LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination

BACKGROUND: Urinalysis has an important place in evaluating kidney and urinary tract infections. Automated urine analyzers enhance productivity and turnover in laboratories and economize time and labor required for analysis. In the present study, we evaluated and compared analytic and diagnostic per...

Descripción completa

Detalles Bibliográficos
Autores principales: Evin, Canpolat Erkan Revşa, Aslan, Özgür
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Society of Medical Biochemists of Serbia, Belgrade 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8882015/
https://www.ncbi.nlm.nih.gov/pubmed/35291498
http://dx.doi.org/10.5937/jomb0-31711
_version_ 1784659605378301952
author Evin, Canpolat Erkan Revşa
Aslan, Özgür
author_facet Evin, Canpolat Erkan Revşa
Aslan, Özgür
author_sort Evin, Canpolat Erkan Revşa
collection PubMed
description BACKGROUND: Urinalysis has an important place in evaluating kidney and urinary tract infections. Automated urine analyzers enhance productivity and turnover in laboratories and economize time and labor required for analysis. In the present study, we evaluated and compared analytic and diagnostic performance of UriSed2 with LX-8000R, which is a novel image-based automated urine sediment analyzer. METHODS: A total of 178 urine samples sent to our laboratory were evaluated by the two urine analyzers and standard manual microscopy. Precision and comparison studies were done in accordance with CLSI guidelines. RESULTS: Sensitivity assessment revealed similar outcomes with both UriSed2 and LX-8000R devices for erythrocyte count (RBC), whereas UriSed2 device yielded higher outcomes for leukocyte count (WBC) and epithelial cells (EPI) than LX-8000R analyzer. Specificity of UriSed2 for WBC and RBC was higher than that of LX-8000R device. According to Gamma statistics, both urine analyzers showed perfect consistency for WBC, RBC and EPI cell counts. Manuel microscopy revealed statistically significant correlation between LX-8000R and UriSed2 in terms of WBC and RBC. Manual evaluation by Bland-Altman analysis demonstrated lower WBC and RBC values and higher EPI as compared to both UriSed2 and LX-8000R devices. As the result of Passing-Bablok regression analysis, both devices were found to be inconsistent with manual microscopy. CONCLUSIONS: We think that evaluation of automated urine analyzers will be more meaningful when they are evaluated together with urine samples and patient clinical findings in addition to comparing with manual microscopy.
format Online
Article
Text
id pubmed-8882015
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Society of Medical Biochemists of Serbia, Belgrade
record_format MEDLINE/PubMed
spelling pubmed-88820152022-03-14 LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination Evin, Canpolat Erkan Revşa Aslan, Özgür J Med Biochem Original Paper BACKGROUND: Urinalysis has an important place in evaluating kidney and urinary tract infections. Automated urine analyzers enhance productivity and turnover in laboratories and economize time and labor required for analysis. In the present study, we evaluated and compared analytic and diagnostic performance of UriSed2 with LX-8000R, which is a novel image-based automated urine sediment analyzer. METHODS: A total of 178 urine samples sent to our laboratory were evaluated by the two urine analyzers and standard manual microscopy. Precision and comparison studies were done in accordance with CLSI guidelines. RESULTS: Sensitivity assessment revealed similar outcomes with both UriSed2 and LX-8000R devices for erythrocyte count (RBC), whereas UriSed2 device yielded higher outcomes for leukocyte count (WBC) and epithelial cells (EPI) than LX-8000R analyzer. Specificity of UriSed2 for WBC and RBC was higher than that of LX-8000R device. According to Gamma statistics, both urine analyzers showed perfect consistency for WBC, RBC and EPI cell counts. Manuel microscopy revealed statistically significant correlation between LX-8000R and UriSed2 in terms of WBC and RBC. Manual evaluation by Bland-Altman analysis demonstrated lower WBC and RBC values and higher EPI as compared to both UriSed2 and LX-8000R devices. As the result of Passing-Bablok regression analysis, both devices were found to be inconsistent with manual microscopy. CONCLUSIONS: We think that evaluation of automated urine analyzers will be more meaningful when they are evaluated together with urine samples and patient clinical findings in addition to comparing with manual microscopy. Society of Medical Biochemists of Serbia, Belgrade 2022-02-02 2022-02-02 /pmc/articles/PMC8882015/ /pubmed/35291498 http://dx.doi.org/10.5937/jomb0-31711 Text en 2022 Canpolat Erkan Revşa Evin, Özgür Aslan, published by CEON/CEES https://creativecommons.org/licenses/by/4.0/This work is licensed under the Creative Commons Attribution 4.0 License.
spellingShingle Original Paper
Evin, Canpolat Erkan Revşa
Aslan, Özgür
LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination
title LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination
title_full LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination
title_fullStr LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination
title_full_unstemmed LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination
title_short LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination
title_sort lx-8000r and urised 2 fully automated urine analyzers comparison to manual microscopic examination
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8882015/
https://www.ncbi.nlm.nih.gov/pubmed/35291498
http://dx.doi.org/10.5937/jomb0-31711
work_keys_str_mv AT evincanpolaterkanrevsa lx8000randurised2fullyautomatedurineanalyzerscomparisontomanualmicroscopicexamination
AT aslanozgur lx8000randurised2fullyautomatedurineanalyzerscomparisontomanualmicroscopicexamination