Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy

INTRODUCTION: The accuracy of dose calculation algorithms depends on the electron density and computed tomography (CT) number of medium scanned. Our study aimed to verify the impact of different CT scanning protocols on Hounsfield unit (HU) and effect on dose calculation algorithms. MATERIALS AND ME...

Descripción completa

Detalles Bibliográficos
Autores principales: Saini, Amit, Pandey, V P, Kumar, Pankaj, Singh, Avtar, Pasricha, Rajesh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8853450/
https://www.ncbi.nlm.nih.gov/pubmed/35261502
http://dx.doi.org/10.4103/jmp.JMP_124_20
_version_ 1784653230598258688
author Saini, Amit
Pandey, V P
Kumar, Pankaj
Singh, Avtar
Pasricha, Rajesh
author_facet Saini, Amit
Pandey, V P
Kumar, Pankaj
Singh, Avtar
Pasricha, Rajesh
author_sort Saini, Amit
collection PubMed
description INTRODUCTION: The accuracy of dose calculation algorithms depends on the electron density and computed tomography (CT) number of medium scanned. Our study aimed to verify the impact of different CT scanning protocols on Hounsfield unit (HU) and effect on dose calculation algorithms. MATERIALS AND METHODS: CIRS thorax phantom with different density material plugs was scanned at varying tube voltages from CT scanner and HU values were measured in treatment planning system (TPS). Calibration curves of electron density at different tube voltages were plotted and used for dose calculation with different calculation algorithms at varying high energy megavoltage photon energies. RESULTS: Insignificant difference is obtained in electron density curves plotted at different tube voltages. The mean variation in HU values was found at different tube voltages for bone, lung, and water are 896.75 (standard deviation [SD] 122.88), −799.25 (SD 5.74), and −17.5 (SD 0.57), respectively. The estimated P values for change in HU values were 0.089, 0.258, and 0.121 for bone, lung, and water, respectively. Pencil beam (PB) convolution and collapsed cone algorithms show no significant dose difference, i.e., <1% variation and Monte Carlo (MC) shows maximum dose difference up to 1.4%. CONCLUSION: Third-generation algorithms such as MC shows dependence on varying tube voltages in dose calculation. Calibration curves plotted at different kVp in TPS advised to be chosen wisely to avoid any dosimetric errors in different medium.
format Online
Article
Text
id pubmed-8853450
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-88534502022-03-07 Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy Saini, Amit Pandey, V P Kumar, Pankaj Singh, Avtar Pasricha, Rajesh J Med Phys Original Article INTRODUCTION: The accuracy of dose calculation algorithms depends on the electron density and computed tomography (CT) number of medium scanned. Our study aimed to verify the impact of different CT scanning protocols on Hounsfield unit (HU) and effect on dose calculation algorithms. MATERIALS AND METHODS: CIRS thorax phantom with different density material plugs was scanned at varying tube voltages from CT scanner and HU values were measured in treatment planning system (TPS). Calibration curves of electron density at different tube voltages were plotted and used for dose calculation with different calculation algorithms at varying high energy megavoltage photon energies. RESULTS: Insignificant difference is obtained in electron density curves plotted at different tube voltages. The mean variation in HU values was found at different tube voltages for bone, lung, and water are 896.75 (standard deviation [SD] 122.88), −799.25 (SD 5.74), and −17.5 (SD 0.57), respectively. The estimated P values for change in HU values were 0.089, 0.258, and 0.121 for bone, lung, and water, respectively. Pencil beam (PB) convolution and collapsed cone algorithms show no significant dose difference, i.e., <1% variation and Monte Carlo (MC) shows maximum dose difference up to 1.4%. CONCLUSION: Third-generation algorithms such as MC shows dependence on varying tube voltages in dose calculation. Calibration curves plotted at different kVp in TPS advised to be chosen wisely to avoid any dosimetric errors in different medium. Wolters Kluwer - Medknow 2021 2021-12-31 /pmc/articles/PMC8853450/ /pubmed/35261502 http://dx.doi.org/10.4103/jmp.JMP_124_20 Text en Copyright: © 2021 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Saini, Amit
Pandey, V P
Kumar, Pankaj
Singh, Avtar
Pasricha, Rajesh
Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy
title Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy
title_full Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy
title_fullStr Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy
title_full_unstemmed Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy
title_short Investigation of tube voltage dependence on CT number and its effect on dose calculation algorithms using thorax phantom in Monaco treatment planning system for external beam radiation therapy
title_sort investigation of tube voltage dependence on ct number and its effect on dose calculation algorithms using thorax phantom in monaco treatment planning system for external beam radiation therapy
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8853450/
https://www.ncbi.nlm.nih.gov/pubmed/35261502
http://dx.doi.org/10.4103/jmp.JMP_124_20
work_keys_str_mv AT sainiamit investigationoftubevoltagedependenceonctnumberanditseffectondosecalculationalgorithmsusingthoraxphantominmonacotreatmentplanningsystemforexternalbeamradiationtherapy
AT pandeyvp investigationoftubevoltagedependenceonctnumberanditseffectondosecalculationalgorithmsusingthoraxphantominmonacotreatmentplanningsystemforexternalbeamradiationtherapy
AT kumarpankaj investigationoftubevoltagedependenceonctnumberanditseffectondosecalculationalgorithmsusingthoraxphantominmonacotreatmentplanningsystemforexternalbeamradiationtherapy
AT singhavtar investigationoftubevoltagedependenceonctnumberanditseffectondosecalculationalgorithmsusingthoraxphantominmonacotreatmentplanningsystemforexternalbeamradiationtherapy
AT pasricharajesh investigationoftubevoltagedependenceonctnumberanditseffectondosecalculationalgorithmsusingthoraxphantominmonacotreatmentplanningsystemforexternalbeamradiationtherapy