Cargando…

Pacemaker lead rupture in a patient with subacute endocarditis: a case report

BACKGROUND: Cardiac implantable electronic device (CIED)-related infections are associated with severe morbidity and mortality. Few cases have previously documented both lead endocarditis and lead rupture simultaneously. CASE SUMMARY: We describe the case of a 73-year-old man with a dual-chamber pac...

Descripción completa

Detalles Bibliográficos
Autores principales: Caratti di Lanzacco, Lorenzo, Wauters, Aurélien
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8843865/
https://www.ncbi.nlm.nih.gov/pubmed/35174311
http://dx.doi.org/10.1093/ehjcr/ytac054
Descripción
Sumario:BACKGROUND: Cardiac implantable electronic device (CIED)-related infections are associated with severe morbidity and mortality. Few cases have previously documented both lead endocarditis and lead rupture simultaneously. CASE SUMMARY: We describe the case of a 73-year-old man with a dual-chamber pacemaker presenting with subacute endocarditis and recurrent cholangitis. A few months prior, the patient was diagnosed with localized colon cancer and Streptococcus sanguinis lead endocarditis based on nuclear imaging. He was given prolonged antibiotic therapy and lead explantation was to be performed after sigmoidectomy. During the following weeks, his condition worsened and he was readmitted for biliary sepsis. A chest X-ray revealed, incidentally, a complete ventricular lead rupture. Pacemaker electrogram showed ventricular undersensing, loss of ventricular capture, and high impedance. As his health declined, removal of the pacemaker was deemed unreasonable and the patient died of biliary sepsis in the next few weeks. DISCUSSION: We describe the case of an asymptomatic intracardiac lead fracture in the setting of colon cancer and a medically managed Streptococcus lead infection. As this complication occurred during lead infection, bacterial damage may have weakened the lead over time. As illustrated by the patient’s outcomes, long-term antibiotic therapy should only be used in cases unsuitable for device removal. Complete hardware removal remains the first-line therapy in patients with CIED-related infections.