The primary stability of the femoral component in cemented single and twin peg Oxford unicompartmental knee arthroplasty under adverse conditions

AIMS: The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation. METHODS: Single and twin pegs were implante...

Descripción completa

Detalles Bibliográficos
Autores principales: Eckert, Johannes Adrian, Bitsch, Rudi G., Sonntag, Robert, Reiner, Tobias, Schwarze, Martin, Jaeger, Sebastian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Editorial Society of Bone & Joint Surgery 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8882328/
https://www.ncbi.nlm.nih.gov/pubmed/35124977
http://dx.doi.org/10.1302/2046-3758.112.BJR-2020-0507.R2
Descripción
Sumario:AIMS: The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation. METHODS: Single and twin pegs were implanted randomly allocated in 12 pairs of human fresh-frozen femora. We generated 5° bone defects at the posterior condyle. Relative movement was simulated using a servohydraulic pulser, and analyzed at 70°/115° knee flexion. Relative movement was surveyed at seven points of measurement on implant and bone, using an optic system. RESULTS: At the main fixation zone, the twin peg shows less relative movement at 70°/115°. At the transition zone, relative movements are smaller for the single peg for both angles. The single peg shows higher compression at 70° flexion, whereas the twin peg design shows higher compression at 115°. X-displacement is significantly higher for the single peg at 115°. CONCLUSION: Bony defects should be avoided in OUKA. The twin peg shows high resilience against push-out force and should be preferred over the single peg. Cite this article: Bone Joint Res 2022;11(2):82–90.