Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System
Indoor air sanitizers contrast airborne diseases and particularly severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2)/Coronavirus disease 2019 (COVID-19). The commercial air sanitizer Zefero (Cf7 S.r.l., San Giovanni La Punta, Italy) works alternatively using a set of integrated disinfectin...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8877762/ https://www.ncbi.nlm.nih.gov/pubmed/35215162 http://dx.doi.org/10.3390/pathogens11020221 |
_version_ | 1784658496300515328 |
---|---|
author | Nicolò, Marco Sebastiano Rizzo, Maria Giovanna Palermo, Nicoletta Gugliandolo, Concetta Cuzzocrea, Salvatore Guglielmino, Salvatore P. P. |
author_facet | Nicolò, Marco Sebastiano Rizzo, Maria Giovanna Palermo, Nicoletta Gugliandolo, Concetta Cuzzocrea, Salvatore Guglielmino, Salvatore P. P. |
author_sort | Nicolò, Marco Sebastiano |
collection | PubMed |
description | Indoor air sanitizers contrast airborne diseases and particularly severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2)/Coronavirus disease 2019 (COVID-19). The commercial air sanitizer Zefero (Cf7 S.r.l., San Giovanni La Punta, Italy) works alternatively using a set of integrated disinfecting technologies (namely Photocatalysis/UV mode) or by generating ozone (Ozone mode). Here we evaluated the virucidal efficacy of Zefero setup modes against human Betacoronavirus OC43 and SARS-CoV-2. For this purpose, we designed a laboratory test system in which each virus, as aerosol, was treated with Photocatalysis/UV or Ozone mode and returned into a recirculation plexiglass chamber. Aerosol samples were collected after different times of exposure, corresponding to different volumes of air treated. The viral RNA concentration was determined by qRT-PCR. In Photocatalysis/UV mode, viral RNA of OC43 or SARS-CoV-2 was not detected after 120 or 90 min treatment, respectively, whereas in Ozone mode, viruses were eliminated after 30 or 45 min, respectively. Our results indicated that the integrated technologies used in the air sanitizer Zefero are effective in eliminating both viruses. As a reliable experimental system, the recirculation chamber developed in this study represents a suitable apparatus for effectively comparing the disinfection capacity of different air sanitizers. |
format | Online Article Text |
id | pubmed-8877762 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-88777622022-02-26 Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System Nicolò, Marco Sebastiano Rizzo, Maria Giovanna Palermo, Nicoletta Gugliandolo, Concetta Cuzzocrea, Salvatore Guglielmino, Salvatore P. P. Pathogens Article Indoor air sanitizers contrast airborne diseases and particularly severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2)/Coronavirus disease 2019 (COVID-19). The commercial air sanitizer Zefero (Cf7 S.r.l., San Giovanni La Punta, Italy) works alternatively using a set of integrated disinfecting technologies (namely Photocatalysis/UV mode) or by generating ozone (Ozone mode). Here we evaluated the virucidal efficacy of Zefero setup modes against human Betacoronavirus OC43 and SARS-CoV-2. For this purpose, we designed a laboratory test system in which each virus, as aerosol, was treated with Photocatalysis/UV or Ozone mode and returned into a recirculation plexiglass chamber. Aerosol samples were collected after different times of exposure, corresponding to different volumes of air treated. The viral RNA concentration was determined by qRT-PCR. In Photocatalysis/UV mode, viral RNA of OC43 or SARS-CoV-2 was not detected after 120 or 90 min treatment, respectively, whereas in Ozone mode, viruses were eliminated after 30 or 45 min, respectively. Our results indicated that the integrated technologies used in the air sanitizer Zefero are effective in eliminating both viruses. As a reliable experimental system, the recirculation chamber developed in this study represents a suitable apparatus for effectively comparing the disinfection capacity of different air sanitizers. MDPI 2022-02-08 /pmc/articles/PMC8877762/ /pubmed/35215162 http://dx.doi.org/10.3390/pathogens11020221 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Nicolò, Marco Sebastiano Rizzo, Maria Giovanna Palermo, Nicoletta Gugliandolo, Concetta Cuzzocrea, Salvatore Guglielmino, Salvatore P. P. Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System |
title | Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System |
title_full | Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System |
title_fullStr | Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System |
title_full_unstemmed | Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System |
title_short | Evaluation of Betacoronavirus OC43 and SARS-CoV-2 Elimination by Zefero Air Sanitizer Device in a Novel Laboratory Recirculation System |
title_sort | evaluation of betacoronavirus oc43 and sars-cov-2 elimination by zefero air sanitizer device in a novel laboratory recirculation system |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8877762/ https://www.ncbi.nlm.nih.gov/pubmed/35215162 http://dx.doi.org/10.3390/pathogens11020221 |
work_keys_str_mv | AT nicolomarcosebastiano evaluationofbetacoronavirusoc43andsarscov2eliminationbyzeferoairsanitizerdeviceinanovellaboratoryrecirculationsystem AT rizzomariagiovanna evaluationofbetacoronavirusoc43andsarscov2eliminationbyzeferoairsanitizerdeviceinanovellaboratoryrecirculationsystem AT palermonicoletta evaluationofbetacoronavirusoc43andsarscov2eliminationbyzeferoairsanitizerdeviceinanovellaboratoryrecirculationsystem AT gugliandoloconcetta evaluationofbetacoronavirusoc43andsarscov2eliminationbyzeferoairsanitizerdeviceinanovellaboratoryrecirculationsystem AT cuzzocreasalvatore evaluationofbetacoronavirusoc43andsarscov2eliminationbyzeferoairsanitizerdeviceinanovellaboratoryrecirculationsystem AT guglielminosalvatorepp evaluationofbetacoronavirusoc43andsarscov2eliminationbyzeferoairsanitizerdeviceinanovellaboratoryrecirculationsystem |