Cargando…

Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry

BACKGROUND: Flow-diverting stents are increasingly used for the minimally-invasive treatment of intracranial aneurysms. However, a correct positioning of such devices can be challenging due to varying vessel diameters as well as the complex anatomy of the neurovasculature. As a consequence, unsucces...

Descripción completa

Detalles Bibliográficos
Autores principales: Roloff, Christoph, Berg, Philipp
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8890739/
https://www.ncbi.nlm.nih.gov/pubmed/35235597
http://dx.doi.org/10.1371/journal.pone.0264688
_version_ 1784661709944782848
author Roloff, Christoph
Berg, Philipp
author_facet Roloff, Christoph
Berg, Philipp
author_sort Roloff, Christoph
collection PubMed
description BACKGROUND: Flow-diverting stents are increasingly used for the minimally-invasive treatment of intracranial aneurysms. However, a correct positioning of such devices can be challenging due to varying vessel diameters as well as the complex anatomy of the neurovasculature. As a consequence, unsuccessful treatment outcomes are increasingly reported requiring an improvement of the understanding of stent-induced flow modification. METHODS: To evaluate the effect of different degrees of flow diverter stent malposition on intra-aneurysmal hemodynamic changes, a controlled hemodynamic configuration was created using an idealized intracranial aneurysms model. Afterwards, four different treatment scenarios were reproduced comprising of 1) the ideal treatment, 2) an insufficient wall apposition in the region of the ostium, 3) a distorted device migrating into the aneurysm sac and 4) an inaccurately deployed stent due to wrong release location. For the assessment of the individual flow modifications, high-resolution stereoscopic particle image velocimetry (PIV) measurements were carried out. RESULTS: The analysis of the precise in-vitro PIV measurements reveals that in all cases a considerable reduction of the cycle-averaged and peak-systolic velocity was obtained. Compared to the untreated aneurysm configuration, the flow reduction ranged from 63% (scenario 4) up to 89% (scenario 3). The ideal treatment reached a reduction of 78%, which is known to be sufficient for a successful therapy. However, inaccurate device positioning leads to increased oscillating flow towards the lateral directions reducing the chances of sufficient thrombus formation. CONCLUSIONS: High-resolution in-vitro PIV measurements enable an accurate quantification of the treatment efficacy for flow-diverting devices. Furthermore, insufficient treatment outcomes can be reproduces allowing for an assessment of intra-aneurysmal hemodynamic changes.
format Online
Article
Text
id pubmed-8890739
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-88907392022-03-03 Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry Roloff, Christoph Berg, Philipp PLoS One Research Article BACKGROUND: Flow-diverting stents are increasingly used for the minimally-invasive treatment of intracranial aneurysms. However, a correct positioning of such devices can be challenging due to varying vessel diameters as well as the complex anatomy of the neurovasculature. As a consequence, unsuccessful treatment outcomes are increasingly reported requiring an improvement of the understanding of stent-induced flow modification. METHODS: To evaluate the effect of different degrees of flow diverter stent malposition on intra-aneurysmal hemodynamic changes, a controlled hemodynamic configuration was created using an idealized intracranial aneurysms model. Afterwards, four different treatment scenarios were reproduced comprising of 1) the ideal treatment, 2) an insufficient wall apposition in the region of the ostium, 3) a distorted device migrating into the aneurysm sac and 4) an inaccurately deployed stent due to wrong release location. For the assessment of the individual flow modifications, high-resolution stereoscopic particle image velocimetry (PIV) measurements were carried out. RESULTS: The analysis of the precise in-vitro PIV measurements reveals that in all cases a considerable reduction of the cycle-averaged and peak-systolic velocity was obtained. Compared to the untreated aneurysm configuration, the flow reduction ranged from 63% (scenario 4) up to 89% (scenario 3). The ideal treatment reached a reduction of 78%, which is known to be sufficient for a successful therapy. However, inaccurate device positioning leads to increased oscillating flow towards the lateral directions reducing the chances of sufficient thrombus formation. CONCLUSIONS: High-resolution in-vitro PIV measurements enable an accurate quantification of the treatment efficacy for flow-diverting devices. Furthermore, insufficient treatment outcomes can be reproduces allowing for an assessment of intra-aneurysmal hemodynamic changes. Public Library of Science 2022-03-02 /pmc/articles/PMC8890739/ /pubmed/35235597 http://dx.doi.org/10.1371/journal.pone.0264688 Text en © 2022 Roloff, Berg https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Roloff, Christoph
Berg, Philipp
Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry
title Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry
title_full Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry
title_fullStr Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry
title_full_unstemmed Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry
title_short Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry
title_sort effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—an experimental framework using stereoscopic particle image velocimetry
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8890739/
https://www.ncbi.nlm.nih.gov/pubmed/35235597
http://dx.doi.org/10.1371/journal.pone.0264688
work_keys_str_mv AT roloffchristoph effectofflowdiverterstentmalpositiononintracranialaneurysmhemodynamicsanexperimentalframeworkusingstereoscopicparticleimagevelocimetry
AT bergphilipp effectofflowdiverterstentmalpositiononintracranialaneurysmhemodynamicsanexperimentalframeworkusingstereoscopicparticleimagevelocimetry