Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery
AIM: To determine the multileaf collimator positional error (MLC-PE) during volumetric modulated arc therapy (VMAT) delivery by studying the time-dependent MLC velocity in mathematically derivable trajectories such as straight line and conic sections. MATERIALS AND METHODS: VMAT delivery is planned...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8853456/ https://www.ncbi.nlm.nih.gov/pubmed/35261498 http://dx.doi.org/10.4103/jmp.JMP_18_21 |
_version_ | 1784653232069410816 |
---|---|
author | Roy, Soumya Sarkar, Biplab Pradhan, Anirudh |
author_facet | Roy, Soumya Sarkar, Biplab Pradhan, Anirudh |
author_sort | Roy, Soumya |
collection | PubMed |
description | AIM: To determine the multileaf collimator positional error (MLC-PE) during volumetric modulated arc therapy (VMAT) delivery by studying the time-dependent MLC velocity in mathematically derivable trajectories such as straight line and conic sections. MATERIALS AND METHODS: VMAT delivery is planned in a way that MLCs are moving in a locus which can be defined by mathematical functions such as linear, parabolic, or circular velocity (PV or CV). The VMAT delivery was interrupted either once or multiple times during the delivery and projection images of the same were acquired in electronic portal imaging device. MLC-PE was then analyzed as a function of dose rate (DR), and MLC speed (SP) and number of interruptions in treatment delivery. In VMAT delivery with linear MLC motion, the delivery was interrupted either once (linear motion single interruption) or multiple (three) times (linear motion multiple interruptions). For PV and CV MLC velocity, the MLC motions are interrupted multiple times. RESULTS: The maximum individual error obtained (DR of 35 MU/min, SP of 2.0 cm/s) was 1.96 ± 0.1 mm. Only 4.4% of MLCs showed ≥ ±1 mm positional error. When the treatment delivery is interrupted multiple times in VMAT delivery, the influence of interruption in MLC-PE overwhelmed the influence by DR and SP. For a sub-group analysis of independent and dependent variables, the mean MLC-PE was 0.18 ± 0.4 mm 0.19 ± 0.42 mm, respectively. CONCLUSION: Determination of MLC-PEs using a mathematical function without approximation indicates that MLC-PE is not a function of MLC speed. In less than 5% of the studied scenarios, the MLC-PE exceeds its tolerance value (±1 mm). The MLC-PE is significantly less in modern machines due to advancements in the delivery mechanism. |
format | Online Article Text |
id | pubmed-8853456 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-88534562022-03-07 Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery Roy, Soumya Sarkar, Biplab Pradhan, Anirudh J Med Phys Original Article AIM: To determine the multileaf collimator positional error (MLC-PE) during volumetric modulated arc therapy (VMAT) delivery by studying the time-dependent MLC velocity in mathematically derivable trajectories such as straight line and conic sections. MATERIALS AND METHODS: VMAT delivery is planned in a way that MLCs are moving in a locus which can be defined by mathematical functions such as linear, parabolic, or circular velocity (PV or CV). The VMAT delivery was interrupted either once or multiple times during the delivery and projection images of the same were acquired in electronic portal imaging device. MLC-PE was then analyzed as a function of dose rate (DR), and MLC speed (SP) and number of interruptions in treatment delivery. In VMAT delivery with linear MLC motion, the delivery was interrupted either once (linear motion single interruption) or multiple (three) times (linear motion multiple interruptions). For PV and CV MLC velocity, the MLC motions are interrupted multiple times. RESULTS: The maximum individual error obtained (DR of 35 MU/min, SP of 2.0 cm/s) was 1.96 ± 0.1 mm. Only 4.4% of MLCs showed ≥ ±1 mm positional error. When the treatment delivery is interrupted multiple times in VMAT delivery, the influence of interruption in MLC-PE overwhelmed the influence by DR and SP. For a sub-group analysis of independent and dependent variables, the mean MLC-PE was 0.18 ± 0.4 mm 0.19 ± 0.42 mm, respectively. CONCLUSION: Determination of MLC-PEs using a mathematical function without approximation indicates that MLC-PE is not a function of MLC speed. In less than 5% of the studied scenarios, the MLC-PE exceeds its tolerance value (±1 mm). The MLC-PE is significantly less in modern machines due to advancements in the delivery mechanism. Wolters Kluwer - Medknow 2021 2021-12-31 /pmc/articles/PMC8853456/ /pubmed/35261498 http://dx.doi.org/10.4103/jmp.JMP_18_21 Text en Copyright: © 2021 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Roy, Soumya Sarkar, Biplab Pradhan, Anirudh Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery |
title | Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery |
title_full | Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery |
title_fullStr | Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery |
title_full_unstemmed | Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery |
title_short | Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery |
title_sort | determination of multileaf collimator positional errors as a function of dose rate, speed, and delivery interruption for volumetric-modulated arc therapy delivery |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8853456/ https://www.ncbi.nlm.nih.gov/pubmed/35261498 http://dx.doi.org/10.4103/jmp.JMP_18_21 |
work_keys_str_mv | AT roysoumya determinationofmultileafcollimatorpositionalerrorsasafunctionofdoseratespeedanddeliveryinterruptionforvolumetricmodulatedarctherapydelivery AT sarkarbiplab determinationofmultileafcollimatorpositionalerrorsasafunctionofdoseratespeedanddeliveryinterruptionforvolumetricmodulatedarctherapydelivery AT pradhananirudh determinationofmultileafcollimatorpositionalerrorsasafunctionofdoseratespeedanddeliveryinterruptionforvolumetricmodulatedarctherapydelivery |