Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm

PURPOSE: The dual‐layer multi‐leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf‐tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target...

Descripción completa

Detalles Bibliográficos
Autores principales: Miyasaka, Ryohei, Cho, SangYong, Hiraoka, Takuya, Chiba, Kohei, Kawachi, Toru, Katayose, Tetsurou, Suda, Yuhi, Hara, Ryusuke
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8906209/
https://www.ncbi.nlm.nih.gov/pubmed/35001518
http://dx.doi.org/10.1002/acm2.13519
_version_ 1784665360446783488
author Miyasaka, Ryohei
Cho, SangYong
Hiraoka, Takuya
Chiba, Kohei
Kawachi, Toru
Katayose, Tetsurou
Suda, Yuhi
Hara, Ryusuke
author_facet Miyasaka, Ryohei
Cho, SangYong
Hiraoka, Takuya
Chiba, Kohei
Kawachi, Toru
Katayose, Tetsurou
Suda, Yuhi
Hara, Ryusuke
author_sort Miyasaka, Ryohei
collection PubMed
description PURPOSE: The dual‐layer multi‐leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf‐tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target volume can be sensitive to the leaf‐tip transmission change. This report evaluates the dosimetric consequence due to the uncertainty of the dual‐layer MLC model in Eclipse through the dose verifications for clinical VMAT. Additionally, the Halcyon leaf‐tip model is empirically adjusted for the VMAT dose calculation with the Acuros XB. MATERIALS AND METHODS: For this evaluation, an in‐house program that analyzes the leaf position in each layer was developed. Thirty‐two clinical VMAT plans were edited into three leaf sequences: dual layer (original), proximal single layer, or distal single layer. All leaf sequences were verified using Delta4 according to the dose difference (DD) and the global gamma index (GI). To improve the VMAT dose calculation accuracy, the dosimetric leaf gap (DLG) was adjusted to minimize the DD in single‐layer leaf sequences. RESULTS: The mean of DD were −1.35%, −1.20%, and −1.34% in the dual‐layer, proximal single‐layer, and distal single‐layer leaf sequences, respectively. The changes in the mean of DD between leaf sequences were within 0.2%. However, the calculated doses differed from the measured doses by approximately 1% in all leaf sequences. The tuned DLG was increased by 0.8 mm from the original DLG in Eclipse. When the tuned DLG was used in the dose calculation, the mean of DD neared 0% and GI with a criterion of 2%/2 mm yielded a pass rate of more than 98%. CONCLUSION: No significant change was confirmed in the dose calculation accuracy between the leaf sequences. Therefore, it is suggested that the dosimetric consequence due to the leaf trailing was negligibly small in clinical VMAT plans. The DLG tuning for Halcyon can be useful for reducing the dose calculation uncertainties in Eclipse VMAT and required in the commissioning for Acuros XB.
format Online
Article
Text
id pubmed-8906209
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-89062092022-03-10 Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm Miyasaka, Ryohei Cho, SangYong Hiraoka, Takuya Chiba, Kohei Kawachi, Toru Katayose, Tetsurou Suda, Yuhi Hara, Ryusuke J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: The dual‐layer multi‐leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf‐tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target volume can be sensitive to the leaf‐tip transmission change. This report evaluates the dosimetric consequence due to the uncertainty of the dual‐layer MLC model in Eclipse through the dose verifications for clinical VMAT. Additionally, the Halcyon leaf‐tip model is empirically adjusted for the VMAT dose calculation with the Acuros XB. MATERIALS AND METHODS: For this evaluation, an in‐house program that analyzes the leaf position in each layer was developed. Thirty‐two clinical VMAT plans were edited into three leaf sequences: dual layer (original), proximal single layer, or distal single layer. All leaf sequences were verified using Delta4 according to the dose difference (DD) and the global gamma index (GI). To improve the VMAT dose calculation accuracy, the dosimetric leaf gap (DLG) was adjusted to minimize the DD in single‐layer leaf sequences. RESULTS: The mean of DD were −1.35%, −1.20%, and −1.34% in the dual‐layer, proximal single‐layer, and distal single‐layer leaf sequences, respectively. The changes in the mean of DD between leaf sequences were within 0.2%. However, the calculated doses differed from the measured doses by approximately 1% in all leaf sequences. The tuned DLG was increased by 0.8 mm from the original DLG in Eclipse. When the tuned DLG was used in the dose calculation, the mean of DD neared 0% and GI with a criterion of 2%/2 mm yielded a pass rate of more than 98%. CONCLUSION: No significant change was confirmed in the dose calculation accuracy between the leaf sequences. Therefore, it is suggested that the dosimetric consequence due to the leaf trailing was negligibly small in clinical VMAT plans. The DLG tuning for Halcyon can be useful for reducing the dose calculation uncertainties in Eclipse VMAT and required in the commissioning for Acuros XB. John Wiley and Sons Inc. 2022-01-09 /pmc/articles/PMC8906209/ /pubmed/35001518 http://dx.doi.org/10.1002/acm2.13519 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Miyasaka, Ryohei
Cho, SangYong
Hiraoka, Takuya
Chiba, Kohei
Kawachi, Toru
Katayose, Tetsurou
Suda, Yuhi
Hara, Ryusuke
Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
title Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
title_full Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
title_fullStr Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
title_full_unstemmed Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
title_short Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
title_sort investigation of halcyon multi‐leaf collimator model in eclipse treatment planning system: a focus on the vmat dose calculation with the acuros xb algorithm
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8906209/
https://www.ncbi.nlm.nih.gov/pubmed/35001518
http://dx.doi.org/10.1002/acm2.13519
work_keys_str_mv AT miyasakaryohei investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm
AT chosangyong investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm
AT hiraokatakuya investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm
AT chibakohei investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm
AT kawachitoru investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm
AT katayosetetsurou investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm
AT sudayuhi investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm
AT hararyusuke investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm