Migration characteristics of the Corail hydroxyapatite-coated femoral stem—a retrospective clinical evaluation and migration measurement with EBRA
PURPOSE: Uncemented stem migration analysis by EBRA-FCA (Einzel-Bild-Roentgen Analyse, Femoral Component Analysis) has been seen to be a good predictive indicator for early implant failure. In this study, we investigated the migration behavior of a cementless press-fit stem after two years follow-up...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8843908/ https://www.ncbi.nlm.nih.gov/pubmed/33999259 http://dx.doi.org/10.1007/s00402-021-03926-9 |
Sumario: | PURPOSE: Uncemented stem migration analysis by EBRA-FCA (Einzel-Bild-Roentgen Analyse, Femoral Component Analysis) has been seen to be a good predictive indicator for early implant failure. In this study, we investigated the migration behavior of a cementless press-fit stem after two years follow-up. Stem type and postoperative gap between collar and femur were evaluated as a risk factor. METHODS: Applying a retrospective study design, we reviewed all consecutive patients who between 2013 and 2017 received a cementless press-fit Corail stem (DePuy Orthopaedics Inc., Warsaw, IN, USA) at our Department. We reviewed medical histories and performed radiological measurements using EBRA-FCA software. RESULTS: A total of 109 stems in 105 patients (female: 60; male: 45) fulfilled our inclusion criteria. Mean age at surgery was 67.8 (range, 21.6–90.5) years. EBRA migration analysis showed a mean subsidence of 1.8 mm (range, 0.0–12.1) at final follow-up. At 18 months mean subsidence of collared stems was significantly lower than in the collarless group [1.3 mm (range, 0.0–7.6) vs. 3.2 mm (range, 0.5–10.7), p = 0.0104]. Collared stems resting on the femoral cut presented a tendency to less subsidence than did collared stems showing a postoperative gap between collar and femur (1.3 vs. 2.0 mm) without finding statistical significance (p > 0.05). CONCLUSIONS: Low subsidence and the migration pattern of the cementless press-fit stem may predict a good long-term result. Collared stems investigated in our study provide good stability and are able to prevent significant subsidence. Trial registration number and date of registration: Number: 20181024-1875; Date: 2018-10-24 |
---|