Cargando…
LC–MS(3) Strategy for Quantification of Carbamazepine in Human Plasma and Its Application in Therapeutic Drug Monitoring
This study developed a detection method based on the strategy of HPLC/MS(3) and verified its suitability by quantifying carbamazepine in human plasma. The high-performance liquid chromatography–tandem mass spectrometry (HPLC/MS(3)) system was performed using a Shimadzu UFLC XR liquid chromatography...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8876099/ https://www.ncbi.nlm.nih.gov/pubmed/35209012 http://dx.doi.org/10.3390/molecules27041224 |
Sumario: | This study developed a detection method based on the strategy of HPLC/MS(3) and verified its suitability by quantifying carbamazepine in human plasma. The high-performance liquid chromatography–tandem mass spectrometry (HPLC/MS(3)) system was performed using a Shimadzu UFLC XR liquid chromatography and a SCIEX QTRAP(®) 5500 linear ion trap triple quadrupole mass spectrometer. The specific operation was as follows: the sample protein was firstly precipitated using methanol, then carbamazepine and carbamazepine-D(2)N(15) were separated on an ACQUITY UPLC HSS T3 column using the gradient elution with solvent A (0.1% formic acid) and solvent B (0.1% formic acid in acetonitrile) at a flow rate of 0.25 mL/min. Each sample was run for 7 min. This method was validated for various parameters including accuracy, precision, selectivity, linearity, LLOQ, etc. Only 5 μL of sample plasma could obtain the result of LLOD 0.5 µg/mL. The intra-day and inter-day precision was <8.23%, and accuracy was between −1.74% and 2.92%. This method was successfully used for monitoring the blood concentration of epilepsy patients after carbamazepine treatment. |
---|