Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
PURPOSE: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model. MATERIAL AND METHODS: A reservoir filled with a diluted contrast agent was placed ins...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8867231/ https://www.ncbi.nlm.nih.gov/pubmed/35233239 http://dx.doi.org/10.5114/jcb.2022.113551 |
_version_ | 1784656009740943360 |
---|---|
author | Scherf, Christian Ramm, Ulla Stein, Thomas Trommel, Martin Tselis, Nikolaos Chatzikonstantinou, Georgios Diefenhardt, Markus Rödel, Claus Köhn, Janett Licher, Jörg |
author_facet | Scherf, Christian Ramm, Ulla Stein, Thomas Trommel, Martin Tselis, Nikolaos Chatzikonstantinou, Georgios Diefenhardt, Markus Rödel, Claus Köhn, Janett Licher, Jörg |
author_sort | Scherf, Christian |
collection | PubMed |
description | PURPOSE: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model. MATERIAL AND METHODS: A reservoir filled with a diluted contrast agent was placed inside a water phantom. A single steel needle applicator was centrally positioned inside the reservoir. Computed tomography (CT) datasets of five different contrast agent dilutions (25 to 300 mg/ml iodine concentration) were acquired, and dose calculations were performed with TG-186 ACE dose calculation formalism of Oncentra(®)Brachy (Elekta). The dose was measured with a PinPoint(®) ionization chamber (PTW) inside the contrast agent. ACE calculated and measured data were compared. RESULTS: For the different contrast agent dilutions, averaged Hounsfield units from 453 ±21 to 2623 ±221 were obtained. Electron densities derived from CT data were significantly higher than corresponding electron densities calculated from chemical compositions. Consequently, the measured dose was higher than corresponding HU-based calculated dose. Relative deviation ranged from 2.5% to 7% per 10 mm penetration depth, depending on contrast agent concentration. CONCLUSIONS: The application of HU-based TG-186 dose formalisms in the presence of high-Z contrast agent bulks overestimates electron densities. Consequently, HU-based dose calculations result in a higher delivered dose than expected from the treatment planning system. |
format | Online Article Text |
id | pubmed-8867231 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Termedia Publishing House |
record_format | MEDLINE/PubMed |
spelling | pubmed-88672312022-02-28 Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy Scherf, Christian Ramm, Ulla Stein, Thomas Trommel, Martin Tselis, Nikolaos Chatzikonstantinou, Georgios Diefenhardt, Markus Rödel, Claus Köhn, Janett Licher, Jörg J Contemp Brachytherapy Original Article PURPOSE: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model. MATERIAL AND METHODS: A reservoir filled with a diluted contrast agent was placed inside a water phantom. A single steel needle applicator was centrally positioned inside the reservoir. Computed tomography (CT) datasets of five different contrast agent dilutions (25 to 300 mg/ml iodine concentration) were acquired, and dose calculations were performed with TG-186 ACE dose calculation formalism of Oncentra(®)Brachy (Elekta). The dose was measured with a PinPoint(®) ionization chamber (PTW) inside the contrast agent. ACE calculated and measured data were compared. RESULTS: For the different contrast agent dilutions, averaged Hounsfield units from 453 ±21 to 2623 ±221 were obtained. Electron densities derived from CT data were significantly higher than corresponding electron densities calculated from chemical compositions. Consequently, the measured dose was higher than corresponding HU-based calculated dose. Relative deviation ranged from 2.5% to 7% per 10 mm penetration depth, depending on contrast agent concentration. CONCLUSIONS: The application of HU-based TG-186 dose formalisms in the presence of high-Z contrast agent bulks overestimates electron densities. Consequently, HU-based dose calculations result in a higher delivered dose than expected from the treatment planning system. Termedia Publishing House 2022-02-18 2022-02 /pmc/articles/PMC8867231/ /pubmed/35233239 http://dx.doi.org/10.5114/jcb.2022.113551 Text en Copyright © 2022 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) ) |
spellingShingle | Original Article Scherf, Christian Ramm, Ulla Stein, Thomas Trommel, Martin Tselis, Nikolaos Chatzikonstantinou, Georgios Diefenhardt, Markus Rödel, Claus Köhn, Janett Licher, Jörg Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy |
title | Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy |
title_full | Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy |
title_fullStr | Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy |
title_full_unstemmed | Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy |
title_short | Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy |
title_sort | effects of iodinated contrast agent on hu-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8867231/ https://www.ncbi.nlm.nih.gov/pubmed/35233239 http://dx.doi.org/10.5114/jcb.2022.113551 |
work_keys_str_mv | AT scherfchristian effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT rammulla effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT steinthomas effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT trommelmartin effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT tselisnikolaos effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT chatzikonstantinougeorgios effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT diefenhardtmarkus effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT rodelclaus effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT kohnjanett effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy AT licherjorg effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy |