Cargando…
Forehead, Temple and Wrist Temperature Assessment of Ethnic Groups using Infrared Technology
Non-contact infrared sensors are widely used as a diagnostic tool for elevated body temperature during initial screening for coronaviruses. The aim of this study was to investigate the thermal differences at three anatomical points: temple, forehead, and wrist, in the initial screening for temperatu...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
IPEM. Published by Elsevier Ltd.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8848569/ https://www.ncbi.nlm.nih.gov/pubmed/35346438 http://dx.doi.org/10.1016/j.medengphy.2022.103777 |
Sumario: | Non-contact infrared sensors are widely used as a diagnostic tool for elevated body temperature during initial screening for coronaviruses. The aim of this study was to investigate the thermal differences at three anatomical points: temple, forehead, and wrist, in the initial screening for temperature indicative of febrile and non-febrile states in skin pigmentation variations in Black, Half-Black and Caucasian skins, correlated with height and weight variables. Temperatures were obtained by means of an infrared thermometer in 289 volunteers with mean age of 18.30 ± 0.76, in a controlled environment according to Singapore Standard, SS582 part 1 and 2, normative standard IEC 80601-2-59, with standard technical protocols established by the International Organization for Standardization, ISO / TR 13154. The data were processed in MATLAB® R2021a, and data normality verified by Kolmogorov-Smirnov test, non-parametric data paired between temple / forehead / wrist were compared using the Wilcoxon signed-rank test. The results show different median temperatures in these anatomical regions, 37.2°C at the temple, 36.8°C at the forehead and 36.4°C at the wrist. As the temple region presents a temperature higher than the other investigated regions and, therefore, close to the core temperature, it should be considered for the initial screening of SARS-CoV-2 when using non-contact infrared thermometers. Furthermore, no significant changes were found due to variation in skin tone, height, or weight. |
---|