Clinical evaluation of a wearable sensor for mobile monitoring of respiratory rate on hospital wards

A wireless and wearable system was recently developed for mobile monitoring of respiratory rate (RR). The present study was designed to compare RR mobile measurements with reference capnographic measurements on a medical-surgical ward. The wearable sensor measures impedance variations of the chest f...

Descripción completa

Detalles Bibliográficos
Autores principales: Järvelä, Kristiina, Takala, Panu, Michard, Frederic, Vikatmaa, Leena
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8894146/
https://www.ncbi.nlm.nih.gov/pubmed/34476669
http://dx.doi.org/10.1007/s10877-021-00753-6
Descripción
Sumario:A wireless and wearable system was recently developed for mobile monitoring of respiratory rate (RR). The present study was designed to compare RR mobile measurements with reference capnographic measurements on a medical-surgical ward. The wearable sensor measures impedance variations of the chest from two thoracic and one abdominal electrode. Simultaneous measurements of RR from the wearable sensor and from the capnographic sensor (1 measure/minute) were compared in 36 ward patients. Patients were monitored for a period of 182 ± 56 min (range 68–331). Artifact-free RR measurements were available 81% of the monitoring time for capnography and 92% for the wearable monitoring system (p < 0.001). A total of 4836 pairs of simultaneous measurements were available for analysis. The average reference RR was 19 ± 5 breaths/min (range 6–36). The average difference between the wearable and capnography RR measurements was − 0.6 ± 2.5 breaths/min. Error grid analysis showed that the proportions of RR measurements done with the wearable system were 89.7% in zone A (no risk), 9.6% in zone B (low risk) and < 1% in zones C, D and E (moderate, significant and dangerous risk). The wearable method detected RR values > 20 (tachypnea) with a sensitivity of 81% and a specificity of 93%. In ward patients, the wearable sensor enabled accurate and precise measurements of RR within a relatively broad range (6–36 b/min) and the detection of tachypnea with high sensitivity and specificity.