Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm
PURPOSE: The dual‐layer multi‐leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf‐tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8906209/ https://www.ncbi.nlm.nih.gov/pubmed/35001518 http://dx.doi.org/10.1002/acm2.13519 |
_version_ | 1784665360446783488 |
---|---|
author | Miyasaka, Ryohei Cho, SangYong Hiraoka, Takuya Chiba, Kohei Kawachi, Toru Katayose, Tetsurou Suda, Yuhi Hara, Ryusuke |
author_facet | Miyasaka, Ryohei Cho, SangYong Hiraoka, Takuya Chiba, Kohei Kawachi, Toru Katayose, Tetsurou Suda, Yuhi Hara, Ryusuke |
author_sort | Miyasaka, Ryohei |
collection | PubMed |
description | PURPOSE: The dual‐layer multi‐leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf‐tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target volume can be sensitive to the leaf‐tip transmission change. This report evaluates the dosimetric consequence due to the uncertainty of the dual‐layer MLC model in Eclipse through the dose verifications for clinical VMAT. Additionally, the Halcyon leaf‐tip model is empirically adjusted for the VMAT dose calculation with the Acuros XB. MATERIALS AND METHODS: For this evaluation, an in‐house program that analyzes the leaf position in each layer was developed. Thirty‐two clinical VMAT plans were edited into three leaf sequences: dual layer (original), proximal single layer, or distal single layer. All leaf sequences were verified using Delta4 according to the dose difference (DD) and the global gamma index (GI). To improve the VMAT dose calculation accuracy, the dosimetric leaf gap (DLG) was adjusted to minimize the DD in single‐layer leaf sequences. RESULTS: The mean of DD were −1.35%, −1.20%, and −1.34% in the dual‐layer, proximal single‐layer, and distal single‐layer leaf sequences, respectively. The changes in the mean of DD between leaf sequences were within 0.2%. However, the calculated doses differed from the measured doses by approximately 1% in all leaf sequences. The tuned DLG was increased by 0.8 mm from the original DLG in Eclipse. When the tuned DLG was used in the dose calculation, the mean of DD neared 0% and GI with a criterion of 2%/2 mm yielded a pass rate of more than 98%. CONCLUSION: No significant change was confirmed in the dose calculation accuracy between the leaf sequences. Therefore, it is suggested that the dosimetric consequence due to the leaf trailing was negligibly small in clinical VMAT plans. The DLG tuning for Halcyon can be useful for reducing the dose calculation uncertainties in Eclipse VMAT and required in the commissioning for Acuros XB. |
format | Online Article Text |
id | pubmed-8906209 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-89062092022-03-10 Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm Miyasaka, Ryohei Cho, SangYong Hiraoka, Takuya Chiba, Kohei Kawachi, Toru Katayose, Tetsurou Suda, Yuhi Hara, Ryusuke J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: The dual‐layer multi‐leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf‐tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target volume can be sensitive to the leaf‐tip transmission change. This report evaluates the dosimetric consequence due to the uncertainty of the dual‐layer MLC model in Eclipse through the dose verifications for clinical VMAT. Additionally, the Halcyon leaf‐tip model is empirically adjusted for the VMAT dose calculation with the Acuros XB. MATERIALS AND METHODS: For this evaluation, an in‐house program that analyzes the leaf position in each layer was developed. Thirty‐two clinical VMAT plans were edited into three leaf sequences: dual layer (original), proximal single layer, or distal single layer. All leaf sequences were verified using Delta4 according to the dose difference (DD) and the global gamma index (GI). To improve the VMAT dose calculation accuracy, the dosimetric leaf gap (DLG) was adjusted to minimize the DD in single‐layer leaf sequences. RESULTS: The mean of DD were −1.35%, −1.20%, and −1.34% in the dual‐layer, proximal single‐layer, and distal single‐layer leaf sequences, respectively. The changes in the mean of DD between leaf sequences were within 0.2%. However, the calculated doses differed from the measured doses by approximately 1% in all leaf sequences. The tuned DLG was increased by 0.8 mm from the original DLG in Eclipse. When the tuned DLG was used in the dose calculation, the mean of DD neared 0% and GI with a criterion of 2%/2 mm yielded a pass rate of more than 98%. CONCLUSION: No significant change was confirmed in the dose calculation accuracy between the leaf sequences. Therefore, it is suggested that the dosimetric consequence due to the leaf trailing was negligibly small in clinical VMAT plans. The DLG tuning for Halcyon can be useful for reducing the dose calculation uncertainties in Eclipse VMAT and required in the commissioning for Acuros XB. John Wiley and Sons Inc. 2022-01-09 /pmc/articles/PMC8906209/ /pubmed/35001518 http://dx.doi.org/10.1002/acm2.13519 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Miyasaka, Ryohei Cho, SangYong Hiraoka, Takuya Chiba, Kohei Kawachi, Toru Katayose, Tetsurou Suda, Yuhi Hara, Ryusuke Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm |
title | Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm |
title_full | Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm |
title_fullStr | Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm |
title_full_unstemmed | Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm |
title_short | Investigation of Halcyon multi‐leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm |
title_sort | investigation of halcyon multi‐leaf collimator model in eclipse treatment planning system: a focus on the vmat dose calculation with the acuros xb algorithm |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8906209/ https://www.ncbi.nlm.nih.gov/pubmed/35001518 http://dx.doi.org/10.1002/acm2.13519 |
work_keys_str_mv | AT miyasakaryohei investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm AT chosangyong investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm AT hiraokatakuya investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm AT chibakohei investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm AT kawachitoru investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm AT katayosetetsurou investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm AT sudayuhi investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm AT hararyusuke investigationofhalcyonmultileafcollimatormodelineclipsetreatmentplanningsystemafocusonthevmatdosecalculationwiththeacurosxbalgorithm |