Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports

[Image: see text] Polyamide noncoated device for adsorption-based microextraction (PANDA microextraction) is a brand new, easy to prepare, environmentally friendly, inexpensive, and efficient sample preparation method created entirely with the use of 3D printing. The proposed method is based on the...

Descripción completa

Detalles Bibliográficos
Autores principales: Kołodziej, Dominika, Sobczak, Łukasz, Goryński, Krzysztof
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2022
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8851416/
https://www.ncbi.nlm.nih.gov/pubmed/35113529
http://dx.doi.org/10.1021/acs.analchem.1c03672
_version_ 1784652815724969984
author Kołodziej, Dominika
Sobczak, Łukasz
Goryński, Krzysztof
author_facet Kołodziej, Dominika
Sobczak, Łukasz
Goryński, Krzysztof
author_sort Kołodziej, Dominika
collection PubMed
description [Image: see text] Polyamide noncoated device for adsorption-based microextraction (PANDA microextraction) is a brand new, easy to prepare, environmentally friendly, inexpensive, and efficient sample preparation method created entirely with the use of 3D printing. The proposed method is based on the extractive proprieties of the unmodified polyamide and carbon fiber blends and is compared with the highly selective thin-film microextraction (TFME). In addition, 3D printing was used to simplify the process of TFME. Prototype sample preparation devices were evaluated by the extraction of oral fluid spiked with 38 small molecules with diverse chemical natures, such as lipophilicity in the log P range of 0.2–7.2. The samples were analyzed by high-performance liquid chromatography coupled with tandem mass spectrometry. The results indicate that chemically and thermally resistant 3D printed supports can be successfully used as a cost-saving, environmentally friendly solution for the preparation of TFME devices, alternative to the conventional metal supports, with only marginal differences in the extraction yield (mean = 4.0%, median = 1.8%, range = 0.0–22.3%, n = 38). Even more remarkably, in some cases, the newly proposed PANDA microextraction method exceeded the reference TFME in terms of the extraction efficacy and offered excellent sample cleanup as favorable matrix effects were observed (mean = −8.5%, median = 7.5%, range = −34.7–20.0%, n = 20). This innovative approach paves the road to the simplified sample preparation with the use of emerging extractive 3D printing polymers.
format Online
Article
Text
id pubmed-8851416
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher American Chemical Society
record_format MEDLINE/PubMed
spelling pubmed-88514162022-02-18 Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports Kołodziej, Dominika Sobczak, Łukasz Goryński, Krzysztof Anal Chem [Image: see text] Polyamide noncoated device for adsorption-based microextraction (PANDA microextraction) is a brand new, easy to prepare, environmentally friendly, inexpensive, and efficient sample preparation method created entirely with the use of 3D printing. The proposed method is based on the extractive proprieties of the unmodified polyamide and carbon fiber blends and is compared with the highly selective thin-film microextraction (TFME). In addition, 3D printing was used to simplify the process of TFME. Prototype sample preparation devices were evaluated by the extraction of oral fluid spiked with 38 small molecules with diverse chemical natures, such as lipophilicity in the log P range of 0.2–7.2. The samples were analyzed by high-performance liquid chromatography coupled with tandem mass spectrometry. The results indicate that chemically and thermally resistant 3D printed supports can be successfully used as a cost-saving, environmentally friendly solution for the preparation of TFME devices, alternative to the conventional metal supports, with only marginal differences in the extraction yield (mean = 4.0%, median = 1.8%, range = 0.0–22.3%, n = 38). Even more remarkably, in some cases, the newly proposed PANDA microextraction method exceeded the reference TFME in terms of the extraction efficacy and offered excellent sample cleanup as favorable matrix effects were observed (mean = −8.5%, median = 7.5%, range = −34.7–20.0%, n = 20). This innovative approach paves the road to the simplified sample preparation with the use of emerging extractive 3D printing polymers. American Chemical Society 2022-02-03 2022-02-15 /pmc/articles/PMC8851416/ /pubmed/35113529 http://dx.doi.org/10.1021/acs.analchem.1c03672 Text en © 2022 The Authors. Published by American Chemical Society https://creativecommons.org/licenses/by/4.0/Permits the broadest form of re-use including for commercial purposes, provided that author attribution and integrity are maintained (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Kołodziej, Dominika
Sobczak, Łukasz
Goryński, Krzysztof
Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports
title Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports
title_full Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports
title_fullStr Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports
title_full_unstemmed Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports
title_short Polyamide Noncoated Device for Adsorption-Based Microextraction and Novel 3D Printed Thin-Film Microextraction Supports
title_sort polyamide noncoated device for adsorption-based microextraction and novel 3d printed thin-film microextraction supports
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8851416/
https://www.ncbi.nlm.nih.gov/pubmed/35113529
http://dx.doi.org/10.1021/acs.analchem.1c03672
work_keys_str_mv AT kołodziejdominika polyamidenoncoateddeviceforadsorptionbasedmicroextractionandnovel3dprintedthinfilmmicroextractionsupports
AT sobczakłukasz polyamidenoncoateddeviceforadsorptionbasedmicroextractionandnovel3dprintedthinfilmmicroextractionsupports
AT gorynskikrzysztof polyamidenoncoateddeviceforadsorptionbasedmicroextractionandnovel3dprintedthinfilmmicroextractionsupports