Cargando…

Digital Morphology Analyzer Sysmex DI-60 vs. Manual Counting for White Blood Cell Differentials in Leukopenic Samples: A Comparative Assessment of Risk and Turnaround Time

BACKGROUND: Digital morphology (DM) analyzers are increasingly being used for white blood cell (WBC) differentials. We assessed the laboratory efficiency of the Sysmex DI-60 system (DI-60; Sysmex, Kobe, Japan) in comparison with manual counting in leukopenic samples. METHODS: In total, 40 peripheral...

Descripción completa

Detalles Bibliográficos
Autores principales: Nam, Minjeong, Yoon, Sumi, Hur, Mina, Lee, Gun Hyuk, Kim, Hanah, Park, Mikyoung, Kim, Hyeong Nyeon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society for Laboratory Medicine 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8859564/
https://www.ncbi.nlm.nih.gov/pubmed/35177560
http://dx.doi.org/10.3343/alm.2022.42.4.398