Cargando…

Automation of DVH Constraint Checks and Physics Quality Control Review Improves Patient Safety in Radiotherapy

This study investigates whether patient safety can be enhanced by the implementation of an automated electronic checklist (PlanCheck) for physics quality control review (QCR) of radiotherapy photon plans. PlanCheck evaluates both technical aspects and DVH constraints. Three hundred and thirty-one co...

Descripción completa

Detalles Bibliográficos
Autores principales: Wahlstedt, Isak, Jensen, Nikolaj
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8853455/
https://www.ncbi.nlm.nih.gov/pubmed/35261505
http://dx.doi.org/10.4103/jmp.JMP_23_21
Descripción
Sumario:This study investigates whether patient safety can be enhanced by the implementation of an automated electronic checklist (PlanCheck) for physics quality control review (QCR) of radiotherapy photon plans. PlanCheck evaluates both technical aspects and DVH constraints. Three hundred and thirty-one consecutively approved radiotherapy plans previously reviewed with manual QCR were retrospectively checked with PlanCheck. Four hundred and thirty-three (3.4%) of the 12783 automated technical checks executed in the 331 plans yielded an error. All errors were scored using the severity rating from the American Association of Physicists in Medicine TG-100 report. Nineteen of these errors (4%) either could have affected or affected target dose (severity 5+) implicating a maximum dose difference to the target or a critical organ at risk of 0.5% to 10% and 3 errors could have resulted in stereotactic brain treatments being delivered to the wrong location (severity 10). Forty-seven breast cancer plans were retrospectively subjected to automated DVH check, 10 undocumented dose constraint violations were found. PlanCheck has been shown to reduce errors in manually reviewed radiotherapy plans and thus to enhance patient safety.