The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study
BACKGROUND: Commercially available wearable (ambulatory) pulse oximeters have been recommended as a method for managing patients at risk of physiological deterioration, such as active patients with COVID-19 disease receiving care in hospital isolation rooms; however, their reliability in usual hospi...
Autores principales: | , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JMIR Publications
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8889481/ https://www.ncbi.nlm.nih.gov/pubmed/35166690 http://dx.doi.org/10.2196/28890 |
_version_ | 1784661410229256192 |
---|---|
author | Santos, Mauro Vollam, Sarah Pimentel, Marco AF Areia, Carlos Young, Louise Roman, Cristian Ede, Jody Piper, Philippa King, Elizabeth Harford, Mirae Shah, Akshay Gustafson, Owen Tarassenko, Lionel Watkinson, Peter |
author_facet | Santos, Mauro Vollam, Sarah Pimentel, Marco AF Areia, Carlos Young, Louise Roman, Cristian Ede, Jody Piper, Philippa King, Elizabeth Harford, Mirae Shah, Akshay Gustafson, Owen Tarassenko, Lionel Watkinson, Peter |
author_sort | Santos, Mauro |
collection | PubMed |
description | BACKGROUND: Commercially available wearable (ambulatory) pulse oximeters have been recommended as a method for managing patients at risk of physiological deterioration, such as active patients with COVID-19 disease receiving care in hospital isolation rooms; however, their reliability in usual hospital settings is not known. OBJECTIVE: We report the performance of wearable pulse oximeters in a simulated clinical setting when challenged by motion and low levels of arterial blood oxygen saturation (SaO(2)). METHODS: The performance of 1 wrist-worn (Wavelet) and 3 finger-worn (CheckMe O2+, AP-20, and WristOx2 3150) wearable, wireless transmission–mode pulse oximeters was evaluated. For this, 7 motion tasks were performed: at rest, sit-to-stand, tapping, rubbing, drinking, turning pages, and using a tablet. Hypoxia exposure followed, in which inspired gases were adjusted to achieve decreasing SaO(2) levels at 100%, 95%, 90%, 87%, 85%, 83%, and 80%. Peripheral oxygen saturation (SpO(2)) estimates were compared with simultaneous SaO(2) samples to calculate the root-mean-square error (RMSE). The area under the receiver operating characteristic curve was used to analyze the detection of hypoxemia (ie, SaO(2)<90%). RESULTS: SpO(2) estimates matching 215 SaO(2) samples in both study phases, from 33 participants, were analyzed. Tapping, rubbing, turning pages, and using a tablet degraded SpO(2) estimation (RMSE>4% for at least 1 device). All finger-worn pulse oximeters detected hypoxemia, with an overall sensitivity of ≥0.87 and specificity of ≥0.80, comparable to that of the Philips MX450 pulse oximeter. CONCLUSIONS: The SpO(2) accuracy of wearable finger-worn pulse oximeters was within that required by the International Organization for Standardization guidelines. Performance was degraded by motion, but all pulse oximeters could detect hypoxemia. Our findings support the use of wearable, wireless transmission–mode pulse oximeters to detect the onset of clinical deterioration in hospital settings. TRIAL REGISTRATION: ISRCTN Registry 61535692; http://www.isrctn.com/ISRCTN61535692 INTERNATIONAL REGISTERED REPORT IDENTIFIER (IRRID): RR2-10.1136/bmjopen-2019-034404 |
format | Online Article Text |
id | pubmed-8889481 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | JMIR Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-88894812022-03-10 The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study Santos, Mauro Vollam, Sarah Pimentel, Marco AF Areia, Carlos Young, Louise Roman, Cristian Ede, Jody Piper, Philippa King, Elizabeth Harford, Mirae Shah, Akshay Gustafson, Owen Tarassenko, Lionel Watkinson, Peter J Med Internet Res Original Paper BACKGROUND: Commercially available wearable (ambulatory) pulse oximeters have been recommended as a method for managing patients at risk of physiological deterioration, such as active patients with COVID-19 disease receiving care in hospital isolation rooms; however, their reliability in usual hospital settings is not known. OBJECTIVE: We report the performance of wearable pulse oximeters in a simulated clinical setting when challenged by motion and low levels of arterial blood oxygen saturation (SaO(2)). METHODS: The performance of 1 wrist-worn (Wavelet) and 3 finger-worn (CheckMe O2+, AP-20, and WristOx2 3150) wearable, wireless transmission–mode pulse oximeters was evaluated. For this, 7 motion tasks were performed: at rest, sit-to-stand, tapping, rubbing, drinking, turning pages, and using a tablet. Hypoxia exposure followed, in which inspired gases were adjusted to achieve decreasing SaO(2) levels at 100%, 95%, 90%, 87%, 85%, 83%, and 80%. Peripheral oxygen saturation (SpO(2)) estimates were compared with simultaneous SaO(2) samples to calculate the root-mean-square error (RMSE). The area under the receiver operating characteristic curve was used to analyze the detection of hypoxemia (ie, SaO(2)<90%). RESULTS: SpO(2) estimates matching 215 SaO(2) samples in both study phases, from 33 participants, were analyzed. Tapping, rubbing, turning pages, and using a tablet degraded SpO(2) estimation (RMSE>4% for at least 1 device). All finger-worn pulse oximeters detected hypoxemia, with an overall sensitivity of ≥0.87 and specificity of ≥0.80, comparable to that of the Philips MX450 pulse oximeter. CONCLUSIONS: The SpO(2) accuracy of wearable finger-worn pulse oximeters was within that required by the International Organization for Standardization guidelines. Performance was degraded by motion, but all pulse oximeters could detect hypoxemia. Our findings support the use of wearable, wireless transmission–mode pulse oximeters to detect the onset of clinical deterioration in hospital settings. TRIAL REGISTRATION: ISRCTN Registry 61535692; http://www.isrctn.com/ISRCTN61535692 INTERNATIONAL REGISTERED REPORT IDENTIFIER (IRRID): RR2-10.1136/bmjopen-2019-034404 JMIR Publications 2022-02-15 /pmc/articles/PMC8889481/ /pubmed/35166690 http://dx.doi.org/10.2196/28890 Text en ©Mauro Santos, Sarah Vollam, Marco AF Pimentel, Carlos Areia, Louise Young, Cristian Roman, Jody Ede, Philippa Piper, Elizabeth King, Mirae Harford, Akshay Shah, Owen Gustafson, Lionel Tarassenko, Peter Watkinson. Originally published in the Journal of Medical Internet Research (https://www.jmir.org), 15.02.2022. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research, is properly cited. The complete bibliographic information, a link to the original publication on https://www.jmir.org/, as well as this copyright and license information must be included. |
spellingShingle | Original Paper Santos, Mauro Vollam, Sarah Pimentel, Marco AF Areia, Carlos Young, Louise Roman, Cristian Ede, Jody Piper, Philippa King, Elizabeth Harford, Mirae Shah, Akshay Gustafson, Owen Tarassenko, Lionel Watkinson, Peter The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study |
title | The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study |
title_full | The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study |
title_fullStr | The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study |
title_full_unstemmed | The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study |
title_short | The Use of Wearable Pulse Oximeters in the Prompt Detection of Hypoxemia and During Movement: Diagnostic Accuracy Study |
title_sort | use of wearable pulse oximeters in the prompt detection of hypoxemia and during movement: diagnostic accuracy study |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8889481/ https://www.ncbi.nlm.nih.gov/pubmed/35166690 http://dx.doi.org/10.2196/28890 |
work_keys_str_mv | AT santosmauro theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT vollamsarah theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT pimentelmarcoaf theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT areiacarlos theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT younglouise theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT romancristian theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT edejody theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT piperphilippa theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT kingelizabeth theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT harfordmirae theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT shahakshay theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT gustafsonowen theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT tarassenkolionel theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT watkinsonpeter theuseofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT santosmauro useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT vollamsarah useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT pimentelmarcoaf useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT areiacarlos useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT younglouise useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT romancristian useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT edejody useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT piperphilippa useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT kingelizabeth useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT harfordmirae useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT shahakshay useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT gustafsonowen useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT tarassenkolionel useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy AT watkinsonpeter useofwearablepulseoximetersinthepromptdetectionofhypoxemiaandduringmovementdiagnosticaccuracystudy |