Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy

PURPOSE: To investigate the effectiveness of deformable image registration (DIR)-based automatic contouring for tandem-ring (T-R) or tandem-ovoid (T-O) 3-dimensional computed tomography (CT)-based image-guided brachytherapy (IGBT). MATERIAL AND METHODS: CT images of 28 patients with intact cervical...

Descripción completa

Detalles Bibliográficos
Autores principales: Yedekci, Yagiz, Gültekin, Melis, Sarı, Sezin Yuce, Yıldız, Ferah
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8867230/
https://www.ncbi.nlm.nih.gov/pubmed/35233238
http://dx.doi.org/10.5114/jcb.2022.112814
_version_ 1784656009503965184
author Yedekci, Yagiz
Gültekin, Melis
Sarı, Sezin Yuce
Yıldız, Ferah
author_facet Yedekci, Yagiz
Gültekin, Melis
Sarı, Sezin Yuce
Yıldız, Ferah
author_sort Yedekci, Yagiz
collection PubMed
description PURPOSE: To investigate the effectiveness of deformable image registration (DIR)-based automatic contouring for tandem-ring (T-R) or tandem-ovoid (T-O) 3-dimensional computed tomography (CT)-based image-guided brachytherapy (IGBT). MATERIAL AND METHODS: CT images of 28 patients with intact cervical cancer were retrospectively analyzed. Selected group had T-R or T-O insertion for IGBT. Hybrid DIR was performed between first fraction CT and subsequent CTs for IGBT. First IGBT CT images were reference images. All DIRs were performed based on these first IGBT CT scans. Contour similarities between manual and automated segmentations were evaluated with dice similarity coefficient (DSC) score. Mean volumes of the structures were delineated manually and automatically compared. Finally, dosimetric comparisons were performed in order to obtain how contour differences affect the doses to target and organs at risk (OARs). RESULTS: In general, mean volumes of the automatic contours were larger than manual contours for both T-R and T-O insertions. However, the difference in volume was statistically significant for the small bowel only (p < 0.05 and p < 0.01 for T-R and T-O, respectively). The DSC scores were small for the small bowel and the sigmoid in both applicator sets. When the two different applicator sets were compared, the performance of DIR-based contour propagation for the rectum was worse in T-O compared to T-R application. Dosimetric comparisons showed that volume differences between the manual and propagated contours did not affect dose-volume parameters. The treatment plans based on manually contoured targets also well-covered DIR contours. The average time for DIR was 2.0 ±0.1 minutes per fraction compared to 14.0 ±0.4 minutes in manual contouring (p < 0.001). CONCLUSIONS: DIR-based automatic contouring of the structures seems successful for both the T-R and T-O applications in cervical IGBT. DIR significantly decreased the time for contouring. Our results indicate that automatic contouring in IGBT is safe and time-saving.
format Online
Article
Text
id pubmed-8867230
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-88672302022-02-28 Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy Yedekci, Yagiz Gültekin, Melis Sarı, Sezin Yuce Yıldız, Ferah J Contemp Brachytherapy Original Paper PURPOSE: To investigate the effectiveness of deformable image registration (DIR)-based automatic contouring for tandem-ring (T-R) or tandem-ovoid (T-O) 3-dimensional computed tomography (CT)-based image-guided brachytherapy (IGBT). MATERIAL AND METHODS: CT images of 28 patients with intact cervical cancer were retrospectively analyzed. Selected group had T-R or T-O insertion for IGBT. Hybrid DIR was performed between first fraction CT and subsequent CTs for IGBT. First IGBT CT images were reference images. All DIRs were performed based on these first IGBT CT scans. Contour similarities between manual and automated segmentations were evaluated with dice similarity coefficient (DSC) score. Mean volumes of the structures were delineated manually and automatically compared. Finally, dosimetric comparisons were performed in order to obtain how contour differences affect the doses to target and organs at risk (OARs). RESULTS: In general, mean volumes of the automatic contours were larger than manual contours for both T-R and T-O insertions. However, the difference in volume was statistically significant for the small bowel only (p < 0.05 and p < 0.01 for T-R and T-O, respectively). The DSC scores were small for the small bowel and the sigmoid in both applicator sets. When the two different applicator sets were compared, the performance of DIR-based contour propagation for the rectum was worse in T-O compared to T-R application. Dosimetric comparisons showed that volume differences between the manual and propagated contours did not affect dose-volume parameters. The treatment plans based on manually contoured targets also well-covered DIR contours. The average time for DIR was 2.0 ±0.1 minutes per fraction compared to 14.0 ±0.4 minutes in manual contouring (p < 0.001). CONCLUSIONS: DIR-based automatic contouring of the structures seems successful for both the T-R and T-O applications in cervical IGBT. DIR significantly decreased the time for contouring. Our results indicate that automatic contouring in IGBT is safe and time-saving. Termedia Publishing House 2022-01-23 2022-02 /pmc/articles/PMC8867230/ /pubmed/35233238 http://dx.doi.org/10.5114/jcb.2022.112814 Text en Copyright © 2022 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) )
spellingShingle Original Paper
Yedekci, Yagiz
Gültekin, Melis
Sarı, Sezin Yuce
Yıldız, Ferah
Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
title Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
title_full Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
title_fullStr Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
title_full_unstemmed Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
title_short Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
title_sort automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8867230/
https://www.ncbi.nlm.nih.gov/pubmed/35233238
http://dx.doi.org/10.5114/jcb.2022.112814
work_keys_str_mv AT yedekciyagiz automaticcontouringusingdeformableimageregistrationfortandemringortandemovoidbrachytherapy
AT gultekinmelis automaticcontouringusingdeformableimageregistrationfortandemringortandemovoidbrachytherapy
AT sarısezinyuce automaticcontouringusingdeformableimageregistrationfortandemringortandemovoidbrachytherapy
AT yıldızferah automaticcontouringusingdeformableimageregistrationfortandemringortandemovoidbrachytherapy