secondary patency were 91.8% (169/184) and 92.9% (171/184), respectively. No stent fractures or migrations were reported. Mean target limb Villalta score decreased from 11.2±5.6 at baseline to 4.1±4.8 at 12 months, and the mean target limb revised Venous Clinical Severity Score decreased from 8.8±4.7 at baseline to 4.3±3.6 at 12 months. Clinically meaningful improvements in quality of life and venous functional assessment scores from baseline were demonstrated through 12 months in all measures. CONCLUSIONS: Symptomatic iliofemoral venous obstruction can be successfully treated with an Abre venous stent. Study outcomes demonstrated a high patency rate with a good safety profile. Patients demonstrated a significant reduction in clinical symptoms and improvement in quality of life that was maintained through 12-month follow-up. REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT03038438. Lippincott Williams & Wilkins 2022-02-02 /pmc/articles/PMC8843393/ /pubmed/35105153 http://dx.doi.org/10.1161/CIRCINTERVENTIONS.121.010960 Text en © 2022 The Authors. https://creativecommons.org/licenses/by-nc-nd/4.0/Circulation: Cardiovascular Interventions is published on behalf of the American Heart Association, Inc., by Wolters Kluwer Health, Inc. This is an open access article under the terms of the Creative Commons Attribution Non-Commercial-NoDerivs (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use, distribution, and reproduction in any medium, provided that the original work is properly cited, the use is noncommercial, and no modifications or adaptations are made.
spellingShingle Original Articles
Murphy, Erin
Gibson, Kathleen
Sapoval, Marc
Dexter, David J.
Kolluri, Raghu
Razavi, Mahmood
Black, Stephen
Pivotal Study Evaluating the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction
title Pivotal Study Evaluating the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction
title_full Pivotal Study Evaluating the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction
title_fullStr Pivotal Study Evaluating the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction
title_full_unstemmed Pivotal Study Evaluating the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction
title_short Pivotal Study Evaluating the Safety and Effectiveness of the Abre Venous Self-Expanding Stent System in Patients With Symptomatic Iliofemoral Venous Outflow Obstruction
title_sort pivotal study evaluating the safety and effectiveness of the abre venous self-expanding stent system in patients with symptomatic iliofemoral venous outflow obstruction
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8843393/
https://www.ncbi.nlm.nih.gov/pubmed/35105153
http://dx.doi.org/10.1161/CIRCINTERVENTIONS.121.010960
work_keys_str_mv AT murphyerin pivotalstudyevaluatingthesafetyandeffectivenessoftheabrevenousselfexpandingstentsysteminpatientswithsymptomaticiliofemoralvenousoutflowobstruction
AT gibsonkathleen pivotalstudyevaluatingthesafetyandeffectivenessoftheabrevenousselfexpandingstentsysteminpatientswithsymptomaticiliofemoralvenousoutflowobstruction
AT sapovalmarc pivotalstudyevaluatingthesafetyandeffectivenessoftheabrevenousselfexpandingstentsysteminpatientswithsymptomaticiliofemoralvenousoutflowobstruction
AT dexterdavidj pivotalstudyevaluatingthesafetyandeffectivenessoftheabrevenousselfexpandingstentsysteminpatientswithsymptomaticiliofemoralvenousoutflowobstruction
AT kolluriraghu pivotalstudyevaluatingthesafetyandeffectivenessoftheabrevenousselfexpandingstentsysteminpatientswithsymptomaticiliofemoralvenousoutflowobstruction
AT razavimahmood pivotalstudyevaluatingthesafetyandeffectivenessoftheabrevenousselfexpandingstentsysteminpatientswithsymptomaticiliofemoralvenousoutflowobstruction
AT blackstephen pivotalstudyevaluatingthesafetyandeffectivenessoftheabrevenousselfexpandingstentsysteminpatientswithsymptomaticiliofemoralvenousoutflowobstruction