Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy

PURPOSE: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model. MATERIAL AND METHODS: A reservoir filled with a diluted contrast agent was placed ins...

Descripción completa

Detalles Bibliográficos
Autores principales: Scherf, Christian, Ramm, Ulla, Stein, Thomas, Trommel, Martin, Tselis, Nikolaos, Chatzikonstantinou, Georgios, Diefenhardt, Markus, Rödel, Claus, Köhn, Janett, Licher, Jörg
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8867231/
https://www.ncbi.nlm.nih.gov/pubmed/35233239
http://dx.doi.org/10.5114/jcb.2022.113551
_version_ 1784656009740943360
author Scherf, Christian
Ramm, Ulla
Stein, Thomas
Trommel, Martin
Tselis, Nikolaos
Chatzikonstantinou, Georgios
Diefenhardt, Markus
Rödel, Claus
Köhn, Janett
Licher, Jörg
author_facet Scherf, Christian
Ramm, Ulla
Stein, Thomas
Trommel, Martin
Tselis, Nikolaos
Chatzikonstantinou, Georgios
Diefenhardt, Markus
Rödel, Claus
Köhn, Janett
Licher, Jörg
author_sort Scherf, Christian
collection PubMed
description PURPOSE: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model. MATERIAL AND METHODS: A reservoir filled with a diluted contrast agent was placed inside a water phantom. A single steel needle applicator was centrally positioned inside the reservoir. Computed tomography (CT) datasets of five different contrast agent dilutions (25 to 300 mg/ml iodine concentration) were acquired, and dose calculations were performed with TG-186 ACE dose calculation formalism of Oncentra(®)Brachy (Elekta). The dose was measured with a PinPoint(®) ionization chamber (PTW) inside the contrast agent. ACE calculated and measured data were compared. RESULTS: For the different contrast agent dilutions, averaged Hounsfield units from 453 ±21 to 2623 ±221 were obtained. Electron densities derived from CT data were significantly higher than corresponding electron densities calculated from chemical compositions. Consequently, the measured dose was higher than corresponding HU-based calculated dose. Relative deviation ranged from 2.5% to 7% per 10 mm penetration depth, depending on contrast agent concentration. CONCLUSIONS: The application of HU-based TG-186 dose formalisms in the presence of high-Z contrast agent bulks overestimates electron densities. Consequently, HU-based dose calculations result in a higher delivered dose than expected from the treatment planning system.
format Online
Article
Text
id pubmed-8867231
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-88672312022-02-28 Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy Scherf, Christian Ramm, Ulla Stein, Thomas Trommel, Martin Tselis, Nikolaos Chatzikonstantinou, Georgios Diefenhardt, Markus Rödel, Claus Köhn, Janett Licher, Jörg J Contemp Brachytherapy Original Article PURPOSE: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model. MATERIAL AND METHODS: A reservoir filled with a diluted contrast agent was placed inside a water phantom. A single steel needle applicator was centrally positioned inside the reservoir. Computed tomography (CT) datasets of five different contrast agent dilutions (25 to 300 mg/ml iodine concentration) were acquired, and dose calculations were performed with TG-186 ACE dose calculation formalism of Oncentra(®)Brachy (Elekta). The dose was measured with a PinPoint(®) ionization chamber (PTW) inside the contrast agent. ACE calculated and measured data were compared. RESULTS: For the different contrast agent dilutions, averaged Hounsfield units from 453 ±21 to 2623 ±221 were obtained. Electron densities derived from CT data were significantly higher than corresponding electron densities calculated from chemical compositions. Consequently, the measured dose was higher than corresponding HU-based calculated dose. Relative deviation ranged from 2.5% to 7% per 10 mm penetration depth, depending on contrast agent concentration. CONCLUSIONS: The application of HU-based TG-186 dose formalisms in the presence of high-Z contrast agent bulks overestimates electron densities. Consequently, HU-based dose calculations result in a higher delivered dose than expected from the treatment planning system. Termedia Publishing House 2022-02-18 2022-02 /pmc/articles/PMC8867231/ /pubmed/35233239 http://dx.doi.org/10.5114/jcb.2022.113551 Text en Copyright © 2022 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) )
spellingShingle Original Article
Scherf, Christian
Ramm, Ulla
Stein, Thomas
Trommel, Martin
Tselis, Nikolaos
Chatzikonstantinou, Georgios
Diefenhardt, Markus
Rödel, Claus
Köhn, Janett
Licher, Jörg
Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
title Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
title_full Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
title_fullStr Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
title_full_unstemmed Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
title_short Effects of iodinated contrast agent on HU-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
title_sort effects of iodinated contrast agent on hu-based dose calculation and dose delivered in iridium-192 high-dose-rate brachytherapy
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8867231/
https://www.ncbi.nlm.nih.gov/pubmed/35233239
http://dx.doi.org/10.5114/jcb.2022.113551
work_keys_str_mv AT scherfchristian effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT rammulla effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT steinthomas effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT trommelmartin effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT tselisnikolaos effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT chatzikonstantinougeorgios effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT diefenhardtmarkus effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT rodelclaus effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT kohnjanett effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy
AT licherjorg effectsofiodinatedcontrastagentonhubaseddosecalculationanddosedeliverediniridium192highdoseratebrachytherapy