Comparison between the Transcutaneous and Total Serum Bilirubin Measurement in Malay Neonates with Neonatal Jaundice

BACKGROUND: This study aims to investigate the reliability of the Dräger Jaundice Meter JM-105 for the screening of neonatal jaundice in Malay neonates. METHODS: A cross-sectional study was conducted in a university hospital involving 130 jaundiced neonates requiring serum bilirubin determination fr...

Descripción completa

Detalles Bibliográficos
Autores principales: Mohamed, Mazrah, Ibrahim, Nor Rosidah, Ramli, Noraida, Abdul Majid, Noorizan, Yacob, Najib Majdi, Nasir, Ariffin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Penerbit Universiti Sains Malaysia 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8887975/
https://www.ncbi.nlm.nih.gov/pubmed/35283687
http://dx.doi.org/10.21315/mjms2022.29.1.5
Descripción
Sumario:BACKGROUND: This study aims to investigate the reliability of the Dräger Jaundice Meter JM-105 for the screening of neonatal jaundice in Malay neonates. METHODS: A cross-sectional study was conducted in a university hospital involving 130 jaundiced neonates requiring serum bilirubin determination from day 2 to day 7 of life. RESULTS: The mean total serum bilirubin (TSB) was 232 μmol/L, whereas the mean transcutaneous bilirubin (TcB) measured at the forehead and sternum were 222 μmol/L and 223 μmol/L, respectively. Further, TcB underestimates TSB with a mean difference of 10.10 μmol/L at the forehead and 9.27 μmol/L at the sternum. A positive linear relationship was observed between TSB with TcB forehead (r = 0.82) and TcB sternum (r = 0.80). A good discriminations ability was observed for both the TcB forehead (receiver operating characteristics [ROC] curve = 89.8%) and sternum (ROC curve = 89.7%) at a TSB level of 205 μmol/L. The sensitivity ranges from 84.4% to 85.3%, while the specificity ranges from 77.4% to 76.4%. CONCLUSION: Our study demonstrates a strong linear relationship and good diagnostic accuracy of TcB values compared to TSB values. To conclude, TcB measured at the forehead or sternum is a good alternative as a non-invasive screening tool for non-severe hyperbilirubinemia in Malay neonates.