Cargando…
Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
Blood gas, electrolyte, glucose, and lactate level measurement have an immediate and critical impact on patient care. We evaluated the performance of i-SmartCare 10 (i-SENS Inc., Seoul, Korea) and conducted a method comparison study of five point-of-care (POC) analyzers with i-SmartCare 10 as the co...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Society for Laboratory Medicine
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8859550/ https://www.ncbi.nlm.nih.gov/pubmed/35177567 http://dx.doi.org/10.3343/alm.2022.42.4.467 |
Sumario: | Blood gas, electrolyte, glucose, and lactate level measurement have an immediate and critical impact on patient care. We evaluated the performance of i-SmartCare 10 (i-SENS Inc., Seoul, Korea) and conducted a method comparison study of five point-of-care (POC) analyzers with i-SmartCare 10 as the comparator, according to the CLSI guidelines. Ten analytes (pH, pCO(2), pO(2), Na(+), K(+), Cl(-), iCa(2+), glucose, lactate, and Hct) were tested on six analyzers: i-SmartCare 10, ABL90 FLEX PLUS (Radiometer Medical ApS, Copenhagen, Denmark), i-Stat (Abbott Point of Care Inc., Princeton, NJ, USA), RapidLab 1265 (Siemens Healthcare Diagnostics Inc., Tarrytown, NY, USA), Stat Profile pHOx Ultra (Nova Biomedical, Waltham, MA, USA), and Gem Premier 5000 (Instrumentation Laboratory, Bedford, MA, USA). The total imprecision and linearity (r(2)>0.99) were excellent, except for a few analytes that narrowly escaped the preset criteria. Interference was noted for Na(+) in the presence of a high K(+) level and for iCa(2+) in the presence of high K(+) and Mg(2+) levels. Forty of 48 items demonstrated either a proportional or systematic difference in regression analysis; the relative mean difference (%) of 14/48 items escaped the allowable total error in the difference plot analysis. i-SmartCare 10 shows acceptable performance, and using a single POC blood gas analyzer is recommended for monitoring. |
---|