A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods

Different laboratory methods are used to measure serum ferritin levels as a marker of iron status in the general population. This study aimed to compare serum ferritin levels using enzyme-linked immunosorbent assay (ELISA) versus immunochemiluminescence (Cobas e411) and immunoturbidimetric (Cobas In...

Descripción completa

Detalles Bibliográficos
Autores principales: Dahman, Lotfi S. Bin, Sumaily, Khalid M., Sabi, Essa M., Hassan, Mohammed A., Thalab, Abeer M. Bin, Sayad, Asrar S., Kolaib, Saleh M. Bin, Alhadhrmi, Fatima M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8870818/
https://www.ncbi.nlm.nih.gov/pubmed/35204412
http://dx.doi.org/10.3390/diagnostics12020320
_version_ 1784656848012443648
author Dahman, Lotfi S. Bin
Sumaily, Khalid M.
Sabi, Essa M.
Hassan, Mohammed A.
Thalab, Abeer M. Bin
Sayad, Asrar S.
Kolaib, Saleh M. Bin
Alhadhrmi, Fatima M.
author_facet Dahman, Lotfi S. Bin
Sumaily, Khalid M.
Sabi, Essa M.
Hassan, Mohammed A.
Thalab, Abeer M. Bin
Sayad, Asrar S.
Kolaib, Saleh M. Bin
Alhadhrmi, Fatima M.
author_sort Dahman, Lotfi S. Bin
collection PubMed
description Different laboratory methods are used to measure serum ferritin levels as a marker of iron status in the general population. This study aimed to compare serum ferritin levels using enzyme-linked immunosorbent assay (ELISA) versus immunochemiluminescence (Cobas e411) and immunoturbidimetric (Cobas Integra 400) methods in terms of sensitivity, specificity and accuracy, and whether they can be used interchangeably. A comparative cross-sectional study enrolled one hundred and six adult Yemeni patients (33 males and 73 females) aged 18–55 years, recruited from the dermatology and cosmetic center of Hadhramout Modern Hospital, Mukalla, Yemen. Serum ferritin levels were measured using ELISA, Cobas e411, and Cobas Integra 400 methods. For method comparison, a paired-sample t-test was used. For the consistency between the three methods, they were analyzed with regression and Pearson correlation coefficient. For determining accuracy, a receiver operating curve (ROC) was used. Bias error between the methods was determined through a Bland–Altman plot analysis. Our results did not show any significant statistical difference between ELISA and Cobas e411 (52.55 ± 7.4 µg/L vs. 52.58 ± 7.5 µg/L, p = 0.967), while there were significantly higher values from Cobas Integra 400 results than Cobas e411 (56.31 ± 7.8 µg/L vs. 52.58 ± 7.5 µg/L, p < 0.001) and ELISA (52.55 ± 7.4 µg/L vs. 56.31 ± 7.8 µg/L, p < 0.001). According to the correlation coefficient and linear regression analysis, a strong association between ELISA with Cobas e411 (r = 0.993, p < 0.001) and Cobas Integra 400 results (r = 0.994, p < 0.001) were revealed. For determining accuracy, Cobas e411 and Cobas Integra 400 results showed higher sensitivity (92.0%; 90.0%) and specificity (97.7%; 99.9%) respectively. Additionally, the Bland–Altman plot analysis showed a high agreement between the ELISA and Cobas e411 methods (bias: −0.035). In contrast, there was a low agreement between the ELISA and Cobas Integra 400 methods (bias: −3.75). Similarly, the agreement between Cobas e411 and Cobas Integra 400 methods was low (bias: −3.72). Serum ferritin levels were measured by Cobas e411, and Cobas Integra 400 methods were strongly correlated with the ELISA results, with higher sensitivity, specificity, and accuracy. However, further investigations with larger samples are required for improved accuracy and more precise results, and to determine whether they can be used interchangeably.
format Online
Article
Text
id pubmed-8870818
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-88708182022-02-25 A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods Dahman, Lotfi S. Bin Sumaily, Khalid M. Sabi, Essa M. Hassan, Mohammed A. Thalab, Abeer M. Bin Sayad, Asrar S. Kolaib, Saleh M. Bin Alhadhrmi, Fatima M. Diagnostics (Basel) Article Different laboratory methods are used to measure serum ferritin levels as a marker of iron status in the general population. This study aimed to compare serum ferritin levels using enzyme-linked immunosorbent assay (ELISA) versus immunochemiluminescence (Cobas e411) and immunoturbidimetric (Cobas Integra 400) methods in terms of sensitivity, specificity and accuracy, and whether they can be used interchangeably. A comparative cross-sectional study enrolled one hundred and six adult Yemeni patients (33 males and 73 females) aged 18–55 years, recruited from the dermatology and cosmetic center of Hadhramout Modern Hospital, Mukalla, Yemen. Serum ferritin levels were measured using ELISA, Cobas e411, and Cobas Integra 400 methods. For method comparison, a paired-sample t-test was used. For the consistency between the three methods, they were analyzed with regression and Pearson correlation coefficient. For determining accuracy, a receiver operating curve (ROC) was used. Bias error between the methods was determined through a Bland–Altman plot analysis. Our results did not show any significant statistical difference between ELISA and Cobas e411 (52.55 ± 7.4 µg/L vs. 52.58 ± 7.5 µg/L, p = 0.967), while there were significantly higher values from Cobas Integra 400 results than Cobas e411 (56.31 ± 7.8 µg/L vs. 52.58 ± 7.5 µg/L, p < 0.001) and ELISA (52.55 ± 7.4 µg/L vs. 56.31 ± 7.8 µg/L, p < 0.001). According to the correlation coefficient and linear regression analysis, a strong association between ELISA with Cobas e411 (r = 0.993, p < 0.001) and Cobas Integra 400 results (r = 0.994, p < 0.001) were revealed. For determining accuracy, Cobas e411 and Cobas Integra 400 results showed higher sensitivity (92.0%; 90.0%) and specificity (97.7%; 99.9%) respectively. Additionally, the Bland–Altman plot analysis showed a high agreement between the ELISA and Cobas e411 methods (bias: −0.035). In contrast, there was a low agreement between the ELISA and Cobas Integra 400 methods (bias: −3.75). Similarly, the agreement between Cobas e411 and Cobas Integra 400 methods was low (bias: −3.72). Serum ferritin levels were measured by Cobas e411, and Cobas Integra 400 methods were strongly correlated with the ELISA results, with higher sensitivity, specificity, and accuracy. However, further investigations with larger samples are required for improved accuracy and more precise results, and to determine whether they can be used interchangeably. MDPI 2022-01-27 /pmc/articles/PMC8870818/ /pubmed/35204412 http://dx.doi.org/10.3390/diagnostics12020320 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Dahman, Lotfi S. Bin
Sumaily, Khalid M.
Sabi, Essa M.
Hassan, Mohammed A.
Thalab, Abeer M. Bin
Sayad, Asrar S.
Kolaib, Saleh M. Bin
Alhadhrmi, Fatima M.
A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods
title A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods
title_full A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods
title_fullStr A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods
title_full_unstemmed A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods
title_short A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods
title_sort comparative study for measuring serum ferritin levels with three different laboratory methods: enzyme-linked immunosorbent assay versus cobas e411 and cobas integra 400 methods
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8870818/
https://www.ncbi.nlm.nih.gov/pubmed/35204412
http://dx.doi.org/10.3390/diagnostics12020320
work_keys_str_mv AT dahmanlotfisbin acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT sumailykhalidm acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT sabiessam acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT hassanmohammeda acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT thalababeermbin acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT sayadasrars acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT kolaibsalehmbin acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT alhadhrmifatimam acomparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT dahmanlotfisbin comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT sumailykhalidm comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT sabiessam comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT hassanmohammeda comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT thalababeermbin comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT sayadasrars comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT kolaibsalehmbin comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods
AT alhadhrmifatimam comparativestudyformeasuringserumferritinlevelswiththreedifferentlaboratorymethodsenzymelinkedimmunosorbentassayversuscobase411andcobasintegra400methods