Cargando…

Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers

Blood gas, electrolyte, glucose, and lactate level measurement have an immediate and critical impact on patient care. We evaluated the performance of i-SmartCare 10 (i-SENS Inc., Seoul, Korea) and conducted a method comparison study of five point-of-care (POC) analyzers with i-SmartCare 10 as the co...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Sang-Mi, Park, Hyung-Doo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society for Laboratory Medicine 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8859550/
https://www.ncbi.nlm.nih.gov/pubmed/35177567
http://dx.doi.org/10.3343/alm.2022.42.4.467
_version_ 1784654485967077376
author Kim, Sang-Mi
Park, Hyung-Doo
author_facet Kim, Sang-Mi
Park, Hyung-Doo
author_sort Kim, Sang-Mi
collection PubMed
description Blood gas, electrolyte, glucose, and lactate level measurement have an immediate and critical impact on patient care. We evaluated the performance of i-SmartCare 10 (i-SENS Inc., Seoul, Korea) and conducted a method comparison study of five point-of-care (POC) analyzers with i-SmartCare 10 as the comparator, according to the CLSI guidelines. Ten analytes (pH, pCO(2), pO(2), Na(+), K(+), Cl(-), iCa(2+), glucose, lactate, and Hct) were tested on six analyzers: i-SmartCare 10, ABL90 FLEX PLUS (Radiometer Medical ApS, Copenhagen, Denmark), i-Stat (Abbott Point of Care Inc., Princeton, NJ, USA), RapidLab 1265 (Siemens Healthcare Diagnostics Inc., Tarrytown, NY, USA), Stat Profile pHOx Ultra (Nova Biomedical, Waltham, MA, USA), and Gem Premier 5000 (Instrumentation Laboratory, Bedford, MA, USA). The total imprecision and linearity (r(2)>0.99) were excellent, except for a few analytes that narrowly escaped the preset criteria. Interference was noted for Na(+) in the presence of a high K(+) level and for iCa(2+) in the presence of high K(+) and Mg(2+) levels. Forty of 48 items demonstrated either a proportional or systematic difference in regression analysis; the relative mean difference (%) of 14/48 items escaped the allowable total error in the difference plot analysis. i-SmartCare 10 shows acceptable performance, and using a single POC blood gas analyzer is recommended for monitoring.
format Online
Article
Text
id pubmed-8859550
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Korean Society for Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-88595502022-07-01 Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers Kim, Sang-Mi Park, Hyung-Doo Ann Lab Med Brief Communications Blood gas, electrolyte, glucose, and lactate level measurement have an immediate and critical impact on patient care. We evaluated the performance of i-SmartCare 10 (i-SENS Inc., Seoul, Korea) and conducted a method comparison study of five point-of-care (POC) analyzers with i-SmartCare 10 as the comparator, according to the CLSI guidelines. Ten analytes (pH, pCO(2), pO(2), Na(+), K(+), Cl(-), iCa(2+), glucose, lactate, and Hct) were tested on six analyzers: i-SmartCare 10, ABL90 FLEX PLUS (Radiometer Medical ApS, Copenhagen, Denmark), i-Stat (Abbott Point of Care Inc., Princeton, NJ, USA), RapidLab 1265 (Siemens Healthcare Diagnostics Inc., Tarrytown, NY, USA), Stat Profile pHOx Ultra (Nova Biomedical, Waltham, MA, USA), and Gem Premier 5000 (Instrumentation Laboratory, Bedford, MA, USA). The total imprecision and linearity (r(2)>0.99) were excellent, except for a few analytes that narrowly escaped the preset criteria. Interference was noted for Na(+) in the presence of a high K(+) level and for iCa(2+) in the presence of high K(+) and Mg(2+) levels. Forty of 48 items demonstrated either a proportional or systematic difference in regression analysis; the relative mean difference (%) of 14/48 items escaped the allowable total error in the difference plot analysis. i-SmartCare 10 shows acceptable performance, and using a single POC blood gas analyzer is recommended for monitoring. Korean Society for Laboratory Medicine 2022-07-01 2022-07-01 /pmc/articles/PMC8859550/ /pubmed/35177567 http://dx.doi.org/10.3343/alm.2022.42.4.467 Text en © Korean Society for Laboratory Medicine https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0 (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Brief Communications
Kim, Sang-Mi
Park, Hyung-Doo
Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
title Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
title_full Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
title_fullStr Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
title_full_unstemmed Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
title_short Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
title_sort performance evaluation of the i-smartcare 10 analyzer and method comparison of six point-of-care blood gas analyzers
topic Brief Communications
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8859550/
https://www.ncbi.nlm.nih.gov/pubmed/35177567
http://dx.doi.org/10.3343/alm.2022.42.4.467
work_keys_str_mv AT kimsangmi performanceevaluationoftheismartcare10analyzerandmethodcomparisonofsixpointofcarebloodgasanalyzers
AT parkhyungdoo performanceevaluationoftheismartcare10analyzerandmethodcomparisonofsixpointofcarebloodgasanalyzers